Browse > Article

A Eliminating Method for Baseline Wander Using Ascending Slope Tracing waves in ECG  

Lee, Ki-Young (관동대학 정보통신공학)
Kim, Jung-Kook (명지대학 전자공학과)
Publication Information
The Transactions of the Korean Institute of Electrical Engineers D / v.55, no.11, 2006 , pp. 471-475 More about this Journal
Abstract
In this paper, we propose a method to eliminate the baseline wander for ECG based on waveform morphology analysis. This method uses the ascending slope tracing waves to approximate the baseline wander in ECG and subtracts these waves from the original ECG to eliminate the baseline wander. This ascending slope tracing waves was developed for efficient enhancement of slope inverting points and sudden slope changing points. This method has been applied to MIT/BIH database to verify its efficacy and validity in practical applications.
Keywords
ECG; Baseline wander; Ascending slope tracing waves; Efficient enhancement; MIT/BIH database;
Citations & Related Records
Times Cited By KSCI : 1  (Citation Analysis)
연도 인용수 순위
1 J. Lee, W. Jo, G. Lee.'Adaptive filtering for baseline wander noise of ECG using neural networks', IEOCE Trans. Inform. & Syst., vol.E87-D, no. 1, pp.261-266. 2004
2 Jane. R., Laguna, P., Thaker, N.V., Caminal, P.,'Adaptive baseline wander removal in the ECG: comparative analysis with cubic spline technique'. Computers in Cardiology, pp, 143-146. 1992   DOI
3 김정국, 정석훈, 권철기, 함광근, 김응주, 박희남, 김영훈,허웅,'Biphasic 자동형 제세동기 시스템 개발', 의공학회지, 제25권, 2호, pp.119-127, 2004   과학기술학회마을
4 J. A. Van Aleste, T. S. Schilder,'Removal of baseline wander and power-line interference from the ECG by an efficient Fill filter with a reduced number of taps', IEEE Trans. Biomed. Eng., vol. BME-32, no. 12. pp. 1052-1060, 1985   DOI   ScienceOn
5 American Heart Association on Electrocardiography, 'Recommendations for standarization on lead and specifications for instruments in ECG/VCG', Circulation. 52. pp. 11-254, 1975
6 B. Mozaffary, M. A. Tinati,'ECG baseline wander elimination using wavelet packets', Enformatika Trans. Eng., Comput. & Tech. V3, ISSN 1305-5313. pp. 22-24, 2004
7 J. A. Van Alste, W. Van Eck, O. E. Herrman, 'ECG baseline wander reduction using linear phase filters'. Comput, Biomed. Res., 1986, Vol. 19, pp : 417-427   DOI   ScienceOn
8 L. Sornmo,'Time varying digital filtering of ECG baseline wander', Med. & Biol. Eng. & Computing, pp. 503-508, 1993   DOI
9 S.H. Oguz, M.H. Asyali,'A morphology based algorithm for baseline wander elimination in ECG records', Proceedings of the 1992 International Conference Biomedical Engineering Days, 18-20 August 1992, Istanbul, Turkey, pp. 157-160, 1992   DOI
10 C. R. Meyer, H. N. Kelser,'Electrocardiogram baseline noise estimation and removal using cubic splines and state-space computation techniques', Computers and Biomedical Research, vol. 10, 00. 459-470, 1977   DOI   ScienceOn