Abstract
Electro interstitial scan shows potential as a non-invasive screening method. It can discriminate some diseases based on electric current response to induce low intensity direct current to limbs or local area of body. DDFAO was invented in France and it is claimed that multi-channel EIS(Electro Interstitial Scan) is useful for various diseases, especially, diagnoses of endocrine system such as diabetics are very effective. In this study, we verified the repeatability and sensitivity of DDFAO by using a RC phantom model and its clinical usefulness using data obtained from normal and diabetes subject groups. As a result, it showed the repeatability and the output change according to change of phantom characteristic, but it was hard to distinguish normal and patient groups non-invasively with just six surface electrodes of DDFAO. The repeatability and the clinical accuracy was not sufficient for screening or diagnostic purposes, as well. In spite of the results with low repeatability and accuracy conducted in this study, we still need further investigations to improve the EIS-based measurement method; EIS is very convenient and simple and it shows potential as a screening tool of the whole body health conditions rather than localized disease diagnosis.
Keywords
Iontophoresis; Cottrell equation; Skin conductance; Bioimpedance; Diabetes;
Citations & Related Records
연도 인용수 순위
  • Reference
1 H.S. Kim, Radiological analysis of liver nodules under diagnostic debate in patients with liver cirrhosis, Korean Association for the Study of the Liver, Single Topic Symposium, pp. 14-29, 2010.
2 J.H. Lee, On the Current State of Blood borne Infectious Diseases Caused by Needle Stick Injuries, The Graduate School of Ewha Womans University, 2010.
3 B. Philippe and T. Lydie, Electrophysiological analysis system and method, PCT/EP2006/ 063452, 2007.
4 P. Brunswick and N. Bocquet, Electrophysiological analysis system, PCT/ EP2008/052211, 2014.
5 A.J. Bard, and L.R. Faulkner, Electrochemical Methods: Fundamentals and Applications 2nd Ed, Wiley, New York. 2001.
6 R. Bayford, Bioimpedance tomography (electrical impedance tomography, Annu Rev Biomed Eng, Vol. 8, pp. 63-91, 2006.   DOI
7 University of California San francisco, Top Trends in Health and Science for 2014, https://www.ucsf.edu/news/2013/12/110666/top-tre nds-health-sciences-2014. 2013.12.
8 EZSCAN & Impeto Medical, http://www.activeage.org/publications/doc_download/48-ezscan-case-study & http://www.impeto-medical.com, 2015.07.
9 M.S. Son, guide for medical screening assessment, Health Insurance Review & Assessment Service, 2014.
10 A. Maarek, Electro interstitial scan system: assessment of 10 years of research and development. Medical Devices: Evidence and Research, Vol. 5, pp. 23-30, 2012.
11 E.H. Schwarz, P. Brunswick, and J. Calvet, EZSCAN a new technology to detect diabetes risk, The British Journal of Diabetes & Vascular Disease, Vol. 11, no. 4, pp. 204-209, 2011.   DOI
12 C.M. Casellini, et al., Sudoscan a noninvasive tool for detecting diabetic small fiber neuropathy and autonomic dysfunction. Diabetes Technol Ther, Vol. 15, No. 11, pp. 948-53, 2013.   DOI
13 http://www.ldteck.com/sudopath.html.
14 http://www.screenyourbody.com/faq.shtml.
15 http://www.helse1.no/dokumenter/DDFAO_T50_Synthesis.pdf.
16 R. Ozaki, et al., A new tool to detect kidney disease in Chinese type 2 diabetes patients: comparison of EZSCAN with standard screening methods. Diabetes Technol Ther, Vol. 13, No. 9, pp. 937-43, 2011.   DOI
17 H.M. Dastjerdi, R. Soltanzadeh, and H. Rabbani, Designing and implementing bioimpedance spectroscopy device by measuring impedance in a mouse tissue. J Med Signals Sens, Vol. 3, No. 3, pp. 187-94, 2013.
18 Medical Centre for the Prevention of Diseases, Validation of the ddfao system/clinical tests definitions of algorithms by inverse problems, http://helse1.no/dokumenter/DDFAO_T50_Synthes is.pdf
19 http://www.highvoltageinfo.com/2008/04/humanbody-resistance-and-capacitance.html.
20 J.H. Bae, Reliability test of a bio-electrical impedance device with varying contact area of electrodes, 10th International Symposium on Electrochemical Impedance Analysis, Borovets, Bulgaria, June, 2014.
21 Guy Holt, http://www.screenlightandgrip.com/html/481_GFCI _Workshop.html. 2012.