Browse > Article

Uncertainties in Pressure Calibration of Laboratory Standard Microphones by Reciprocity Technique  

서상준 (한국표준과학연구원 물리표준부 음향진동그룹)
권휴상 (한국표준과학연구원 물리표준부 음향진동그룹)
이용봉 (한국표준과학연구원 물리표준부 음향진동그룹)
서재갑 (한국표준과학연구원 물리표준부 음향진동그룹)
Abstract
According to the Mutual Recognition Arrangement (MRA), the calibration and/or test laboratories should satisfy the management and technical requirements ISO 17025 or equivalent. Chapter 5, Section 5.10.4 of the technical requirement of ISO 17025 suggests the required informations for calibration certificates, one of them is to state the uncertainty of measurement. The uncertainties of measurement in reciprocity calibration of standard laboratory microphone were calculated. The expanded uncertainties for 1 and 1/2 inch microphones were 0.03 dB in the middle frequency range and they increased up to 0.10 dB and 0.11 dB at 20 Hz, 0.07 dB and 0.08 dB at high frequency, respectively.
Keywords
Standard microphone; Reciprocity calibration; Pressure sensitivity level; Mutual recognition arrangement(MRA); Expanded uncertainty;
Citations & Related Records
연도 인용수 순위
  • Reference
1 General Requirements for the Competance of Testing and Calibration Laboratories /
[] / ISO 17025
2 The static pressure and temperature coefficients of laboratory standard microphones /
[ Rasmussen,K. ] / Metrologia   ScienceOn
3 /
[] / Instruction Manual of B&K 4160 and 4180
4 /
[] / B&K Technical Documents of Reciprocity Calibration System Type 9699
5 /
[ 한국표준과학연구원 ] / 측정불확도 표현 지침, KRISS-99-070-SP
6 Radial wave-motion in cylindrical plane-wave couplers /
[ Rasmussen,K. ] / Acta Acustika
7 Measurement Microphones - Part 1, Specifications for Laboratory Standard Microphones /
[] / IEC 61094-1
8 /
[ ISO ] / Guide to the Expression of Uncertainty in Measurement
9 The influence of environmental conditions on the pressure sensitivity of measurement microphones /
[ Rasmussen,K. ] / B&K Technical Review
10 Measurement Microphones - Part 2, Primary Method for Pressure Calibration of Laboratory Standard Microphones by the Reciprocity Technique /
[] / IEC 61094-2
11 Controlled environment for reciprocity calibration of laboratory standard microphones and measurement of sensitivity pressure correction /
[ Wong,G.S.K.;Wu,L. ] / Metrologia   ScienceOn