A study to 3D dose measurement and evaluation for Respiratory Motion in Lung Cancer Stereotactic Body Radiotherapy Treatment  

Choi, Byeong-Geol (Department of Radiation Oncology, VHS Medical Center)
Choi, Chang-Heon (Department of Radiation Oncology, VHS Medical Center)
Yun, Il-Gyu (Department of Radiation Oncology, VHS Medical Center)
Yang, Jin-Seong (Department of Radiation Oncology, VHS Medical Center)
Lee, Dong-Myeong (Department of Radiation Oncology, VHS Medical Center)
Park, Ju-Mi (Department of Radiation Oncology, VHS Medical Center)
Publication Information
The Journal of Korean Society for Radiation Therapy / v.26, no.1, 2014 , pp. 59-67 More about this Journal
Abstract
Purpose : This study aims to evaluate 3D dosimetric impact for MIP image and each phase image in stereotactic body radiotherapy (SBRT) for lung cancer using volumetric modulated arc therapy (VMAT). Materials and Methods : For each of 5 patients with non-small-cell pulmonary tumors, a respiration-correlated four-dimensional computed tomography (4DCT) study was performed. We obtain ten 3D CT images corresponding to phases of a breathing cycle. Treatment plans were generated using MIP CT image and each phases 3D CT. We performed the dose verification of the TPS with use of the Ion chamber and COMPASS. The dose distribution that were 3D reconstructed using MIP CT image compared with dose distribution on the corresponding phase of the 4D CT data. Results : Gamma evaluation was performed to evaluate the accuracy of dose delivery for MIP CT data and 4D CT data of 5 patients. The average percentage of points passing the gamma criteria of 2 mm/2% about 99%. The average Homogeneity Index difference between MIP and each 3D data of patient dose was 0.03~0.04. The average difference between PTV maximum dose was 3.30 cGy, The average different Spinal Coad dose was 3.30 cGy, The average of difference with $V_{20}$, $V_{10}$, $V_5$ of Lung was -0.04%~2.32%. The average Homogeneity Index difference between MIP and each phase 3d data of all patient was -0.03~0.03. The average PTV maximum dose difference was minimum for 10% phase and maximum for 70% phase. The average Spain cord maximum dose difference was minimum for 0% phase and maximum for 50% phase. The average difference of $V_{20}$, $V_{10}$, $V_5$ of Lung show bo certain trend. Conclusion : There is no tendency of dose difference between MIP with 3D CT data of each phase. But there are appreciable difference for specific phase. It is need to study about patient group which has similar tumor location and breathing motion. Then we compare with dose distribution for each phase 3D image data or MIP image data. we will determine appropriate image data for treatment plan.
Keywords
Stereotactic body radiotherapy(SBRT); MIP image; respiratory motion;
Citations & Related Records
연도 인용수 순위
  • Reference
1 Wunderink W, Mendez Romero A, de Kruijf W, et al.: Reduction of respiratory tumor motion by abdominal compression in stereotactic body frame, analyzed by tracking fiducial markers implanted in liver. Int J Radiat Oncol Biol Phys 2008;71:907-915   DOI   ScienceOn
2 Case RB, Sonke JJ, Moseley DJ, et al.: Inter- and intrafraction variability in liver position in non-breath hold stereotactic body radiotherapy. Int J Radiat Oncol Biol Phys 2009; 75:302-308   DOI   ScienceOn
3 Korreman S, Medin J, Kjar-Kristoffersen F: Dosimetric verification of RapidArc treatment delivery. Acta Oncol 2009;48:185-191   DOI   ScienceOn
4 Schreibmann E, Dhabaan A, Elder E, Fox T: Patientspecific quality assurance method for VAMT treatment delivery. Med Phys 2009;36:4530-4535   DOI   ScienceOn
5 Low DA, Harms WB, Mutic S, et al: A technique for the quantitative evaluation of dose distributions. Med Phys 1998;25:656-661   DOI   ScienceOn
6 Low DA, Dempsey JF: Evaluation of the gamma dose distribution comparison method. Med Phys 2003;30:2455-2464   DOI   ScienceOn
7 Depuydt T, Van Esch A, Huyskens DP, et al: A quantitative evaluation of IMRT dose distributions: refinement and clinical assessment of the gamma evaluation. Radiother Oncol 2002;62:309-319   DOI   ScienceOn
8 Matsugi K, Narita Y, Sawada A, et al.: Measurement of interfraction variation in position and size of target volumes in stereotactic body radiotherapy for lung cancer. Int J Radiat Oncol Biol Phys. 2009;75:543-548   DOI   ScienceOn
9 Han K, Cheung P, Basran PS, et al.: A comparison of two immobilization systems for stereotactic body radiation therapy of lung tumors. Radiother Oncol 2010;95:103-108   DOI   ScienceOn
10 Mendez Romero A, Zinkstok RT, Wunderink W, et al.: Stereotactic body radiation therapy for liver tumors: Impact of daily setup corrections and day-to-day anatomic variations on dose in target and organs at risk. Int J Radiat Oncol Biol Phys 2009;75:1201-1208   DOI   ScienceOn