Safety Requirements and Test Methods of a Radiofrequency Stimulator  

Park, Sang-Geon (Division of Mechatronics Engineering, Silla University)
Abstract
In this study, we investigate the safety requirements and test methods of a radiofrequency stimulator. The main test items include controls of a minimum output, accommodation range, and output parameters that have been known as the safety requirements in conformity with international standards. As the test criteria for controlling the minimum output, an increase or decrease in a unit of 1 mA or 1 V or less was applied to the output amplitude regulator for both continuous and discontinuous control, and the output at the minimum setting was manipulated to not exceed 2% of the maximum setting. For controlling the output parameters, one of the representative test criteria states that the current limit of 250 mA should be equal to or less than 1,500 Hz. Consequently, when applying the radiofrequency stimulator on the human body, we need to ensure that the safety requirements conform to the international standards.
Keywords
Output parameter; Radiofrequency; Stimulator;
Citations & Related Records
연도 인용수 순위
  • Reference
1 S. Erdine, A. Yucel, A. Cimen, S. Aydin, A. Sav, and A. Bilir, "Effects of pulsed versus conventional radiofrequency current on rabbit dorsal root ganglion morphology," European Journal of Pain, vol. 9, no. 3, pp. 251-256, 2005.   DOI
2 K. M. Malik, "Pulsed radiofrequency, water-cooled radiofrequency, and cryoneurolysis," in Essentials of Pain Medicine. St. Louis, MO: Elsevier, pp. 431-438, 2011.
3 W. H. Sweet, "Controlled thermocoagulation of trigeminal ganglion and rootlets for differential destruction of pain fibers: facial pain other than trigeminal neuralgia," Clinical Neurosurgery, vol. 23, pp. 96-102, 1976.   DOI
4 S. S. Uematsu, "Percutaneous electrothermocoagulation of spinal nerve trunk, ganglion and rootlets," in Current Techniques in Operative Neurosurgery. New York: Grunne & Stratton, pp. 469-490, 1977.
5 M. E. Sluijter, "Pulsed radiofrequency," Anesthesiology, vol. 103, no. 6, p. 1313, 2005.   DOI
6 M. E. Sluijter, E. R. Cosman, W. B. Rittman III, and M. Van Kleef, "The effects of pulsed radiofrequency fields applied to the dorsal root ganglion: a preliminary report," The Pain Clinic, vol. 11, no. 2, pp. 109-117, 1998.
7 E. R. Cosman Jr. and E. R. Cosman Sr., "Electric and thermal field effects in tissue around radiofrequency electrodes," Pain Medicine, vol. 6, no. 6, pp. 405-424, 2005.   DOI
8 M. Barbieri and M. Bellini, "Radiofrequency neurotomy for the treatment of chronic pain: interference with implantable medical devices," Anaesthesiology Intensive Therapy, vol. 46, no. 3, pp. 162-165, 2014.   DOI
9 J. P. Mitchell, and G. N. Lumb, A Handbook of Surgical Diathermy. Bristol: John Wright & Sons, 1966.
10 US Food and Drug Administration, "Guidance Document for Powered Muscle Stimulator 510(k)s," 1999 [Internet]. Available: http://www.fda.gov/downloads/medicaldevices/deviceregulationandguidance/guidancedocuments/ucm073783.pdf.
11 "Medical electrical equipment - Part 2-10: Particular requirements for the basic safety and essential performance of nerve and muscle stimulators," IEC 60601-2-10:2012, 2012.
12 "Medical electrical equipment - Part 2-10: Particular requirements for the basic safety and essential performance of nerve and muscle stimulators," IEC 60601-2-10:2012+AMD1:2016 CSV Consolidated version, 2016.