Improvement of the Accuracy of Wrist Noninvasive Blood Pressure Measurement Using Multiple Bio-signals  

Jung, Woon-Mo ((주)누가의료기 기술연구소)
Sim, Myeong-Heon (연세대학교 의공학과 대학원)
Jung, Sang-O ((주)누가의료기 기술연구소)
Kim, Min-Yong (연세대학교 의공학과 대학원)
Yoon, Chan-Sol (연세대학교 의공학과 대학원)
Jung, In-Chol ((주)누가의료기 기술연구소)
Yoon, Hyung-Ro (연세대학교 의공학과)
Publication Information
The Transactions of The Korean Institute of Electrical Engineers / v.60, no.8, 2011 , pp. 1606-1616 More about this Journal
Abstract
The blood pressure measuring equipment, which is being supplied and used most widely by being recognized convenience and accuracy now generally, is oscillometric blood pressure monitor. However, a change in blood pressure is basically influenced by diverse elements such as each individual's physiological status and physical condition. Thus, the measurement of blood pressure, which used single element called oscillation in blood pressure of being conveyed to cuff, is not considered on physiological elements such as cardiovascular system status and blood vessel stiffness index, and on external elements, thereby being quite in error. Accordingly, this study detected diverse bio-signals and body informations in each individual as the measurement subject such as ECG, PPG, and Korotkoff Sound in order to enhance convenience and accuracy of measuring blood pressure in the complex measurement equipment, thereby having extracted regression method for compensation in error of oscillometric blood pressure measurement on the wrist, and having improved accuracy of measuring blood pressure. To verify a method of improving accuracy, the blood pressure value in each of SBP, DBP, MAP was acquired through 4-stage experimental procedure targeting totally 51 subjects. Prior to experiment, the subjects were divided into two groups such as the experimental group for extracting regression method and the control group for verifying regression method. Its error was analyzed by comparing the reference blood pressure value, which was obtained through the auscultatory method, and the oscillometric blood pressure value on the wrist. To reduce the detected error, the blood pressure compensation regression method was calculated through multiple linear regression analysis on elements of blood pressure, individual body information, PTT, HR, K-Sound PSD change. Verification was carried out on improving significance and accuracy by applying the regression method to the data of control group. In the experimental results, as a result of confirming error on the reference blood pressure value in SBP, DBP, and MAP, which were acquired through applying regression method, the results of $-0.47{\pm}7.45$ mmHg, $-0.23{\pm}7.13$ mmHg, $0.06{\pm}6.39$ mmHg could be obtained. This is not only the numerical value of satisfying the sphygmomanometer reference of AAMI, but also shows the lower result than the numerical value in SBP : $-2.5{\pm}12.2$ mmHg, DBP : $-7.5{\pm}8.4$ mmHg, which is the mean error in the experimental results of Brram's research for verifying accuracy of Omron RX-M, which shows relatively high accuracy among wrist sphygmomanometers. Thus, the blood pressure compensation could be confirmed to be made within significant level.
Keywords
Wrist NiBP; Auscultatory method; Multi parameter; BP Compensation; Vascular status;
Citations & Related Records
Times Cited By KSCI : 2  (Citation Analysis)
Times Cited By SCOPUS : 0
연도 인용수 순위
1 T.K Kim, A.R. Cho, B.C Choi, S.W. Beak, "The Changes of Differences of Pulse Transit Time between Toe and Finger in General and Epidural Anesthesia" Korean J Anesthesiology, vol.53, no.1, pp. 21-26, 2007.   DOI   ScienceOn
2 Parry Fung, Guy Dumont, Craig Ries, Chris Mott, Mark Ansermino, "Continuous Noninvasive Blood Pressure Measurement by Pulse Transit Time", Proceedings of the 26th Annual International Conference of the IEEE EMBS San Francisco, CA, USA, pp. 1-5, 2004. 9.
