Browse > Article
http://dx.doi.org/10.5478/MSL.2013.4.4.91

Simulation Study to Improve Ion Transmission Efficiency in Mass Spectrometry with a Dual Ion Funnel Ion Source  

Baek, Sun Jong (Division of Mass Spectrometry Research, Korea Basic Science Institute)
Kim, Seung Yong (Division of Mass Spectrometry Research, Korea Basic Science Institute)
Kim, Taeman (MS Development Group, Genomic System Research Center, Bioneer Corporation)
Kim, Hyun Sik (Division of Mass Spectrometry Research, Korea Basic Science Institute)
Publication Information
Mass Spectrometry Letters / v.4, no.4, 2013 , pp. 91-94 More about this Journal
Abstract
We performed computer simulations to improve transmission efficiencies of a dual ion funnel system implemented on an FT-ICR MS. We found that the low m/z range from 50 to 150 could be significantly improved by operating the two ion funnels at different RF amplitudes. These new operational conditions could be applied to analyze metabolome samples, which require high sensitivity in the m/z range from 50 to 1,000.
Keywords
FT-ICR; ion funnel; trajectory calculation; simulations; SIMION;
Citations & Related Records
연도 인용수 순위
  • Reference
1 Shaffer, S. A.; Tang, K.; Anderson, G. A.; Prior, D. C.; Udseth, H. R.; Smith, R. D. Rapid Communications in Mass Spectrometry 1997, 11, 1813   DOI
2 Kim, T.; Tolmachev, A. V.; Harkewicz, R.; Prior, D. C.; Anderson, G.; Udseth, H. R.; Smith, R. D. Anal. Chem. 2000, 72, 2247   DOI   ScienceOn
3 Kim, T.; Park, M. A. "Ion guide for mass spectrometers", US Patent 7,495,212.
4 Cobb, J. S.; Easterling, M. L.; Agar, J. N. J. Am. Soc. Mass Spectrom. 2010, 21 (6), 949.   DOI   ScienceOn
5 Dahl, D. A. SIMION 3D (Version 8.1) Idaho National Engineering and Environmental Laboratory; Idaho Falls, ID, 2000.
6 Tolmachev, A. V.; Kim, T.; Udseth, H. R.; Smith, R. D.; Bailey, T. H.; Futrell, J. H. Int. J. Mass Spectrom. 2000, 203, 31.   DOI   ScienceOn
7 Kim, T. Ph.D. Thesis, 1997, p154, McGill University.