Browse > Article
http://dx.doi.org/10.5370/KIEE.2012.61.6.885

Design of Noninvasive Arterial Blood Pressure Measurement System by Using Double-Cuffs Oscillometric Method  

Lee, Pil-Jae (건국대 의학공학부)
Lee, Young-Jae (건국대 의학공학부)
Kim, Dong-Jun (청주대 공대 전자정보공학부)
Lee, Jae-Ho (건국대 의학공학부)
Lim, Min-Gyu (건국대 의학공학부)
Lee, Jeong-Whan (건국대 의료생명대 의학공학부, 건국대 의공학실용기술연구소)
Publication Information
The Transactions of The Korean Institute of Electrical Engineers / v.61, no.6, 2012 , pp. 885-890 More about this Journal
Abstract
In this study, a new blood pressure measuring system was proposed and implemented. An additional small-cuff was placed on the center of a inner cuff to measure morphological signals and new oscillometric ratio. The proposed BP-measuring system is composed of an external cuff, an inner cuff and a small-cuff. Oscillation signal from small-cuff is interpolated with 7th-order fitting polynomials and SBP, DBP ratio were 22.2% and 87.7%. Experimental data were gathered from 20 volunteers ($25{\pm}4$ years) and arterial blood pressure values were compared with auscultation, sphygmomanometers, small-cuff and inner-cuff. As a result, the difference in systolic BP between auscultation and the small-cuff was 1.93(${\pm}1.28$) mmHg, and the inner-cuff was 4.53(${\pm}4.39$) mmHg, and sphygmomanometer was 6.68(${\pm}3.99$) mmHg, and the corresponding difference in diastolic BP was 2.50(${\pm}2.04$) mmHg, 3.50(${\pm}3.19$) mmHg, 7.35(${\pm}5.62$), respectively.
Keywords
Systolic BP; Diastolic BP; Mean arterial pressure; Small cuff; Oscillometric method;
Citations & Related Records
연도 인용수 순위
  • Reference
1 L. Shing-Hong,L. Chin-Teng, "A Model-Based Fuzzy Logic Controller With Kalman Filtering for Tracking Mean Arterial Pressure," IEEE Transactions on Systems, Vol. 31, No. 6, pp. 676-686, 2001.
2 L. Yuansheng, "New Blood Pressure Measure Method Based on Characteristic Point," Electronic Measurement and Instruments, ICEMI '07. 8th International Conference on, pp. 327-330, Aug. 2007.
3 J. W. Yeol, "Digital Signal Processing of the Wideband External Pulse Recorded During Cuff Deflation: a New Way to Measure Blood Pressure," Engineering in Medicine and Biology Society, pp. 5985-5988, 2006.
4 D. K. Park, "Novel method of automatic auscultation for blood pressure measurement using pulses in cuff pressure and korotkoff sound," Computers in Cardiology, pp. 181-184, 2008.
5 G. L. Peters, S. K. Binder, N. R. Campbell, "The effect of crossing legs on-60-blood pressure: a randomized single-blind cross-overstudy," Blood Press Monit, pp. 97-101, April, 1999.
6 J. M. Bland, D. G. Altman, "Statistical method for assessing agreement between two methods of clinical measurement," The Lancet, Vol. 327, pp. 307-310, 1986.   DOI   ScienceOn
7 T. Tamura, I. Mizukura, M. Sekine, Y. Kimura, "Monitoring and Evaluation of Blood Pressure Changes With a Home Healthcare System," Information Technology in Biomedicine, IEEE Transactions on, Vol. 15, pp. 602-607, 2011.
8 L. Wun-Jin, L. Yuan-Long, C. Yao-Shun, L. Yuan-Hsiang, "A wireless blood pressure monitoring system for personal health management," Engineering in Medicine and Biology Society, pp. 2196-2199, 2010.
9 L. A. Geddes, M. Voelz, C. combs, D. Reiner, and C. F. Babbs, "Characterization of the Oscillometric Method for Measuring Indirect Blood Pressure," Ann. Biomed. Eng, Vol. 10, pp. 271-280, 1982.   DOI   ScienceOn
10 J.C.T.B Moreas, M. Cerulli, P. S. Ng, "Development of a New Oscillometric Blood Pressure System," IEEE Computers in Cardiology Conf, pp. 467-470, 1999.