Evaluation of the Usefulness of Exactrac in Image-guided Radiation Therapy for Head and Neck Cancer  

Baek, Min Gyu (Department of Radiation Oncology, Changwon Gyengsang National University Hospital)
Kim, Min Woo (Department of Radiation Oncology, Changwon Gyengsang National University Hospital)
Ha, Se Min (Department of Radiation Oncology, Changwon Gyengsang National University Hospital)
Chae, Jong Pyo (Department of Radiation Oncology, Changwon Gyengsang National University Hospital)
Jo, Guang Sub (Department of Radiation Oncology, Changwon Gyengsang National University Hospital)
Lee, Sang Bong (Department of Radiation Oncology, Changwon Gyengsang National University Hospital)
Publication Information
The Journal of Korean Society for Radiation Therapy / v.32, no., 2020 , pp. 7-15 More about this Journal
Abstract
Purpose: In modern radiotherapy technology, several methods of image guided radiation therapy (IGRT) are used to deliver accurate doses to tumor target locations and normal organs, including CBCT (Cone Beam Computed Tomography) and other devices, ExacTrac System, other than CBCT equipped with linear accelerators. In previous studies comparing the two systems, positional errors were analysed rearwards using Offline-view or evaluated only with a Yaw rotation with the X, Y, and Z axes. In this study, when using CBCT and ExacTrac to perform 6 Degree of the Freedom(DoF) Online IGRT in a treatment center with two equipment, the difference between the set-up calibration values seen in each system, the time taken for patient set-up, and the radiation usefulness of the imaging device is evaluated. Materials and Methods: In order to evaluate the difference between mobile calibrations and exposure radiation dose, the glass dosimetry and Rando Phantom were used for 11 cancer patients with head circumference from March to October 2017 in order to assess the difference between mobile calibrations and the time taken from Set-up to shortly before IGRT. CBCT and ExacTrac System were used for IGRT of all patients. An average of 10 CBCT and ExacTrac images were obtained per patient during the total treatment period, and the difference in 6D Online Automation values between the two systems was calculated within the ROI setting. In this case, the area of interest designation in the image obtained from CBCT was fixed to the same anatomical structure as the image obtained through ExacTrac. The difference in positional values for the six axes (SI, AP, LR; Rotation group: Pitch, Roll, Rtn) between the two systems, the total time taken from patient set-up to just before IGRT, and exposure dose were measured and compared respectively with the RandoPhantom. Results: the set-up error in the phantom and patient was less than 1mm in the translation group and less than 1.5° in the rotation group, and the RMS values of all axes except the Rtn value were less than 1mm and 1°. The time taken to correct the set-up error in each system was an average of 256±47.6sec for IGRT using CBCT and 84±3.5sec for ExacTrac, respectively. Radiation exposure dose by IGRT per treatment was measured at 37 times higher than ExacTrac in CBCT and ExacTrac at 2.468mGy and 0.066mGy at Oral Mucosa among the 7 measurement locations in the head and neck area. Conclusion: Through 6D online automatic positioning between the CBCT and ExacTrac systems, the set-up error was found to be less than 1mm, 1.02°, including the patient's movement (random error), as well as the systematic error of the two systems. This error range is considered to be reasonable when considering that the PTV Margin is 3mm during the head and neck IMRT treatment in the present study. However, considering the changes in target and risk organs due to changes in patient weight during the treatment period, it is considered to be appropriately used in combination with CBCT.
Keywords
Image-guided Radiation Therapy; 6D Online Setup Correction; CBCT; ExacTrac;
Citations & Related Records
연도 인용수 순위
  • Reference
1 Quality assurance for image-guided radiation therapy utilizing CT-based technologies: A report of the AAPM TG-179
2 Lee N, et al.: Int J Radiat Oncol Biol Phys. 2003: Intensity-modulated radiation therapy for head-and-neck cancer: The UCSF experience focusing on target volume delineation: 2003 Sep 1;57(1):49-60.   DOI
3 Miyabe Y, et al.: Med Phys. 2011: Positioning accuracy of a new image-guided radiotherapy system: 2011 May;38(5):2535-41   DOI
4 Hansen EK, Larson DA, Aubin M, et al.: Image-guided radiotherapy using megavoltage cone-beam computed tomography for treatment of parasipinous tumors in the presence of orthopedic hardware: Int J Radiat Oncol Biol Phys 2006;66:323-6   DOI
5 Ryu SI, Chang SD, Kim DH, et al: Image-guided radiosurgery for the spine and pancreas: Comput Aided Surg 2000;5:278-88.   DOI
6 Murphy, M.J., Adler, J.R. Jr., Bodduluri, M. et al: Image-guided radiosurgery for the spine and pancreas: Comput Aided Surg. 2000;5:278-288.   DOI
7 Yin, F.F., Ryu, S., Ajlouni, M. et al: Image-guided procedures for intensity-modulated spinal radiosurgery: Technical note. J Neurosurg. 2004;101:419-424.   DOI
8 Yin, F.F., Wang, Z., Yoo, S. et al: Integration of conebeam CT in stereotactic body radiation therapy: Technol Cancer Res Treat. 2008;7:133-139.   DOI
9 van Herk, M.: Different styles of image-guided radiotherapy: Semin Radiat Oncol. 2007;17:258-267.   DOI
10 Guckenberger, M., Meyer, J., Wilbert, J., Baier, K., Sauer, O., Flentje, M.: Precision of image-guided radiotherapy (IGRT) in six degrees of freedom and limitations in clinical practice: Strahlenther Onkol. 2007;183:307-313.   DOI
11 Jin, J.Y., Yin, F.F., Tenn, S.E., Medin, P.M., Solberg, T.D.: Use of the BrainLAB ExacTrac X-ray 6D system in image-guided radiotherapy: Med Dosim. 2008;33:124-134.   DOI
12 Moseley, D.J., White, E.A., Wiltshire, K.L. et al: Comparison of localization performance with implanted fiducial markers and cone-beam computed tomography for on-line image-guided radiotherapy of the prostate: Int J Radiat Oncol Biol Phys. 2007;67:942-953.   DOI
13 Murphy, M.J., Balter, J., Balter, S. et al: The management of imaging dose during image-guided radiotherapy: report of the AAPM Task Group 75. Med Phys. 2007;34:4041-4063.   DOI
14 Kan, M.W., Leung, L.H., Wong, W., Lam, N.: Radiation dose from cone beam computed tomography for image-guided radiation therapy: Int J Radiat Oncol Biol Phys. 2008;70:272-279.   DOI
15 Clemente S, et al.: Is Exactrac X-ray System an alternative to CBCT for Positioning Patients Head and Neck Cancers?: Med Phys. 2013 Nov ; 40(11): 111725.   DOI
16 Chang Z, et al.: 6D image guidance for spinal noninvasive stereotactic body radiation therapy: Comparison between ExacTrac X-ray 6D with kilovoltage cone-beam CT: Radiother Oncol. 2010 Apr;95(1):116-21.   DOI
17 Se an Oh, et al.: Evaluations of the setup discrepancy between BrainLAB 6D ExacTrac and conebeam computed tomography used with the imaging guidance system Novalis-Tx for intracranial stereotactic radiosurgery: PLos ONE 12(5):e0177798.   DOI
18 Ma J, et al.: ExacTrac X-ray 6 degree-of-freedom image-guidance for intracranial non-invasive stereotactic radiotherapy: comparison with kilovoltage cone-beam CT: Radiother Oncol. 2009 Dec;93(3):602-8.   DOI