Comparison treatment planning with the measured change the dose of each Junction section according to the error of setup CSI Treatment with Conventional, IMRT, VMAT  

Lee, Ho Jin (Department of Radiation Oncology, Seoul University hospital)
Jeon, Chang Woo (Department of Radiation Oncology, Seoul University hospital)
Ahn, Bum Suk (Department of Radiation Oncology, Seoul University hospital)
Yu, Sook Hyeon (Department of Radiation Oncology, Seoul University hospital)
Park, So Yeon (Department of Radiation Oncology, Seoul University hospital)
Publication Information
The Journal of Korean Society for Radiation Therapy / v.26, no.2, 2014 , pp. 217-224 More about this Journal
Abstract
Purpose : Conventional, IMRT, at CSI treatment with VMAT, this study compare the treatment plan with dose changes measured at Junction field according to the error of Setup. Materials and Methods : This study established Conventional, the IMRT, VMAT treatment planning for CSI therapy using the Eclipse 10.0 (Eclipse10.0, Varian, USA) and chose person in Seoul National University Hospital. Verification plan was also created to apply IMRT QA phantom for each treatment plan to the film measurements. At this time, the error of Setup was applied to the 2, 4, 6mm respectively with the head and foot direction. ("+" direction of the head, "-" means that the foot direction.) Using IMRT QA Phantom and EBT2 film, was investigated by placing the error of Setup for each Junction. We check the consistency of the measured Film and plan dose distribution by gamma index (Gamma index, ${\gamma}$). In addition, we compared the error of Setup by the dose distribution, and analyzing the uniformity of the dose distribution within the target by calculating the Homogeneity Index (HI). Results : It was figured out that 90.49%-gamma index we obtained with film is agreement with film scan score and dose distribution of treatment plan. Also, depend on the dose distribution on distance, if we make the error of Setup 2, 4, 6mm in the head direction, it showed that 3.1, 4.5, 8.1 at $^*Diff$(%) of Conventional, 1.1, 3.5, 6.3 at IMRT, and 1.6, 2.5, 5.7 at VMAT. In the same way, if we make the error of Setup 2, 4, 6mm in the foot direction, it showed that -1.6, -2.8, -4.4 at $^*Diff$(%) of Conventional, -0.9, -1.6, -2.9 at IMRT, and -0.5, -2.2, -2.5 at VMAT. Homogeneity Index(HI)s are 1.216 at Conventional, 1.095 at IMRT and 1.069 at VMAT. Discussion and Conclusion : The dose-change depend on the error of Setup at the CSI RT(radiation therapy) using IMRT and VMAT which have advantages, Dose homogeneity and the gradual dose gradients on the Junction part is lower than that of Conventional CSI RT. This a little change of dose means that there is less danger on patients despite of the error of Setup generated at the CSI RT.
Keywords
Citations & Related Records
Times Cited By KSCI : 1  (Citation Analysis)
연도 인용수 순위
1 이충일 김회남 오택열 등 : 전산화 단층 촬영 장치를 이용 한 뇌척수 조사의 치료 계획. The journal of the korean radiotherapeutic technology, v.11 no.1, 1999년 : 53-59
2 Matthew T. Studenski, Ph.D.,* Xinglei Shen, M.D. et al : Intensity-modulated radiation therapy and volumetricmodulated arc therapy for adult craniospinal irradiation-A comparison with traditional techniques. Medical Dosimetry 38 (2013) : 48-54   DOI   ScienceOn
3 Tejinder Kataria, Kuldeep Sharma, Vikraman Subramani, K. P. Karrthick, and Shyam S. Bisht : Homogeneity Index: An objective tool for assessment of conformal radiation treatments. J Med Phys. 2012 Oct-Dec; 37(4): 207-213   DOI   ScienceOn
4 Myonggeun Yoon, Sung Yong Park, Dongho Shin, et al. : A new homogeneity index based on statistical analysis of the dose-volume histogram. J Appl Clin Med Phys. 2007 Mar 20 : 9-17
5 우헌, 박장필, 민제순, 이제희, 유숙현 : 용적세기조절회 전치료 치료계획 확인에 사용되는 MapPHAN의 유용성 평가. 대한방사선치료학회지 제 5권 제호 2013 : 115-121
6 Daniel A. Low, William B. Harms, Sasa Mutic, and James A. Purdy : A technique for the quantitative evaluation of dose distributions. Medical Physics, Vol. 25, No. 5, May 1998 : 656-661   DOI   ScienceOn
7 William A. Parker, Carolyn R. Freeman : A simple technique for craniospinal radiotherapy in the supine position. Radiotherapy and Oncology 78 (2006) : 217?222   DOI   ScienceOn
8 ROLAND B. HAWKINS, M.D., PH.D. : A simple method of radiation of radiation treatment of craniospinal fields with patient supine. Int. J. Radiation Oncology Biol. Phys., Vol. 49, No. 1, 2001 : 261-264   DOI   ScienceOn
9 Jianzhou Chen, Chuangzhen Chen, Todd F. Atwood, et al. : Volumetric modulated arc therapy planning method for supine craniospinal irradiation. Journal of Radiation Oncology September 2012, Volume 1, Issue 3 : 291-297   DOI