Development of Joint Angle Measurement System for the Feedback Control in FES Locomotion  

Moon, Ki-Wook (건국대 의료생명대 의학공학부)
Kim, Chul-Seung (고려대학교)
Kim, Ji-Won (건국대 의료생명대 의학공학부)
Lee, Jea-Ho (건국대 의료생명대 의학공학부)
Kwon, Yu-Ri (건국대 의료생명대 의학공학부)
Kang, Dong-Won (건국대 의료생명대 의학공학부)
Khang, Gon (경희대 전자정보대 동서의료공학과)
Kim, Yo-Han (건국대학교 충주병원)
Eom, Gwang-Moon (건국대 의료생명대 의학공학부, 건국대 의공학실용기술연구소)
Publication Information
The Transactions of The Korean Institute of Electrical Engineers / v.58, no.1, 2009 , pp. 203-209 More about this Journal
Abstract
The purpose of this study is to develop a minimally constraint joint angle measurement system for the feedback control of FES (functional electrical stimulation) locomotion. Feedback control is desirable for the efficient FES locomotion, however, the simple on-off control schemes are mainly used in clinic because the currently available angle measurement systems are heavily constraint or cosmetically poor. We designed a new angle measurement system consisting of a magnet and magnetic sensors located below and above the ankle joint, respectively, in the rear side of ipsilateral leg. Two magnetic sensors are arranged so that the sensing axes are perpendicular each other. Multiple positions of sensors attachment on the shank part of the ankle joint model and also human ankle joint were selected and the accuracy of the measured angle at each position was investigated. The reference ankle joint angle was measured by potentiometer and motion capture system. The ankle joint angle was determined from the fitting curve of the reference angle and magnetic flux density relationship. The errors of the measured angle were calculated at each sensor position for the ankle range of motion (ROM) $-20{\sim}15$ degrees (dorsiflexion as positive) which covers the ankle ROM of both stroke patients and normal subjects during locomotion. The error was the smallest with the sensor at the position 1 which was the nearest position to the ankle joint. In case of human experiment, the RMS (root mean square) errors were $0.51{\pm}1.78(0.31{\sim}0.64)$ degrees and the maximum errors were $1.19{\pm}0.46(0.68{\sim}1.58)$ degrees. The proposed system is less constraint and cosmetically better than the existing angle measurement system because the wires are not needed.
Keywords
FES; Locomotion; Foot Drop; Hall Sensor; Magnet; Joint Angle; Gait Events; Magnetic Flux Density;
Citations & Related Records
Times Cited By KSCI : 1  (Citation Analysis)
Times Cited By SCOPUS : 0
연도 인용수 순위
1 Kovetic, R, Marsolais, E. B., 'Synthesis of Paraplegic gait with Multichannel Functional Neuromuscular Stimulation,' IEEE tans. Rehab. Eng., Vol. 2, pp. 66-79, 1994   DOI   ScienceOn
2 D. Graupe and K. H. Kohn, 'Functional neuromuscular stimulator for short-distance ambulation by certain thoracic-level spinal-cord-injured paraplegics,' Surg, Neurol., Vol. 50, no. 3, pp.202-207, Sept. 1998   DOI   ScienceOn
3 Colborne GR, Olney SJ, 'Feedback of joint angle and EMG in gait of able-bodied subjects.' Arch Phys Med Rehabil. 1990 Jun;71(7):478-83   PUBMED
4 박래준, 최상준, 조미숙, 조정선, 이윤미, 조용호, 박소현, '족저근막염 환자의 미세전류발생 신발 착용이 족부 근피로 및 통증 완화에 미치는 영향,' 대한물리치료학회지, Vol. 18, No. 1, pp. 1-10, 2006   과학기술학회마을
5 Philip R. Troyk., Robert J. Jaeger., Michael Haklin., James Poyezdala., Thomas Bajzek., 'Design and implementation of an Implantable goniometer' IEEE Trans. Biomedical Eng, Vol. BME-33, No.2, 1986   DOI   ScienceOn
6 P. T. Kolen., J. P. Rhode., P. R. Francis., 'Absolute angle measurement using the earth-field-referenced hall effect sensors,' J Biomechanics, Vol. 26, No.3, pp.265-270, 1993   DOI   ScienceOn
7 Gerald F. Harris., Peter A. Smith., Richard M. Marks., 'Foot and ankle motion analysis,' Clinical treatment and technology, pp.131-143, 2006
8 Alojz, Kralj, Tadej, Bajd, Gon, Khang, 'Functional Electrical Stimulation,' Ryomoongak, 1995
9 G.M. Eom, T. Watanabe, N. Hoshimiya, G. Khang, 'Gradual potentiation of isometric muscle force during constant electrical stimulation.' Med. BioI. Eng. Comput., 2002, 40, 137-143   DOI   ScienceOn
10 R. Nienhuis., J. M. Siegel., 'Analysis of head movement and position using hall effect devices', Physiology & Behevior, vol. 45, pp.199-203, 1989   DOI   ScienceOn
11 Pappas, I. P. I., Keller, T., Mangold, S., Popovic M.R., Dietz, V., Morai M., 'A reliable gyroscope-based gait-phase detection sensor embedded in a shoe insole' IEEE Sensors Journal, Vol.4, No.2, pp. 268-274, 2004   DOI   ScienceOn
12 T. Nakamura, T. Kusuhara, and Y. Yamamoto, 'Impedance characteristics and data processing for analysis of human movement,' Proc. XII Int. Conf. Elect, Bio-Imp., pp.641644, Oslo, Norway, 2001