Browse > Article
http://dx.doi.org/10.5125/jkaoms.2011.37.2.114

Accuracy of simulation surgery of Le Fort I osteotomy using optoelectronic tracking navigation system  

Bu, Yeon-Ji (Division of Dentistry, Seoul National University)
Kim, Soung-Min (Division of Dentistry, Seoul National University)
Kim, Ji-Youn (Department of Oral & Maxillofacial Surgery, School of Dentistry, Seoul National University)
Park, Jung-Min (Dental Research Institute, Seoul National University)
Myoung, Hoon (Division of Dentistry, Seoul National University)
Lee, Jong-Ho (Division of Dentistry, Seoul National University)
Kim, Myung-Jin (Division of Dentistry, Seoul National University)
Publication Information
Journal of the Korean Association of Oral and Maxillofacial Surgeons / v.37, no.2, 2011 , pp. 114-121 More about this Journal
Abstract
Introduction: The aim of this study was to demonstrate that the simulation surgery on rapid prototype (RP) model, which is based on the 3-dimensional computed tomography (3D CT) data taken before surgery, has the same accuracy as traditional orthograthic surgery with an intermediate splint, using an optoelectronic tracking navigation system. Materials and Methods: Simulation surgery with the same treatment plan as the Le Fort I osteotomy on the patient was done on a RP model based on the 3D CT data of 12 patients who had undergone a Le Fort I osteotomy in the department of oral and maxillofacial surgery, Seoul National University Dental Hospital. The 12 distances between 4 points on the skull, such as both infraorbital foramen and both supraorbital foramen, and 3 points on maxilla, such as the contact point of both maxillary central incisors and mesiobuccal cuspal tip of both maxillary first molars, were tracked using an optoelectronic tracking navigation system. The distances before surgery were compared to evaluate the accuracy of the RP model and the distance changes of 3D CT image after surgery were compared with those of the RP model after simulation surgery. Results: A paired t-test revealed a significant difference between the distances in the 3D CT image and RP model before surgery.(P<0.0001) On the other hand, Pearson's correlation coefficient, 0.995, revealed a significant positive correlation between the distances.(P<0.0001) There was a significant difference between the change in the distance of the 3D CT image and RP model in before and after surgery.(P<0.05) The Pearson's correlation coefficient was 0.13844, indicating positive correlation.(P<0.1) Conclusion: Theses results suggest that the simulation surgery of a Le Fort I osteotomy using an optoelectronic tracking navigation system I s relatively accurate in comparing the pre-, and post-operative 3D CT data. Furthermore, the application of an optoelectronic tracking navigation system may be a predictable and efficient method in Le Fort I orthognathic surgery.
Keywords
Optoelectronic tracking navigation system; Le Fort I osteotomy; Simulation surgery; Rapid prototype (RP) model;
Citations & Related Records
연도 인용수 순위
  • Reference
1 Watzinger F, Birkfellner W, Wanschitz F, Ziya F, Wagner A, Kremser J, et al. Placement of endosteal implants in the zygoma after maxillectomy: a Cadaver study using surgical navigation. Plast Reconstr Surg 2001;107:659-67.   DOI   ScienceOn
2 Ewers R, Schicho K, Wagner A, Undt G, Seemann R, Figl M, et al. Seven years of clinical experience with teleconsultation in craniomaxillofacial surgery. J Oral Maxillofac Surg 2005;63:1447-54.   DOI   ScienceOn
3 Wagner A, Millesi W, Watzinger F, Truppe M, Rasse M, Enislidis G, et al. Clinical experience with interactive teleconsultation and teleassistance in craniomaxillofacial surgical procedures. J Oral Maxillofac Surg, 1999;57:1413-8.   DOI
4 Eliashar R, Sichel JY, Gross M, Hocwald E, Dano I, Biron A, et al. Image of guided navigation system-a new technology for complex endoscopic endonasal surgery, Postgrad Med J 2003; 79:686-90.
5 Wagner A, Ploder O, Enislidis G, Truppe M, Ewers R. Virtual image guided navigation in tumor surgery-technical innovation, J Craniomaxillofac Surg 1995;23:271-3.   DOI   ScienceOn
6 Wagner A, Rasse M, Millesi W, et al., Virtual reality for orthognathic surgery: The augmented reality environment concept, J Oral Maxillofac Surg, 55, 456 (1997).   DOI   ScienceOn
7 Ellis E, 3rd. Accuracy of model surgery: evaluation of an old technique and introduction of a new one. J Oral Maxillofac Surg 1990;48:1161-7.   DOI   ScienceOn
8 Hassfeld S, Muhling J, Zoller J. Intraoperative navigation in oral and maxillofacial surgery. Int J Oral Maxillofac Surg 1995; 24:111-9.   DOI   ScienceOn
9 Hassfeld S, Muhling J. Computer assisted oral and maxillofacial surgery-a review and an assessment of technology. Int J Oral Maxillofac Surg 2001;30:2-13.   DOI   ScienceOn
10 Wagner A, Ploder O, Enislidis G, Truppe M, Ewers R. Imageguided surgery, Int J Oral Maxillofac Surg 1996;25:147-51.   DOI   ScienceOn
11 Nizam A, Gopal RN, Naing L, Hakim AB, Samsudin AR. Dimensional Accuracy of the Skull Models Produced by Rapid Prototyping Technology Using Stereolithography Apparatus. Archives of Orofacial Sciences 2006;1:60-6.
12 Choi JY, Choi JH, Kim NK, Kim Y, Lee JK, Kim MK, et al. Analysis of errors in medical rapid prototyping models. Int J Oral Maxillofac Surg 2002;31:23-32.   DOI   ScienceOn
13 Jones RM, Khambay BS, McHugh S, Ayoub AF. The validity of a computer-assisted simulation system for orthognathic surgery (CASSOS) for planning the surgical correction of class III skeletal deformities: single-jaw versus bimaxillary surgery. Int J Oral Maxillofac Surg 2007;36:900-8.   DOI   ScienceOn
14 Ewers R, Schicho K, Undt G, Wanschitz F, Truppe M, Seemann R, et al. Basic research and 12 years of clinical experience in computer-assisted navigation technology: a review. Int J Oral Maxillofac Surg 2005;34:1-8.   DOI   ScienceOn
15 Klug C, Schicho K, Ploder O, Yerit K, Watzinger F, Ewers R, et al. Point-to-point computer-assisted navigation for precise transfer of planned zygoma osteotomies from the stereolithographic model into reality. J Oral Maxillofac Surg 2006;64:550-9.   DOI   ScienceOn
16 Schicho K, Figl M, Seemann R, Ewers R, Lambrecht JT, Wagner A, et al. Accuracy of treatment planning based on stereolithography in computer assisted surgery. Med Phys 2006;33:3408-17.   DOI   ScienceOn