Browse > Article

Implementation of Wearable Heart Activity Monitoring System having Modified Bipolar Electrode and Correlation Analysis with Clinical Electrocardiograph(ECG)  

Lee, Kang-Hwi (건국대학교 의학공학부)
Lee, Jeong-Whan (건국대 의료생명대 의학공학부)
Lee, Young-Jae (건국대학교 의학공학부)
Kim, Kyeong-Seop (건국대학교 의학공학부)
Yang, Heui-Koung (청주대학교 전자정보공학부)
Shin, Kun-Su (삼성종합기술원)
Lee, Myoung-Ho (연세대 공대 전기전자공학과)
Publication Information
The Transactions of The Korean Institute of Electrical Engineers / v.57, no.6, 2008 , pp. 1102-1108 More about this Journal
Abstract
Wearable physiological signal monitoring systems are regarded as an important sensing unit platforms in ubiquitous/mobile healthcare application. In this paper, we suggested the modified bipolar electrodes implemented on the portable heart activity monitoring system, which minimized the distance of electrodes formed on a attachable pad. The proposed electrode configuration is useful in mobile measurement environments, but has a disadvantage of reduced amplitude of the heart action potential. In order to overcome the shortcoming of the suggested electrode configuration, we implemented the amplifying circuit to increase the signal-gain and decrease the artifacts. For evaluations, we analyzed the specificity of measured cardiography using the proposed electrodes through the comparing of heart activity monitoring system with standard clinical ECG(lead2) by pearson correlation coefficients. The result showed that the average correlation coefficient is $0.903{\pm}0.036,\;0.873{\pm}0.072$ at V3, V4 chest lead position, respectively. Thus, the modified bipolar electrode is quite suitable to monitor the electrical activity of the heart in the situation of the mobile environment, and could be considered having high similarity with standard clinical ECG.
Keywords
ECG; Electrical Activity of the Heart; Electrode; Bio-signal Amplifier; Patch type ECG; Wearable ECG;
Citations & Related Records

Times Cited By SCOPUS : 0
연도 인용수 순위
  • Reference
1 Russel K.Hobbie, "생명과학을 위한 의학물리", 청문각, 3rd, 213p-241p
2 Brusce B.Winter and John G Webster, "Dreven-Right Leg Circuit Design", IEEE, Trans.Biomed.Eng, Vol1, pp.62-66, 1983
3 M.P.Wattts and D.B.Shoat, "Trends in electrocardiograph design", Journal of the Institution of electronic and Radio Eng, Vol57, No.4, pp.140-150, 1987   DOI   ScienceOn
4 지경용 외 14인, "유비쿼터스 시대의 보건의료", u-Health 서비스와 솔루션, 35p-84p, 2005
5 Enrique Mario Spinelli and Ramon Palls-Areny, " AC-Coupled Front-End for Biopotential Measurements", IEEE, Trans. Biomed. Eng, Vol50, No3, pp.391-395, 2003   DOI   ScienceOn
6 P.Zipp and H.Ahrens, "A Model of Bioelectrode Motion Artefact and Reduction of Artefact By Amplifier Input Stage Design", J.Biomed Eng, Vol1, pp.273-276, 1979   DOI
7 George E.Bergey, Russell D.Squires and William C.Sipple, "Electrocardiogram Recording with Pasteless Electrodes", IEEE, Trans,Biomed.Eng, Vol1 BME-18, No.3, pp.206-211, 1971
8 Willis j. Tompkins, "Biomedical Digital Signal Processing", Prentice-Hall International, 1993
9 김경호 외 2인, "의용기기개론", 문운당, 85p-90p, 2007
10 John G. Webster, "Medical Instrumentation: Application and Design," Wiley, 1998
11 James C. Huhta and John G. Webster, "60-Hz Interference in Electrocardiography", IEEE, Trans. Biomed. Eng 20, Vol 2, pp.91-101,1973