Abstract
Microscopic specimen was captured by an integral imaging microscope and displayed as a three-dimensional image by an integral imaging display system. We applied the generalized relationship between pickup and display using two different lens arrays to our integral imaging microscope and display system. In order to display three-dimensional microscopic image, scaling of the captured elemental images is required. We analyzed the effect of the scaling coefficient in terms of the distortion of the displayed three-dimensional image and the loss of the captured elemental images. In our experiment, microscopic specimen is picked up by an integral imaging microscope having $125{\mu}m$ elemental lens pitch and displayed as three-dimensional image by an integral imaging display system having 1mm elemental lens pitch. The scaling coefficient was chosen to minimize the elemental image loss.
Keywords
Integral Imaging; Microscope; Scaling Coefficient; 3D Display; 3D Image Acquisition;
Citations & Related Records
Times Cited By KSCI : 2  (Citation Analysis)
연도 인용수 순위
1 J.-H. Park, G. Baasantseren, N. Kim, G. Park, J.-M. Kang, and B. Lee, 'View image generation in perspective and orthographic projection geometry based on integral imaging,' Opt. Express, 16, 8800-8813, 2008   DOI   ScienceOn
2 J.-H. Park, H. Choi, Y. Kim, and B. Lee, 'Scaling of three-dimensional integral imaging,' Japanese Journal of Applied Physics, 44, 216-224, 2005   DOI   ScienceOn
3 J.-S. Jang and B. Javidi, 'Three-dimensional integral imaging of micro-objects,' Opt. Lett., 29, 1230-1232, 2004   DOI   ScienceOn
4 권기철, 김남, 홍종면, '편광방식 입체영상 디스플레이 시스템의 의료분야 응용기술 구현,' 한국정보디스플레이학회지, 제6권, 14-20, 2005   과학기술학회마을
5 Y.-T. Lim, J.-H. Park, N. Kim, and K.-C. Kwon, 'Dense light field microscopy,' Proc. SPIE, 7237, 72371Q, 2009   DOI
6 J. Arai, H. Kawai, and F. Okano, 'Microlens arrays for integral imaging system,' Appl. Opt., 45, 9066-9078, 2006   DOI   ScienceOn
7 M. Levoy, R. Ng., A. Adams, M. Footer, and M. Horowitz, 'Light Field Microscopy', ACM Transactions on Graphics, 25, 31-42, 2006
8 N. Kim, 'Bio/Medical Applications Using High Definition 3-D Stereo Camera and Monitor System,' Optics and Photonics Congress, DTuA6, 2008
9 D.-H. Shin and E.-S. Kim, 'Computational integral imaging reconstruction of 3D Object using a depth conversion technique,' J. Opt. Soc. K., 12, 131-135, 2008   과학기술학회마을   DOI   ScienceOn
10 H. Yamanoue, M. Okui, and F. Okano, 'Geometrical analysis of puppet-theater and cardboard effects in stereoscopic HDTV images,' IEEE Trans. Circuits Syst. Video Technol., 16, 744-752, 2006   DOI   ScienceOn