Browse > Article
http://dx.doi.org/10.4046/trd.2020.0137

Measurement of Fractional Exhaled Nitric Oxide in Adults: Comparison of Two Different Analyzers (NIOX VERO and NObreath)  

Kang, Sung-Yoon (Division of Pulmonology and Allergy, Department of Internal Medicine, Gachon University Gil Medical Center)
Lee, Sang Min (Division of Pulmonology and Allergy, Department of Internal Medicine, Gachon University Gil Medical Center)
Lee, Sang Pyo (Division of Pulmonology and Allergy, Department of Internal Medicine, Gachon University Gil Medical Center)
Publication Information
Tuberculosis and Respiratory Diseases / v.84, no.3, 2021 , pp. 182-187 More about this Journal
Abstract
Background: Fractional exhaled nitric oxide (FeNO) is a non-invasive marker for eosinophilic airway inflammation and a good predictor of response to corticosteroids. There is a need for a reliable and accurate measurement method, as FeNO measurements have been widely used in clinical practice. Our study aimed to compare two FeNO analyzers and derive a conversion equation for FeNO measurements in adults. Methods: We included 99 participants who had chief complaints of chronic cough and difficulty in breathing. The participants underwent concurrent FeNO measurement using NIOX VERO (Circassia AB) and NObreath (Bedfont). We compared the values of the two devices and analyzed their correlation and agreement. We then formulated an equation to convert FeNO values measured by NObreath into those obtained by NIOX VERO. Results: The mean age of the participants was 51.2±17.1 years, with a female predominance (58.6%). Approximately 60% of the participants had asthma. The FeNO level measured by NIOX VERO (median, 27; interquartile range [IQR], 15-45) was significantly lower than that measured by NObreath (median, 38; IQR, 22-58; p<0.001). There was a strong positive correlation between the two devices (r=0.779, p<0.001). Additionally, Bland-Altman plots and intraclass correlation coefficient demonstrated a good agreement. Using linear regression, we derived the following conversion equation: natural log (Ln) (NObreath)=0.728×Ln (NIOX VERO)+1.244. Conclusion: The FeNO values of NIOX VERO and NObreath were in good agreement and had positive correlations. Our proposed conversion equation could help assess the accuracy of the two analyzers.
Keywords
Exhalation; Nitric Oxide; Biomarkers; Respiratory Disease;
Citations & Related Records
연도 인용수 순위
  • Reference
1 Tanabe Y, Harada N, Ito J, Matsuno K, Takeshige T, Harada S, et al. Difference between two exhaled nitric oxide analyzers, NIOX VERO((R)) electrochemical hand-held analyzer and NOA280i((R)) chemiluminescence stationary analyzer. J Asthma 2019;56:167-72.   DOI
2 Gustafsson LE, Leone AM, Persson MG, Wiklund NP, Moncada S. Endogenous nitric oxide is present in the exhaled air of rabbits, guinea pigs and humans. Biochem Biophys Res Commun 1991;181:852-7.   DOI
3 Maniscalco M, Vitale C, Vatrella A, Molino A, Bianco A, Mazzarella G. Fractional exhaled nitric oxide-measuring devices: technology update. Med Devices (Auckl) 2016;9:151-60.
4 Pisi R, Aiello M, Tzani P, Marangio E, Olivieri D, Chetta A. Measurement of fractional exhaled nitric oxide by a new portable device: comparison with the standard technique. J Asthma 2010;47:805-9.   DOI
5 Menzies D, Nair A, Lipworth BJ. Portable exhaled nitric oxide measurement: comparison with the "gold standard" technique. Chest 2007;131:410-4.   DOI
6 Brozek JL, Bousquet J, Agache I, Agarwal A, Bachert C, Bosnic-Anticevich S, et al. Allergic Rhinitis and its Impact on Asthma (ARIA) guidelines-2016 revision. J Allergy Clin Immunol 2017;140:950-8.   DOI
7 Global strategy for asthma management and prevention [Internet]. Fontana, WI: Global Initiative for Asthma; 2020 [cited 2020 Oct 15]. Available from: http://www.ginasthma.org.
8 Cicchetti DV. Guidelines, criteria, and rules of thumb for evaluating normed and standardized assessment instruments in psychology. Psychol Assess 1994;6:284-90.   DOI
9 Kapande KM, McConaghy LA, Douglas I, McKenna S, Hughes JL, McCance DR, et al. Comparative repeatability of two handheld fractional exhaled nitric oxide monitors. Pediatr Pulmonol 2012;47:546-50.   DOI
10 Fukuhara A, Saito J, Sato S, Sato Y, Nikaido T, Saito K, et al. Validation study of asthma screening criteria based on subjective symptoms and fractional exhaled nitric oxide. Ann Allergy Asthma Immunol 2011;107:480-6.   DOI
11 Alving K, Anolik R, Crater G, LaForce CF, Rickard K. Validation of a new portable exhaled nitric oxide analyzer, NIOX VERO: randomized studies in asthma. Pulmo Ther 2017;3: 207-18.   DOI
12 Alving K, Weitzberg E, Lundberg JM. Increased amount of nitric oxide in exhaled air of asthmatics. Eur Respir J 1993;6: 1368-70.
13 American Thoracic Society; European Respiratory Society. ATS/ERS recommendations for standardized procedures for the online and offline measurement of exhaled lower respiratory nitric oxide and nasal nitric oxide, 2005. Am J Respir Crit Care Med 2005;171:912-30.   DOI
14 Inoue Y, Sato S, Manabe T, Makita E, Chiyotanda M, Takahashi K, et al. Measurement of exhaled nitric oxide in children: a comparison between NObreath(R) and NIOX VERO(R) analyzers. Allergy Asthma Immunol Res 2018;10:478-89.   DOI
15 Khalili B, Boggs PB, Bahna SL. Reliability of a new hand-held device for the measurement of exhaled nitric oxide. Allergy 2007;62:1171-4.   DOI
16 Korn S, Wilk M, Voigt S, Weber S, Keller T, Buhl R. Measurement of fractional exhaled nitric oxide: comparison of three different analysers. Respiration 2020;99:1-8.   DOI
17 Tsuburai T, Kamide Y, Nakamura Y, Tomita Y, Hamada Y, Watai K, et al. Differences in fraction of exhaled nitric oxide values measured by two hand-helded analyzers (NObreath and NIOX Vero). Arerugi 2017;66:204-8.
18 Alving K, Janson C, Nordvall L. Performance of a new handheld device for exhaled nitric oxide measurement in adults and children. Respir Res 2006;7:67.   DOI
19 Morice AH, Millqvist E, Bieksiene K, Birring SS, Dicpinigaitis P, Domingo Ribas C, et al. ERS guidelines on the diagnosis and treatment of chronic cough in adults and children. Eur Respir J 2020;55:1901136.   DOI
20 Bland JM, Altman DG. Comparing methods of measurement: why plotting difference against standard method is misleading. Lancet 1995;346:1085-7.   DOI
21 Song WJ, Kwon JW, Kim EJ, Lee SM, Kim SH, Lee SY, et al. Clinical application of exhaled nitric oxide measurements in a Korean population. Allergy Asthma Immunol Res 2015;7:3-13.   DOI
22 Fokkens WJ, Lund VJ, Mullol J, Bachert C, Alobid I, Baroody F, et al. EPOS 2012: European position paper on rhinosinusitis and nasal polyps 2012: a summary for otorhinolaryngologists. Rhinology 2012;50:1-12.   DOI
23 Antus B, Horvath I, Barta I. Assessment of exhaled nitric oxide by a new hand-held device. Respir Med 2010;104:1377-80.   DOI