Abstract
A novel dual beam heterodyne Laser Doppler Vibrometer (LDV) in conjunction with FM demodulators, which utilizes a residual beam to eliminate the perturbationdue due to the vibrometer body vibration without any external reference surface, has been developed. Residual laser beam from the beam splitter is used to pick up the vibration of damper, which is mounted in the vibrometer, and combined with reference beam at the photodetector. The output signal of this detector and main signal are processed to extract the object vibration, using a least mean square adaptive algorithm. It is shown experimentally that the body vibration of 1-5 Hz can be effectively removed from the measured signal using DSP technology to extract unperturbed 100 Hz original signal.
Keywords
LDV; Vibration compensation; Adaptive Fillter;
Citations & Related Records
Times Cited By KSCI : 1  (Citation Analysis)
연도 인용수 순위
1 S. Aatola, Vibration measurements of automobile catalyst, First international Conference on Vibration Measurements by Laser Techniques, SPIE Vol. 2358, 359 364 (1994)
2 G. Liotta and C.Wang, Vibration measurement by a laser Doppler displacement meter ,Second international Conference on Vibration Measurements by Laser Techniques, SPIE Vol. 2868, 337 343 (1996)   DOI
3 J. Wang and E. Zhang, Application of laser vibrometer in vibration measurement of Microstructures ,Second international Conference on Vibration Measurements by Laser Techniques, SPIE Vol. 2868, 318 325 (1996)   DOI
4 A. Link, J. Gerhardt, and H. von Martens, Amplitude and phase calibration of accelerometers in the nanometer range, 37-48 (1966)
5 V.A. Grechikhin and B.S. Rinkevochius, Digital Hilbert transform for processing of laser Doppler vibrometer signals, Second international Conference on Vibration Measurements by Laser Techniques, SPIE Vol. 2868, 89-96 (1996)   DOI
6 Seonghoon Kim, Jinhwan Kim, Hoseong Kim, Development of high resolution laser Doppler Vibrometer, Trans. KIEE, Vol. 49C, No.2, 125-132 (2000)   과학기술학회마을
7 Simon Haykin, Adaptive Filter Theory, Prentice Hall, 1996