The Potential Role of Grid-Like Software in Bedside Chest Radiography in Improving Image Quality and Dose Reduction: An Observer Preference Study  

Ahn, Su Yeon (Department of Radiology, Seoul National University College of Medicine, and Institute of Radiation Medicine, Seoul National University Medical Research Center)
Chae, Kum Ju (Department of Radiology, Seoul National University College of Medicine, and Institute of Radiation Medicine, Seoul National University Medical Research Center)
Goo, Jin Mo (Department of Radiology, Seoul National University College of Medicine, and Institute of Radiation Medicine, Seoul National University Medical Research Center)
Publication Information
Korean Journal of Radiology / v.19, no.3, 2018 , pp. 526-533 More about this Journal
Abstract
Objective: To compare the observer preference of image quality and radiation dose between non-grid, grid-like, and grid images. Materials and Methods: Each of the 38 patients underwent bedside chest radiography with and without a grid. A grid-like image was generated from a non-grid image using SimGrid software (Samsung Electronics Co. Ltd.) employing deep-learning-based scatter correction technology. Two readers recorded the preference for 10 anatomic landmarks and the overall appearance on a five-point scale for a pair of non-grid and grid-like images, and a pair of grid-like and grid images, respectively, which were randomly presented. The dose area product (DAP) was also recorded. Wilcoxon's rank sum test was used to assess the significance of preference. Results: Both readers preferred grid-like images to non-grid images significantly (p < 0.001); with a significant difference in terms of the preference for grid images to grid-like images (p = 0.317, 0.034, respectively). In terms of anatomic landmarks, both readers preferred grid-like images to non-grid images (p < 0.05). No significant differences existed between grid-like and grid images except for the preference for grid images in proximal airways by two readers, and in retrocardiac lung and thoracic spine by one reader. The median DAP were 1.48 (range, 1.37-2.17) $dGy*cm^2$ in grid images and 1.22 (range, 1.11-1.78) $dGy*cm^2$ in grid-like images with a significant difference (p < 0.001). Conclusion: The SimGrid software significantly improved the image quality of non-grid images to a level comparable to that of grid images with a relatively lower level of radiation exposure.
Keywords
Bedside chest radiography; Scatter correction; Software; Image quality; Dose reduction; Grid;
Citations & Related Records
Times Cited By KSCI : 1  (Citation Analysis)
연도 인용수 순위
1 Mentrup D, Jockel S, Menser B, Neitzel U. Iterative scatter correction for grid-less bedside chest radiography: performance for a chest phantom. Radiat Prot Dosimetry 2016;169:308-312   DOI
2 Renger B, Brieskorn C, Toth V, Mentrup D, Jockel S, Lohofer F, et al. Evaluation of dose reduction potentials of a novel scatter correction software for bedside chest X-ray imaging. Radiat Prot Dosimetry 2016;169:60-67   DOI
3 Goo JM, Im JG, Kim JH, Seo JB, Kim TS, Shine SJ, et al. Digital chest radiography with a selenium-based flat-panel detector versus a storage phosphor system: comparison of soft-copy images. AJR Am J Roentgenol 2000;175:1013-1018   DOI
4 Floyd CE Jr, Baker JA, Lo JY, Ravin CE. Measurement of scatter fractions in clinical bedside radiography. Radiology 1992;183:857-861   DOI
5 Anderson DW. Introduction of grids to mobile ICU radiography in a teaching hospital. Br J Radiol 2006;79:315-318   DOI
6 Moore CS, Wood TJ, Avery G, Balcam S, Needler L, Smith A, et al. Investigating the use of an antiscatter grid in chest radiography for average adults with a computed radiography imaging system. Br J Radiol 2015;88:20140613   DOI
7 Uffmann M, Schaefer-Prokop C. Digital radiography: the balance between image quality and required radiation dose. Eur J Radiol 2009;72:202-208   DOI
8 Mentrup D, Specht S, Spieckermann J, Zorba G. Gridless imaging with SkyFlow: time savings and workflow improvements. Web site. https://www.usa.philips.com/b-dam/b2bhc/us/topics/radiology-patience-experience/SkyFlowwhitepaper-452299111471_LR.PDF. Published June 2015. Accessed July 13, 2017
9 Rubinowitz AN, Siegel MD, Tocino I. Thoracic imaging in the ICU. Crit Care Clin 2007;23:539-573   DOI
10 Eisenhuber E, Schaefer-Prokop CM, Prosch H, Schima W. Bedside chest radiography. Respir Care 2012;57:427-443   DOI
11 Wandtke JC. Bedside chest radiography. Radiology 1994;190:1-10   DOI
12 Kawamura T, Naito S, Okano K, Yamada M. Improvement in image quality and workflow of x-ray examinations using a new image processing method, "Virtual Grid Technology". Web site. http://www.fujifilm.com/about/research/report/060/pdf/index/ff_rd060_004_en.pdf. Published 2015. Accessed July 13, 2017
13 Chae KJ, Goo JM, Ahn SY, Yoo JY, Yoon SH. Application of deconvolution algorithm of point spread function in improving image quality: an observer preference study on chest radiography. Korean J Radiol 2018;19:147-152   DOI
14 MacMahon H, Yasillo NJ, Carlin M. Laser alignment system for high-quality portable radiography. Radiographics 1992;12:111-120   DOI
15 Scott AW, Gauntt DM, Yester MV, Barnes GT. High-ratio grid considerations in mobile chest radiography. Med Phys 2012;39:3142-3153   DOI
16 Lee K, Lee K, Lee S, Kim H. Workflow improvement with Samsung SimGrid SW. Web site. https://www.samsungmedicalsolution.com/en/common/whitePaper. Published February 22, 2016. Accessed July 13, 2017
17 Gauntt DM, Barnes GT. An automatic and accurate x-ray tube focal spot/grid alignment system for mobile radiography: system description and alignment accuracy. Med Phys 2010;37:6402-6410   DOI