Browse > Article
http://dx.doi.org/10.5125/jkaoms.2014.40.6.266

Reconstruction plates used in the surgery for mandibular discontinuity defect  

Seol, Guk-Jin (Department of Oral and Maxillofacial Surgery, Kyungpook National University School of Dentistry)
Jeon, Eun-Gyu (Department of Oral and Maxillofacial Surgery, Kyungpook National University School of Dentistry)
Lee, Jong-Sung (Department of Oral and Maxillofacial Surgery, Kyungpook National University School of Dentistry)
Choi, So-Young (Department of Oral and Maxillofacial Surgery, Kyungpook National University School of Dentistry)
Kim, Jin-Wook (Department of Oral and Maxillofacial Surgery, Kyungpook National University School of Dentistry)
Kwon, Tae-Geon (Department of Oral and Maxillofacial Surgery, Kyungpook National University School of Dentistry)
Paeng, Jun-Young (Department of Oral and Maxillofacial Surgery, Kyungpook National University School of Dentistry)
Publication Information
Journal of the Korean Association of Oral and Maxillofacial Surgeons / v.40, no.6, 2014 , pp. 266-271 More about this Journal
Abstract
Objectives: The purpose of this study was to analyze the survival rate of reconstruction plates that were used to correct mandibular discontinuity defects. Materials and Methods: We analyzed clinical and radiological data of 36 patients. Only discontinuous mandibular defect cases were included in the study. Reconstruction plate survival rate was analyzed according to age, gender, location of defect, defect size, and whether the patient underwent a bone graft procedure, coronoidectomy, and/or postoperative radiation therapy (RT). Results: Plate-related complications developed in 8 patients, 7 of which underwent plate removal. No significant differences were found in plate survival rate according to age, gender, location of defect, defect size, or whether a bone graft procedure was performed. However, there were differences in the plate survival rate that depended on whether the patient underwent coronoidectomy or postoperative RT. In the early stages ($9.25{\pm}5.10months$), plate fracture was the most common complication, but in the later stages ($35.75{\pm}17.00months$), screw loosening was the most common complication. Conclusion: It is important to establish the time-related risk of complications such as plate fracture or screw loosening. Coronoidectomy should be considered in most cases to prevent complications. Postoperative RT can affect the survival rate and hazard rate after a reconstruction plate is fitted.
Keywords
Mandibular reconstruciton; Complications; Survival rates;
Citations & Related Records
연도 인용수 순위
  • Reference
1 Klotch DW, Gal TJ, Gal RL. Assessment of plate use for mandibular reconstruction: has changing technology made a difference? Otolaryngol Head Neck Surg 1999;121:388-92.   DOI
2 Maurer P, Eckert AW, Kriwalsky MS, Schubert J. Scope and limitations of methods of mandibular reconstruction: a long-term followup. Br J Oral Maxillofac Surg 2010;48:100-4.   DOI
3 Peacock ZS, Afshar S, Lukas SJ, Kaban LB. Customized repair of fractured mandibular reconstruction plates. J Oral Maxillofac Surg 2012;70:e563-73.   DOI
4 Knott PD, Suh JD, Nabili V, Sercarz JA, Head C, Abemayor E, et al. Evaluation of hardware-related complications in vascularized bone grafts with locking mandibular reconstruction plate fixation. Arch Otolaryngol Head Neck Surg 2007;133:1302-6.   DOI
5 Szypryt P, Forward D. The use and abuse of locking plates. Orthop Trauma 2009;23:281-90.   DOI
6 Probst FA, Mast G, Ermer M, Gutwald R, Schmelzeisen R, Pautke C, et al. MatrixMANDIBLE preformed reconstruction plates--a two-year two-institution experience in 71 patients. J Oral Maxillofac Surg 2012;70:e657-66.   DOI
7 Katakura A, Shibahara T, Noma H, Yoshinari M. Material analysis of AO plate fracture cases. J Oral Maxillofac Surg 2004;62:348-52.   DOI
8 Lindqvist C, Soderholm AL, Salo A, Subasinghe J, Ylijoki S, Skutnabb K, et al. A comparative study on four screw-plate locking systems in sheep: a clinical and radiological study. Int J Oral Maxillofac Surg 2001;30:160-6.   DOI
9 Faulkner MG, Hatcher DC, Hay A. A three-dimensional investigation of temporomandibular joint loading. J Biomech 1987;20:997-1002.   DOI
10 Kimura A, Nagasao T, Kaneko T, Tamaki T, Miyamoto J, Nakajima T. Adaquate fixation of plates for stability during mandibular reconstruction. J Craniomaxillofac Surg 2006;34:193-200.
