Abstract
Purpose Cadmium-zinc-telluride (CZT) camera with semiconductor detector is capable of dynamic myocardial perfusion SPECT for coronary flow reserve (CFR). Image acquisition with the heart positioned within 2 cm in the center of the quality field of view (QFOV) is recommended because the CZT detector based on focused multi-pinhole collimators and is stationary gantry without rotation. The aim of this study was to investigate the optimal method for measuring position of the heart within the center of the QFOV when performing dynamic myocardial perfusion SPECT with the Discovery NM 530c camera. Materials and Methods From June to September 2018, 45 patients were subject to dynamic myocardial perfusion SPECT with D530c. For accurate heart positioning, the patient's heart was scanned with a mobile ultrasound and marked at the top of the probe where the mitral valve (MV) was visible in the parasternal long-axis view (PLAX). And, the marked point on the patient's body matched with the reference point indicated CZT detector in dynamic stress. The heart was positioned to be in the center of the QFOV in rest. The coordinates of dynamic stress and rest were compared statistically. Results The coordinates of the dynamic stress using mobile ultrasound and those taken of the rest were recorded for comparative analysis with regard to the position of the couch and analyzed. There were no statistically significant differences in the coordinates of Table in & out, Table up & down, and Detector in & out (P > 0.05). The difference in distance between the 2 groups was measured at $0.25{\pm}1.00$, $0.24{\pm}0.96$ and $0.25{\pm}0.82cm$ respectively, with no difference greater than 2 cm in all categories. Conclusion The position of the heart taken using mobile ultrasound did not differ significantly from that of the center of the QFOV. Therefore, The use of mobile ultrasound in dynamic stress will help to select the correct position of the heart, which will be effective in clinical diagnosis by minimizing the image quality improvement and the patient's exposure to radiation.
Keywords
Coronary flow reserve; CZT camera; QFOV; D530c;
Citations & Related Records
연도 인용수 순위
  • Reference
1 Gloud KL, Kirkeeide RL, Buchi M. Coronary flow reserve as a physiologic measure of stenosis severity. J Am Coll Cardiol 1990;15:459-74.   DOI
2 이병일, 범희승. 양전자방출단층촬영을 이용한 심근혈류 및 관상동맥 혈류예비능 평가. 대한핵의학회지 2005; 39:118-123
3 Yasuyuki Takahashi, Masao Miyagawa, Yoshiko Nishiko Nishiyama, Hayato Ishimura, Teruhito Mochizuki. Performance of a semiconductor SPECT system: comparison with a conventional Anger-type SPECT instrument. Ann Nucl Med 2013;27:11-16   DOI
4 Masao Miyagawa, Yoshiko Nishiyama, Rami Tashiro, Hayato Ishimura, Yasuyuki Takahashi, Teruhito Mochizuki. Novel Cardiac SPECT Technology with Semiconductor Detectors: Emerging Trends and Future Perspective. Ann Nucl Med 2015;1:18-26
5 한국심초음파학회. 임상 심초음파학. 제2판. 대한의학서적; 2008. p. 81-82
6 Standard scanning method : Transthoracic Echocardiography. Japanese Society of Sonographers Web site. Available at: http: http://www.jss.org/english/standard/standardization_ ca/index.html. Accessed March 19, 2019.
7 Derliz Mereles. Transthoracic examination. Echobasics Web site. Available at: http://www.echobasics.de/tteen.html. Accessed April 7, 2019.