Evaluation of HalcyonTM Fast kV CBCT effectiveness in radiation therapy in cervical cancer patients of childbearing age who performed ovarian transposition  

Lee Sung Jae (Department of Radiation Oncology, Asan Medical Center)
Shin Chung Hun (Department of Radiation Oncology, Asan Medical Center)
Choi So Young (Department of Radiation Oncology, Asan Medical Center)
Lee Dong Hyeong (Department of Radiation Oncology, Asan Medical Center)
Yoo Soon Mi (Department of Radiation Oncology, Asan Medical Center)
Song Heung Gwon (Department of Radiation Oncology, Asan Medical Center)
Yoon In Ha (Department of Radiation Oncology, Asan Medical Center)
Publication Information
The Journal of Korean Society for Radiation Therapy / v.34, no., 2022 , pp. 73-82 More about this Journal
Abstract
Purpose: The purpose of this study is to evaluate the effectiveness of reducing the absorbed dose to the ovaries and the quality of the CBCT image when using the HalcyonTM Fast kV CBCT of cervical cancer patients of child-bearing age who performed ovarian transposition Materials and Methods : Contouring of the cervix and ovaries required for measurement was performed on the computed tomography images of the human phantom (Alderson Rando Phantom, USA), and three Optically Stimulated Luminescence Dosimeter(OSLD) were attached to the selected organ cross-section, respectively. In order to measure the absorbed dose to the cervix and ovaries in the TruebeamTM pelvis mode (Hereinafter referred to as TP), The HalcyonTM Pelvis mode (Hereinafter referred to as HP) and The HalcyonTM Pelvis Fast mode (Hereinafter referred to as HPF), An image was taken with a scan range of 17.5 cm and also taken an image that reduced the Scan range to 12.5cm. A total of 10 cumulative doses were summed, It was replaced with a value of 23 Fx, the number of cervical cancer treatments, and compared In additon, uniformity, low contrast visibility, spatial resolution, and geometric distortion were compared and analyzed using Catphan 504 phantom to compare CBCT image quality between equipment. Each factor was repeatedly measured three times, and the average value was obtained by analysing with the Doselab (Mobius Medical Systems, LP. Versions: 6.8) program. Results: As a result of measuring absorbed dose by CBCT with OSLD, TP and HP did not obtain significant results under the same conditions. The mode showing the greatest reduction value was HPF versus TP. In HPF, the absorbed dose was reduced by 39.8% in the cervix and 19.8% in the ovary compared to the TP in the scan range of 17.5 cm. the scan range was reduced to 12.5 cm, absorbed dose was reduced by 34.2% in the cervix and 50.5% in the ovary. In addition, result of evaluating the quality of the image used in the above experiment, it complied with the equipment manufacturer's standards with Geometric Distortion within 1mm (SBRT standard), Uniformity HU, LCV within 2.0%, Spatial Resolution more than 3 lp/mm. Conclusion: According to the results of this experiment, HalcyonTM can select more various conditions than TruebeamTM in treatment of fertility woman who have undergone ovarian Transposition , because it is important to reduce the radiation dose by CBCT during radiation therapy. So finally we recommend HalcyonTM Fast kV CBCT which maintains image quality even at low mAs. However, it is consider that the additional exposure to low doses can be reduced by controlling the imaging range for patients who have undergone ovarian transposition in other treatment machines.
Keywords
$Halcyon^{TM}$; Fast kV CBCT; Ovarian Transposition; Absorbed dose;
Citations & Related Records
Times Cited By KSCI : 1  (Citation Analysis)
연도 인용수 순위
1 Seri Hong, Young-Joo Won, Jae Jun Lee et al. Cancer Statistics in Korea: Incidence, Mortality, Survival, and Prevalence in 2018
2 JY Wo, AN. Viswanathan. Impact of radiotherapy on fertility, pregnancy, and neonatal outcomes in female cancer patients Int J Radiat Oncol Biol Phys, 73 (2009), pp. 1304-1312   DOI
3 Lina Yin, Saiquan Lu, et al. Ovarian transposition before radiotherapy in cervical cancer patients: functional outcome and the adequate dose constraint
4 Michael JZ, Zvi Fuks, Margie Hunt, et al.: High-dose intensity modulated radiation therapy for prostate cancer: early toxicity and biochemical outcome in 772 patients. Int J Radiat Oncol Biol Phys, 2002;53(5):1111-6   DOI
5 Choi So Young,Kim Tae Won,Kim Min Su, et al. : Phantom Study = Comparative evaluation of dose according to changes in rectal gas volume during radiation therapy for cervical cancer : Phantom Study The journal of the korean society for radio-therapeutic technology 33(-), 89-97.
6 D. W. Kim, W. K. Chung, M. G. Yoon, "Imaging doses and secondary cancer risk from kilovoltage cone-beam CT in radiation therapy," Health physics, Vol. 104, No. 5, pp. 499-503, 2013.   DOI
7 Bin Cai 1, Eric Laugeman 1, Thomas R Mazur, et al : Characterization of a prototype rapid kilovoltage x-ray image guidance system designed for a ring shape radiation therapy unit 2019 Mar;46(3):1355-1370
8 E.G. Yukihara, S.W. McKeever. Optically stimulated luminescence (OSL) dosimetry in medicine. Phys Med Biol, 53 (2008), pp. R351-R379
9 Zhenia Gopalakrishnanm, Raghuram K.Nair, Saju Bhasi et al: Verification of treatment planning algorithms using optically stimulated luminescent dosimeters in a breast phantom. J med Phys. Vol 43(4) 2018:264-269   DOI
10 Su Chul Han, Kum Bae Kim, Sang Hyoun Choi et al.: Changes of Optically Stimulated Luminescence Dosimeter Sensitivity with High Dose. Vol 27(2) 2016:98
11 Jung Seong Ahn A study of Peripheral Area Image Quality in CBCT of Medical Linear Accelerator Department of Health Science, Graduate School, Eulji University
12 Sameer Taneja 1, David L Barbee 1, Anthony J Rea et al.: CBCT image quality QA: Establishing a quantitative program J Appl Clin Med Phys. 2020 Nov;21(11):215-225.