A inductive voltage divider(IVD) is widely used as a ratio arm of precise impedance measurement bridges at low frequencies of audio frequency range and a well fabricated IVD has ratio error of $10^{-7}$ order without any calibration. Recently, the order of $10^{-8}$ of the best measurement uncertainty is needed for calibration and maintenance of impedance standards as national standards. In order to achieve that uncertainty, the IVD which is used for a ratio arm of precise impedance measurement bridge should be calibrated within the uncertain of order of $10^{-8}$ For this purpose, a ratio calibration bridge for IVDs has been developed. The measurement uncertainties of both inphase and quadrature of the bridge are analyzed less then $3{\times}10^{-8}$ respectively at 1 kHz and 25 V.
Keywords
IVD(Inductive Voltage Divider); Totoidal Core; Inphase; Quadrature Error; Bridge;
Citations & Related Records
Times Cited By KSCI : 1  (Citation Analysis)
연도 인용수 순위
1 European cooperation for Accreditation, 'Expression of the Uncertainty of Measurement in Calibration', Publication Reference EA-4/02, December, 1999
2 이래덕, '측정불확도 평가 및 표현방법 -전기분야-', 한국표준과학연구원, KRISS-99-079-ET, 1999.
3 Thompson, A.M., 'A.C. Bridge Method for the Measurement of Three Terminal Admittance', IEEE Trans. on Instr. and Meas., Vol. IM-13, No.4, pp. 189-197, December, 1964   DOI
4 Cutkosky, R.D., 'Evaluation of the NBS Unit of Resistance Based on Computable Capacitor', J. Res. NBS, Vol. 65A, No.3, pp, 147-158, 1961   DOI
5 R. D. Cutkosky, J. Q. Shield, 'The Precision Measurement of Transformer Ratios', IRE trans. on Instrumentation, Vol. 1-9, Number 2, September, 1960
6 Raymond V. Lisle , Thomas L. Zapf, 'Comparison Calibration of Inductive Voltage Divider', ISA Transaction, Vol. 3, pp. 238-242, 1964
7 J. J. Hill, A P. Miller, 'A Seven-decade adjustable-ratio inductively coupled Voltage Divider with 0.1 part per Million Accuracy', Proc. Instn Elect. Engrs(GB), 109B, pp, 157-62, 1962
8 김한준, 이래덕, 강전홍, 한상옥, '저 주파수용 7-dial Inductive Voltage Divider 제작과 특성분석', Trans. KIEE., Vol. 53B, No.5, pp. 305-308, May, 2004   과학기술학회마을
9 http://kcdb.bipm.fr/appendixC/
10 Singer instrumentation, 'Theoretical analysis of accuracy of ratio transformers', Ratiotrans Engineering Bulletins, No.1, 1962