A Study on Dose and Image Quality according to X-ray Photon Detection Method in Digital Radiography System  

Hong, Sun Suk (Department of Radiologg Asan Medical Center, Department of Radiologic Technology, Eulji University, Graduate School)
Kim, Ho Chul (Department of Radiological Science, Eulji University)
Publication Information
Journal of the Institute of Electronics and Information Engineers / v.50, no.12, 2013 , pp. 247-253 More about this Journal
Abstract
The purpose is a comparative evaluation in the DR System according to the dosimetry and image quality of the quantitative and objective via Direct digital radiography, Indirect digital radiography, Image intensifier (Charge Coupled Device type) digital radiography. The experimental method used rando phantom and measured the entrance surface dose. And through using the measured entrance surface dose and then using the PCXMC program were evaluated risk due to irradiation and the effective dose. SNR and NPS and CNR were measured and analyzed by using 21cm acryl phantom. Significance of measured value was evaluated by statistics method. Entrance surface dose, major organ dose, effective dose all of them were measured the lowest rated in direct DR when it is on the basis of direct DR dose, high-dose ratio were measured in I.I DR approximately 1.3 times, indirect DR approximately 2.4 times. Risk in accordance with radiation also was measured same as dose ratio. On the conclusion that SNR measurement result based on direct DR SNR measurements, low-SNR ratio were measured in I.I DR approximately 7.25 times, indirect DR approximately 1.48 times. On the conclusion that CNR measurement result based on direct DR CNR measurements, high-dose ratio were measured in I.I type DR approximately 1.16 tims and low-dose ratio were measured in indirect DR approximately 0.87 times. Therefore Direct DR system using a-selenium sensing element to detect x-ray photon is thought effectively at the examination such as infant to sensitive irradiation and the genital gland. Because quality image is built by low dose. Also when it is necessary that image test requiring many diagnosis information, indirect DR system is thought effectively.
Keywords
direct digital radiography; indirect digital radiography; Image intensifier digital radiography; image quality; dose; image receptor;
Citations & Related Records
연도 인용수 순위
  • Reference
1 M. Yaffe and J. Rowlands, "X-ray detectors for digital radiography". Physics in Medicine and Biology., 42, pp. 1, 1997.   DOI   ScienceOn
2 M. Korner, C. H. Weber, S. Wirth, K.-J. Pfeifer, M. F. Reiser, and M. Treitl, "Advances in Digital Radiography: Physical Principles and System Overview1"., Radiographics., vol.27, pp.675-686, 2007.   DOI   ScienceOn
3 Tae-wook. Kang, "Performance evaluation of indirect digital radiography system Performance evaluation of indirect digital radiography system"., Graduate school Korea university of bio-medical sciences, Seoul, Master's Thesis, 2011.
4 E. Alcon and R. Lopes, "Slot scintillation detector modeling for digital radiography"., Radiation Physics and Chemistry, vol. 61, pp. 411-414, 2001.   DOI   ScienceOn
5 D. Williams and P. Burns, "Diagnostics for digital capture using MTF"., in IS AND TS PICS CONFERENCE, pp. 227-232, 2007
6 Jung-min Kim, "Compare the quality and characteristics of a digital fluoroscopy unit"., Journal of Health Science & Medical Technology, vol. 28, pp.92-102, 2002.
7 A. G. Davis, A. R. Cowen, S. M. Kengyelics, J. Moore, and M. U. Sivananthan, "Do flat detector cardiac X-ray system convey advantages over image-intersifire-based system study comparing X-ray dose and image quality., European Radiology, vol.17, pp.1787-1794, 2007.   DOI   ScienceOn
8 D. Thomas, S. Darby, F. Fagnani, P. Hubert, M. Vaeth, and K. Weiss, "Definition and estimation of lifetime detriment from radiation exposures: principles and methods"., Health physics, vol.63, pp.259-272, 1992.   DOI   ScienceOn
9 M. Tapiovaara and T. Siiskonen, "A Monte Carlo program for calculating patient doses in medical x-ray examinations", Radiation and Nuclear Safety Authority, Report ISNN 978-952-478-397-2 STUK-A231, Finland, 2008.
10 K. Bacher, P. Smeets, L. Vereecken, A. De Hauwere, P. Duyck, R. De Man, et al., "Image quality and radiation dose on digital chest imaging: comparison of amorphous silicon and amorphous selenium flat-panel systems"., American Journal of Roentgenology, vol.187, pp.630-637, 2006.   DOI   ScienceOn