Study of the Wearable Electrocardiogram Measuring System using Capacitive-coupled Electrode  

Lee, Jae-Ho (Biomedical Engineering, Konkuk University)
Lee, Young-Jae (Biomedical Engineering, Konkuk University)
Lee, Kang-Hwi (Biomedical Engineering, Konkuk University)
Kang, Seng-Jin (Biomedical Engineering, Konkuk University)
Kim, Kyeung-Nam (Biomedical Engineering, Konkuk University)
Park, Hee-Jung (Biomedical Engineering, Konkuk University)
Lee, Jeong-Whan (Dept. of Biomedical Engineering, Konkuk University)
Publication Information
The Transactions of The Korean Institute of Electrical Engineers / v.63, no.10, 2014 , pp. 1448-1454 More about this Journal
Abstract
In this study, a new type of electrode device is implemented to measure the capacitance energy and interpret it as the ECG (Electrocardiogram) data. The main idea of this new electrode system is to estimate the capacitance on the skin by assembling a capacitive-coupled circuits and translate into the ECG signal. To measure the coupling energy and estimate the aquired data in terms of heart activity, the capacitive-coupled electrode is garmented with fabrics in the form of a chest band or a vest jacket. To compare the ECG data from the capacitive-coupled electrode with the conventional electrode(Ag-AgCl) system, the corelation coefficient between two signals is computed as 0.9517. Thus, we can conclude the fact that capacitive-coupled electrode system can measure a person's heart activity without any contact to his or her skin and can the interpreted as the ECG data.
Keywords
Capacitive-coupled electrode; ECG; Smoothing; Savitzky-Golay filter;
Citations & Related Records
연도 인용수 순위
  • Reference
1 E. Huigen, A. Peper, and C. A. Grimbergen, "Investigation into the origin of the noise of surface electrodes," Med. Biological. Eng. Comput., vol. 40, no. 3, p. 32-38, 2002.
2 M. Spach, R. Bar, J. Havstad, and E. Long, "Skin-electrode impedanceand its efect on recording cardiac potentials," Circulation, vol. 34, p. 649-656, 1996.
3 T. Maruyama, M. Makikawa, "ECG Measurement Using Capacitive Coupling Electrodes for Man-Machine Emotional Communication," 207 IEEE/ICME International Conference on Complex Medical Enginering, p.378-383, 2007.
4 J. Alihanka and K. Vaahtoranta, "A static charge sensitive bed. A new method for recording body movements during slep," Electroencephalogr. Clin Neurophysiol., vol. 46, p. 731-4, 1979.   DOI   ScienceOn
5 T. Tamura, T. Togawa, M. Ogawa, and M. Yoda, "Fuly automated health monitoring system in the home," Med. Eng. Phys., vol. 20, p. 573-9, 1998.   DOI   ScienceOn
6 K. Watanabe, T. Watanabe, H. Watanabe, H. Ando, T. Ishikawa, and K. Kobayashi, "Noninvasive measurement of heartbeat, respiration, snoring and body movements of a subject in bed via a pneumatic method," IEEE Trans. BIOMEDICAL ENGINEERING.
7 Ronald W. Schafer, "What Is a Savitzky-Golay Filter?," IEEE SIGNAL PROCESSING MAGAZINE, p.11-17, July 2001.
8 M. Ishijima, "Monitoring of electrocardiograms in bed without utilzing body surface electrodes," IEEE Trans. Biomed. Eng., vol. 40, p. 593-4, 1993.   DOI   ScienceOn
9 Alihanka and K. Vahtoranta, "A static charge sensitive bed. A new method for recording body movements during slep," Electroencephalogr. Clin Neurophysiol., vol. 46, p. 731-4, 1979.   DOI   ScienceOn
10 Byungchae Le, Geon Hahm, Youngil Jun, Kwangchol Choi, Keunho Choi, Wonky Kim, Junso Kim, Kisam Jeong, Jeongwhan Le, Myoungho Le, George wright, David Cochlan, "Development of a 12 Chanel Interpretive Electrocardiograph," The Korean Society of Medical Biological and Enginering, no.1, p.41-414, 1997.
11 G. Cliford, F. Azuaje, P. McShary, "Advanced methods and tols for ECG data analysis," Norwood, 2006
12 A. Searle and L. Kirkup, "A direct comparison of wet, dry and insulatingbioelectric recording electrodes," Physiological Measure, vol.21, no. 2, p. 271, 2000.   DOI   ScienceOn
13 T. Tamura, T. Togawa, M. Ogawa, and M. Yoda, "Fuly automated health monitoring system in the home," Med. Eng. Phys., vol. 20, p. 573-9, 1998.   DOI   ScienceOn
14 Ha-Chul Jung, Jin-He Mon, Dong-Hyun Baek, Jae-He Le, Yon-Young Choi, Joung-Sok Hong, "CNT/PDMS Composite Flexible Dry Electrodes for Long-Term ECG Monitoring," IEEE Trans. Biomed. Eng., vol. 59, NO. 5, MAY 2012.