Browse > Article

A Study on Blood Flow Measurement Method using Independent Component Analysis  

Cho, Seog-Bin (Dept. of Electronics Engineering, Pusan National University)
Lim, Dong-Seok (LG Electronic Inc.)
Baek, Kwang-Ryul (Dept. of Electronics Engineering, Pusan National University)
Publication Information
Abstract
The echo signal on ultrasonic transducer is a mixed signal from tissues, blood vessel walls, blood cells and noise. In this mixed-signal, the signal reflected from tissues and blood vessel walls is called clutter. It is necessary to extract pure blood signal from this mixed-signal, when measuring blood flow velocity with medical ultrasonic system The quality of measured blood flow velocity is highly dependent on sufficient attenuation of the clutter signals. In this paper, we suggest a clutter rejection method using ICA For simulation, the echo signals are generated by Field n ultrasonic simulation program In this echo signals, independent signals are separated by using ICA Then the blood signal is obtained from the separated signals. Blood flow velocity is measured by 2D autocorrelation method. We compare ICA clutter rejection method with PCA-based eigen filter method using both measured blood flow velocity profiles by 2D autocorrelation. In simulation results, ICA clutter rejection method can be better applied measuring blood flow velocity in noisy echo signals.
Keywords
Medical ultrasound; ICA; Blood velocity measurement; Clutter rejection;
Citations & Related Records
연도 인용수 순위
  • Reference
1 J.A. Jensen, Field II, a program for simulating ultrasound system, http://www.es.oersted.dtu.dk/staff/jaj/field/
2 T. Loupas, J.T. Powers and R.W. Gill, 'An axial velocity estimator for ultrasound blood flow imaging, based on a full evaluation of the doppler equation by means of a two-dimensional autocorrelation approach', IEEE trans. Ultrason., Ferroelect., Freq., Contr., vol. 42, no. 4, pp. 672-688, 1995   DOI   ScienceOn
3 W.D. Hedrick, D.L. Hykes, D.E. Starchman, Ultrasound physics and instrumentation, Elsevier Mosby, 2005
4 C.M. Gallippi, K.R. Nightingale and G.E. Trahey, 'BSS-based filtering of physiological and ARFI-induced tissue and blood motion', Ultrasound in Med. & Biol., vol. 29, no. 11, pp. 1583-1592, 2003   DOI   ScienceOn
5 김영길, '초음파 펄스 도플러 시스템의 stationary canceler에 대한 연구', 의공학회지, 6권, 1호, 1985
6 A. Hyvarinen, J. Karhunen and E. Oja, Independent component analysis, Wiley interscience, 2001
7 S. Bjaerum, H. Torp and K. Kristoffersen, 'Clutter filters design for ultrasound color flow imaging', IEEE trans, Ultrason., Ferroelect., Freq., Contr., vol. 49, no. 2, 2002
8 C.M. Gallippi and G.E. Trahey, 'Adaptive clutter filtering via blind signal separation for lateral blood velocity measurement', IEEE Ultrasonics Symp., pp.1517-1521, 2002
9 S. Bjaerum and H. Torp, 'Clutter filters adapted to tissue motion in ultrasound color flow imaging', IEEE trans, Ultrason., Ferroelect., Freq., Contr., vol. 49, no. 6, pp. 693-704, 2002   DOI   ScienceOn
10 S. Bjaerum and H. Torp, 'Optimal adaptive clutter filtering in color flow imaging', in Proc. IEEE Ultrasonic Symp., vol. 2, pp. 1223-1226, 1997
11 C.M. Gallippi and G.E. Trahey, 'Complex blind source separation for acoustic radiation force impulse imaging in the peripheral vasculature, in vivo', IEEE Ultrasonics Symp., pp.596-601, 2004