A Study on Usefulness Verification Technique of the Measurement System by the Difference Between Caculated and Experimental Values of Ratio Error/phase Annie Error in Current Transformer  

정재갑 (한국표준과학연구원)
권성원 (한국표준과학연구원)
이상화 (한국표준과학연구원)
박영태 (한국표준과학연구원)
Publication Information
The Transactions of the Korean Institute of Electrical Engineers C / v.53, no.4, 2004 , pp. 213-217 More about this Journal
Abstract
A current transformer(CT) used for the estabilishment of high current national standard, has generally very small ratio error and phase angle error. Both the errors of CT depend critically on the external burden used. When both the ratio and phase angle errors at two different burdens including zero burden are known, those at any other burdens are calculated theoretically. The theoretical values are well consistent with the experimental results within the $82{\times}10$-6, implying the measurement system of CT in KRISS is well maintained.
Keywords
Current transformer; Ratio error; Phase angle error; Burden; National standard;
Citations & Related Records
Times Cited By KSCI : 3  (Citation Analysis)
연도 인용수 순위
1 G. Jones, 'Current Transformer Measurement' NPL Special Publication, 1994
2 W. J. M. Moore and N. L. Kusters, 'Direct reading ratio-error sets for the calibration of current transformers,' IEEE Trans. Instrum. Meas., Vol. IM-19, No. 3, p161, 1970
3 정재갑 등, '고전압 비율 국가표준확립', KRISS 보고서, 2003
4 W. J. M. Moore, P. N. Miljanic, 'The current comparator', Peter Peregrinus Ltd., London, United Kingdom, 1988
5 'American National Standard Requirements for Instrument Transformers', American National Standards Institute, ANSI/IEEE C57.13,, 1986
6 John D. Ramboz and O. Petersons, 'A Calibration Service for Current Transformers' NIST Special Punlication, vol. 250, No. 36, p66, 1991
7 권성원 등, '교류전압 전류의 정밀측정과 계기용 변성기', KRISS-99-091-ET, 1999