11 Shibahara T, Noma H, Furuya Y, Takaki R. Fracture of mandibular reconstruction plates used after tumor resection. J Oral Maxillofac Surg 2002;60:182-5.   DOI   ScienceOn
12 Arias-Gallo J, Maremonti P, Gonzalez-Otero T, Gomez-Garcia E, Burgueno-Garcia M, Chamorro Pons M, et al. Long term results of reconstruction plates in lateral mandibular defects. Revision of nine cases. Auris Nasus Larynx 2004;31:57-63.   DOI
13 Nagasao T, Miyamoto J, Tamaki T, Kawana H. A comparison of stresses in implantation for grafted and plate-and-screw mandible reconstruction. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 2010;109:346-56.   DOI
14 Schoning H, Emshoff R. Primary temporary AO plate reconstruction of the mandible. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 1998;86:667-72.   DOI
15 Mariani PB, Kowalski LP, Magrin J. Reconstruction of large defects postmandibulectomy for oral cancer using plates and myocutaneous flaps: a long-term follow-up. Int J Oral Maxillofac Surg 2006;35:427-32.   DOI
16 Chepeha DB, Teknos TN, Fung K, Shargorodsky J, Sacco AG, Nussenbaum B, et al. Lateral oromandibular defect: when is it appropriate to use a bridging reconstruction plate combined with a soft tissue revascularized flap? Head Neck 2008;30:709-17.   DOI
17 Olascoaga A, Vilar-Compte D, Poitevin-Chacon A, Contreras-Ruiz J. Wound healing in radiated skin: pathophysiology and treatment options. Int Wound J 2008;5:246-57.   DOI   ScienceOn
18 Klotch DW, Prein J. Mandibular reconstruction using AO plates. Am J Surg 1987;154:384-8.   DOI
19 Marx RE. Osteoradionecrosis: a new concept of its pathophysiology. J Oral Maxillofac Surg 1983;41:283-8.   DOI   ScienceOn
20 Head C, Alam D, Sercarz JA, Lee JT, Rawnsley JD, Berke GS, et al. Microvascular flap reconstruction of the mandible: a comparison of bone grafts and bridging plates for restoration of mandibular continuity. Otolaryngol Head Neck Surg 2003;129:48-54.   DOI
21 Schmoker R, Spiessl B, Mathys R. A total mandibular plate to bridge large defects of the mandible. In: Spiessl B, Bassetti C, eds. New concepts in maxillofacial bone surgery. Berlin, New York: Springer-Verlag; 1976:156-9.
22 Gellrich NC, Suarez-Cunqueiro MM, Otero-Cepeda XL, Schon R, Schmelzeisen R, Gutwald R. Comparative study of locking plates in mandibular reconstruction after ablative tumor surgery: THORP versus UniLOCK system. J Oral Maxillofac Surg 2004;62:186-93.   DOI
23 Boyd TG, Huber KM, Verbist DE, Bumpous JM, Wilhelmi BJ. Case report removal of exposed titanium reconstruction plate after mandibular reconstruction with a free fibula osteocutaneous flap with large surgical pin cutters: a case report and literature review. Eplasty 2012;12:e42.