Browse > Article

The Automatic Precision Measurement of RF Voltage using Power and Impedance Standards  

Shin, Jin-Kook (경기공업대학 컴퓨터정보시스템과)
Abstract
In this paper, the automatic precision measurement of RF voltage has been done using the power and impedance standards [1] in the frequency range of 50 to 1000 MHz. A coaxial microcalorimeter and an automatic network analyzer were used for the determination of the RF-DC differences and the total uncertainty is about 1.0 %. A HP computer, a commodore computer and IEEE-488 interface bus were used for measuring the effective efficiency of thermistor mount and the RF-DC difference of thermal voltage converter, All processes of measurement were accomplished by self-developed program automatically.
Keywords
RF voltage; Microcalorimeter; Thermistor mount; Effective efficiency; RF-DC difference;
Citations & Related Records
연도 인용수 순위
  • Reference
1 Nak Sam Chung, Joon Shin, Horst Bayer, and Rolf Honigbaum, 'Coaxial and Waveguide Microcalorimeters for RF and Microwave Power Standards,'IEEE Trans. Instrum. Meas., vol. 38, No. 2, pp. 460-464, April 1989   DOI   ScienceOn
2 R. F. Clark and A. P. Jurkus, 'Calibration of thermal transfer standard of RF voltage, 'IEEE Trans. Instrum. Meas., vol. IM-16, pp. 232-237, Sept. 1967
3 J. A. Stratton, Electromagnetic Theory, McGRAWHILL Book Company, New York and London, 1941
4 N. Marcuwitz, Waveguide Handbook, New York: Dover Pub., Inc., Ch. 5, pp. 310-312, 1965