3 S.W Kim, J.H Lee, K.C Nam, S.C. Kim, E.J Cha, D.W. Kim, "Comparison of peripheral vascular compliance between normal and diabetic groups using the second derivative of photoplethysmogram." The Institute of Electronics Engineers of Korea, vol. 44, pp. 15-20, 2007.
4 Ministry for Health and Welfare of Korea, "The survey of health and nutrition for nations", Korea Centers for Disease Control and Prevention, 2007.
5 Jason WD, Paul AM, James DS and Charles LS, "Effects of caffeine on blood pressure, heart rate, and forearm blood flow during dynamic leg exercise." Journal of Applied Physiology, Vol. 85, pp. 154-159, 1998.
6 Ali Reza Khoshdel, Shane Carney, Alastair Gillies, "The impact of arm position and pulse pressure on the validation of a wrist-cuff blood pressure measurement device in a high risk population", International Jounal of General Medicine, Vol. 3, pp. 119-125, 2010.
7 Kamal Rahmouni, Marcelo L. G. Correia, William G. Haynes, Allyn L. Mark, "Obesity-associated Hypertension : New Insights Into Mechanisms, Hypertension", Hypertension, Vol. 45, pp. 9-14, 2005.   DOI
8 Geddes, L.A., M.L., Voelz, M., Combs, C., Reiner, D., "Characterization of the oscillometric methods for measuring indirect blood pressure", Ann. Biomed. Eng., Vol. 10, pp. 271-280, 1983 .
9 Association for the Advancement of Medical Instrumentation, American national standard. "Electronic or automated sphygmomanometers, ANSI/AAMI SP 10-1987", Arlington, VA: AAMI, 1987.
10 Chobanian AV, Bakris GL, Black HR, Cushman WC, Green LA, Izzo JL, Jr, Jones DW, Materson BJ, Oparil S, Wright JT, Jr, Roccella EJ, "Seventh Report of The Joint National Committee on Prevention, Detection, and Treatment of High Blood Pressure: the JNC 7 Complete report", Joint National Committee on Prevention, Detection, Evaluation, and Treatment of High Blood Pressure; National Heart, Lung, and Blood Institute; National High Blood Pressure Education Program Coordinating Committee, 2004
11 Spiegelman D, Israel RG, Bouchard C, Willett WC, "Absolute fat mass percent body fat, and body fat distribution : Which is the real determinant of blood pressure and serum glucose?", American Journal of Clinical Nutrition, Vol. 55, pp. 1033-1044, 1992.   DOI
12 E. O'Brien, J. Petrie, W. Littler et al., "The British Hypertension Society protocol for the evaluation of automated and semi-automated blood pressure measuring device with special reference to ambulatory systems", J. Hypertension Vol. 8, pp. 607-619, 1990.   DOI   ScienceOn
13 V. Jazbinsek, J. Luznik, Z. Trontelj, "Non-Invasive Blood Pressure Measurements : Separation of The Arterial Pressure Oscillometric Waveform From the Deflation using Digital Filtering", EMBEC'05, 2005. 11.
14 H. C. Lukaski, "Comparison of proximal and distal placement of electrodes to assess human body composition by bioelectrical impedance", Basic Life Sci. Vol. 26, pp. 39-52, 1993
15 Webster JG, "Design of Pulse Oximeters", Institute of Physics Publishing Bristol and Philadelphia, 1997
16 Ursino, M., Cristalli, C., "Mathematical modeling of noninvasive blood pressure estimation techniques - part II : Brachial hemodynamincs", Journal of Boimechanical Engineering, Vol. 117, pp. 117-125, 1995.   DOI   ScienceOn
17 Ursino, M., Cristalli, C., "A mathematical study of some biomechanical factors affecting the oscillometric blood pressure measurement", IEEE Transactions on Biomedical Engineering, Vol. 43, pp. 761-778, 1996.   DOI   ScienceOn
18 Allen J, "Photoplethysmography and its application in clinical physiological measurement." Physiological Measurement, vol. 28, pp. R1-R39, 2007.   DOI   ScienceOn