Precision Measurement Technique of DC Voltage/Resistance Ratio  

Kim, Kyu-Tae (한국표준과학연구원)
Yu, Kwang-Min (한국표준과학연구원)
Publication Information
The Transactions of the Korean Institute of Electrical Engineers D / v.51, no.3, 2002 , pp. 96-99 More about this Journal
Abstract
We developed 1:1 resistance ratio measurement system for precision measurement of a series-resistor type dc voltage divider. By using active guard technique and exchange technique, any leakage effects, which are one of the most critical error sources, were successfully removed. The best measurement uncertainty with the developed ratio measurement system was estimated to be approximately $10^{-8}\;for\;10\;k{\Omega}\;and\;100\;k{\Omega}$.
Keywords
표준;정밀측정;전압;저항;직류비율;누설;
Citations & Related Records
연도 인용수 순위
  • Reference
1 ISO 'Guide to the expression of uncertainty in measurements,' 1991
2 국제단위계, 제7 개정판, KRISS-99-101-SP,1999
3 L. Bliek et al., 'Measurements of the quantized Hall resistance h/e2 with a reproducibility of 10-8 and its application for a novel determination of the ohm,' PTB-Mitt., vol.93, pp.19-20 and pp.44-45, 1988
4 T.J. Witt, D. Reymann and D. Avrons, 'An accurate $10-k{\Omega}$ resistqance measurement,' IEEE Trans. Instrum. Meas., vol.40, pp.271-273, Apr. 1991   DOI   ScienceOn
5 ISO, 'Guide to the expression of uncertainty in measurements,' 1991
6 K.-T. Kim, S.I. Park, Y.M. Cho and S.-W. Kwon, 'Common error sources in using a series-resistors type voltage divider below level of 0.01 ppm,' IEEE Trans. Instrum. Meas., vol.44, pp. 318-321, Apr. 1995