Abstract
We measured radial arterial pulse signals using a prototype of a clip-type pulsimeter equipped with a permanent magnet and a Hall device, which produced signals through a voltage-detecting circuit. The systolic peak time and the reflective peak time for a temporally pulsed signal were analyzed for an arbitrary pulse wave at one position of a small permanent magnet. The measured value of the peripheral pulse wave velocity was about 1.25-1.52 m/s, demonstrating the accuracy of this new method. To measure the peripheral blood flow velocity, we simultaneously connected the radial artery pulsimeter to a photoplethysmography meter. The average value of the peripheral blood flow velocity was about 0.27-0.50 m/s.
Keywords
radial artery pulse wave; clip-type pulsimeter; photoplethysmography; systolic peak time; reflective peak time; peripheral blood flow velocity; peripheral pulse wave velocity;
Citations & Related Records
Times Cited By KSCI : 4  (Citation Analysis)
연도 인용수 순위
1 S. S. Lee, D. H. Nam, Y. S. Hong, W. B. Lee, I. H. Son, K. H. Kim, and J. G. Choi, Sensors 11, 1784 (2011).   DOI
2 D. H. Nam, W. B. Lee, Y. S. Hong, and S. S. Lee, Sensors 13, 4714 (2013).   DOI   ScienceOn
3 R. Asmar, A. Benetos, G. M. London, C. Hughe, Y. Weiss, J. Topouchian, B. Laloux, and M. Safar, Blood Pressure 4, 48 (1995).   DOI
4 Q. Yu, J. Zhou, and Y. C. Fung, Am. J. Physiol. Heart Circ. Physiol. 265, 52 (1993).
5 M. F. P. O'Rourke, R. P. Kelly, and A. P. Avolio, The Arterial Pulse, 1st Ed.; Lea & Febiger, Philadelphia, PA, USA (1992).
6 A. Yamashina, H. Tomiyama, T. Arai, Y. Koji, M. Yambe, H. Motobe, Z. Glunizia, Y. Yamamoto, and S. Hori, Hypertens. Res. 26, 801 (2003).   DOI   ScienceOn
7 J. C. Bramwell and A. V. Hill, Proceedings of the Royal Society, London (1922) p. 298.
8 John Allen, Photoplethysmography and its Application in Clinical Physiological Measurement 28, R1 (2007).
9 K. Shelley and S. Shelley, Pulse Oximeter Waveform: Olethysmography, in Clinical Monitorning, Carol Lake, R. Hines, and C. Blitt, Eds.: W. B. Saunders Company (2001) p. 420.
10 A. Fronek, Noninvasive diagnosis in vascular disease, McGraw Hill, New York (1989).
11 I. H. Son and S. S. Lee, J. Korean Magn. Soc. 21, 104 (2011).   DOI   ScienceOn
12 M. C. Ahn, J. G. Choi, I. H. Son, S. S. Lee, and K. H. Kim, J. Korean Magn. Soc. 20, 106 (2010).   DOI   ScienceOn
13 S. S. Lee, M. C. Ahn, and S. H. Ahn, J. Magn. 14, 132 (2009).   DOI   ScienceOn
14 S. S. Lee, I. H. Son, J. G. Choi, D. H. Nam, Y. S. Hong, and W. B. Lee, J. Korean Phys. Soc. 58, 349 (2011).   DOI   ScienceOn