Feasibility of Using the Clamp Meter in Measuring X-Ray Tube Current  

Kim, Sung-Chul (Department of Radiological science, Gachon University Medical campus)
Publication Information
Abstract
The clamp meter maintains electric safety as a non-invasive method while measuring the absolute value of tube current with it has been recently developed for an X-ray high-tension cable. Especially this can show high accuracy at short X-ray exposure time. Considering such a condition, this study is to evaluate the feasibility of a clamp meter in measuring X-ray tube current by taking the measurements and comparing with those of the Dynalyzer III which has been considered as a standard measuring device. From measuring the tube current accuracy depending on changes in tube voltage and exposure time, the clamp meter showed higher accuracy rate which was -1.3~4.2% difference. Thus clamp meter can be used for clinical radiologists who are not familiar electric circuit to manage X-ray devices easily and correctly in the future.
Keywords
X-ray tube current; Clamp meter; Non-invasive method; X-ray tube voltage;
Citations & Related Records
Times Cited By KSCI : 2  (Citation Analysis)
연도 인용수 순위
1 S.S. Kang, D.G. Kwon, and K.K. Kim, Diagnostic X-ray Equipment 2nd ed., Chung-ku Pub. co., 2011.
2 J.M. Kim, S.C. Kim and S.G. Ko, X-ray Equipment & Management Experiments, Daihak Pub. co., 2000.
3 R.R. Carlton, A.M. Adler, Principles of Radiographic imaging-An art and a science 4th Ed., Thomson, 2006.
4 T.T. Thompson, A practical Approach to Modern X-ray Equipment, Brown and co., 1978.
5 S.S. Kang, C.S. Kim and S.J. Ko, "Usefulness of Non- Invasive Measurement Tool on Performance Evaluation of Inverter Type X-ray Unit", J. of Radiological Science & Technology, vol. 31, no. 2, 2008, pp. 123-127.
6 B.S. Kang, K.M. Lee, W.Y. Shim, S.C. Park, H.D. Choi and Y.K. Cho, "Analyze for the Quality Control of General X-ray Systems in Capital region", J. of Radiological Science & Technology, vol. 35, no. 2, 2012, pp. 93-102.
7 I.C. Im, "An ability test for the use of Indirect Radiographic unit", J. of Radiological Science & Technology, vol. 27, no. 4, 2004, pp. 37-41.
8 J.H. Park, I.C. lm, K.R. Dong and S.S. Kang, "A Performance Evaluation of Diagnostic X-ray Unit Depends on the Hospitals Size", J. of Radiation Protection, vol. 34 no. 1, 2009, pp. 31-36.   과학기술학회마을
9 J.K. Park, "Medical Clinics'quality Management of X-ray Units in Gyeongbuk Area", J. of the Korea Contents Association, vol. 10, no. 9, 2010, pp. 267-275.
10 P. Carter, A. Paterson, M. Thornton, A. Hyatt, A. Milne, and J. Pirrie, Chesneys' Equipment for Student Radiographers 4th Ed., Blackwell scientific Pub., 1994, pp. 64-76.
11 J.D. Ramboz, "A Highly Accurate hand-held Clamp-on Current Transformer", IEEE trans. Instrum., vol. 45, no. 2, 1996, pp. 445-448.   DOI   ScienceOn
12 E. So, D.A. Bennett, "A low-current Multistage Clampon Current Transformer with Ratio errors below 50 ppm", IEEE trans. Instrum., vol. 46, no. 2, 1997, pp. 454-458.   DOI   ScienceOn
13 Y.M. chang, S.I. kim, "Development of Clamp Current Meter using a Flexible Rogowski coil", Kieeme J. of Annual autumn conference, 2002, pp. 472-475.