Abstract
Humidity calibration for a temperature-stable quartz oscillator (TSQO) was investigated to exclude the influences of relative humidity on the TSQO output in order to use the corresponding devices outdoors. The TSQO output is a voltage that is inversely proportional to the electric impedance of the quartz oscillator, which depends on the viscosity and density of the measured gas. The TSQO output was humidity calibrated using its humidity dependence, which was obtained by varying the relative humidity (RH) from 0 to 100 RH% while other conditions were kept constant. The humidity dependencies of the TSQO output were fit by a linear function. Subtracting the change in the TSQO output induced by the change in humidity, calculated with the function from the experimentally measured TSQO output for a range of 0-100RH%, eliminated the influence of humidity on the TSQO output. The humidity calibration succeeded in reducing the fluctuations of the TSQO output from 0.4-3% to 0.1-0.3% of the average values for a range of 0-100RH%, at constant temperatures. The necessary stability of the TSQO output for application in hydrogen sensors was below one-third of the change observed for a hydrogen leakage of 1 vol.% hydrogen concentration, corresponding to 0.33% of the change in each background. Therefore, the results in this study indicate that the present humidity calibration effectively suppresses the influence of humidity, for the TSQO output for use as an outdoor hydrogen sensor.
Keywords
Quartz oscillator; Humidity calibration; Pressure gauge; Hydrogen sensing; Viscosity measurement;
Citations & Related Records
연도 인용수 순위
  • Reference
1 K. Kokubun, M. Hirata, H. Murakami, Y. Toda, and M. Ono, Vacuum 34 731 (1984).   DOI
2 A. Kurokawa, K. Odaka, and S. Ichimura, Vacuum 73 301 (2004).   DOI
3 A. Suzuki, A. Kurokawa, H. Nonaka, and S. Ichimura, Sens. Actuators, A 127 37 (2006).   DOI
4 A. Suzuki and H. Nonaka, Jpn. J. Appl. Phys. 47 3661 (2008).   DOI
5 A. Suzuki and H. Nonaka, Int. J. Hydrogen Energy 33 6385 (2008).   DOI
6 A. Suzuki, A. Kurokawa, and H. Nonaka, Plasma Monitoring Using a Quartz Sensor, Proc. 18th Int. Symp. Pla. Chem., 2008, p. 30P.
7 A. Suzuki and H. Nonaka, Rev. Sci. Instrum. 80 095109 (2009).   DOI
8 A. Suzuki and H. Nonaka, Vacuum 84 554 (2009).   DOI
9 A. Suzuki and H. Nonaka, Measurement for Dissociation Ratio of Ammonia in Plasmas Using a Quartz Sensor, Proceedings of 27th Symposium on Plasma Processing, 2010, p. A5.
10 A. Suzuki and H. Nonaka, Vacuum 84 1389 (2010).   DOI
11 A. Suzuki and H. Nonaka, Jpn. J. Appl. Phys. 50 01AA03 (2011).   DOI
12 A. Suzuki, Hydrogen Sensing Characteristics of a Quartz Oscillator, 2012.
13 A. Suzuki and S. Asahina, Jpn. J. Appl. Phys. 51 01AA03 (2012).   DOI
14 A. Suzuki and S. Asahina, Pressure dependence of optical emissions in nitrogen-hydrogen plasmas, Proceedings of 30th Symposium on Plasma Processing, 2013, p. A4.
15 A. Suzuki and S. Asahina, Jpn. J. Appl. Phys. 52 in press. (2013).
16 A. Suzuki and S. Asahina, Spatial distributions of non-molecular species in N2-H2 plasmas measured by a quartz oscillator, Proc. 21st Int. Symp. Pla. Chem., 2013, p. 303.
17 A. Suzuki and S. Asahina, Jpn. J. Appl. Phys. 52 11NC04 (2013).   DOI
18 A. Suzuki and S. Asahina, Jpn. J. Appl. Phys. 53 11RC03 (2014).   DOI
19 A. Suzuki, H. Hojo, and T. Kobayashi, Vacuum 121 255 (2015).   DOI
20 A. Suzuki, J. Vac. Sci. Technol. A 34 031604 (2016).   DOI
21 B. Dulmet and R. Bourquin, Revue De Physique Appliquee 18 619 (1983).   DOI
22 I. F. Golubev, Viscosity of gases and gas mixtures, A Handbook, Israel Program for Scientific Translations, Jerusalem, 1970.
23 O. Tetens and Z. Geophys. 6 297 (1930).