Abstract
Micro-weighing device by using null balance method is being essential part in fields of high-technology industries such as precision semiconductor industry, precision chemistry, biotechnology and genetics etc. Also, requirements for high resolution and for large measurement range increase more and more. The performance of the micro-weighing device can be determined by the mechanism design and analysis. The analytical design method has been proposed for the performance improvement such as resolution, measurement range and fast response. The 2-stage displacement amplification is designed to overcome the limit of conventional force transmitting lever. The parallel spring is designed for the measurement result independent of the input force position variation. Also, the natural frequency of mechanism is analyzed for the fast response. After each analysis, optimal design has been carried out. To verify the analysis and design result, characteristics experiments had been carried out after construction. Finally, the system had been controlled.
Keywords
Null balance; force; Measurement; Repeatability; Parallel spring; Resolution;
Citations & Related Records
연도 인용수 순위
  • Reference
1 Inverse Kinematic Modeling of a Coupled Flexure Hinge Mechanism /
[ Ryu J.W.;Lee S.Q.;Gwon D.G.;Moon K.S. ] / Mechatronics   DOI   ScienceOn
2 Cheshmehdoost, A. and Jones, B.E., 1995, Design and Performance Characteristics of an Integrated High Capacity DETF-Based Force Sensor, The 8th Int. Conf. on Solid-state Sensors and Actuators and Eurosensors IX Stockholm, Sweden, June25-29 pp.608-611
3 Hara, S., Kasai, S., Nishitani, T. and Nagashima, A., 1999, Development of Steady-State Magnetic Sensor, Rev. of Sci. Instrum., Vol.70, No.1, pp.435-438   DOI   ScienceOn
4 Kurosu, S., 1996, Perfomance of Gyroscopic Force Measuring Apparatus, Proc. of the Measurement of Force and Mass, pp.129-134
5 Peterson, I. R., 1997, Design Considerations for an Electrobalance Microforce Sensor, Rev. Sci. Instrum. Vol.68, No.2, pp.1130-1136   DOI   ScienceOn
6 Ryu J.W., Lee S.Q., Gwon D.G. and Moon K.S., 1999 Inverse Kinematic Modeling of a Coupled Flexure Hinge Mechanism, Mechatronics Vol.9, 657-674   DOI   ScienceOn
7 Maier, R. and Schmidt, G .. , 1989, Integrated Digital Control and Filtering for an E1ectrodynamically Compensated Weighcell, IEEE transactions on Instrumentation and Measurement, Vol.38, No.5 pp.998-1003   DOI   ScienceOn
8 Holford, K.M., Bakopoulos, c.P. and White N.M., 1990, The Development of a High Pressure, Thick-Film Sensor Proc Proc. of I Mech E Mechatronics Conf. pp.47-59
9 Quinn, T.J., 1992, The Beam Balance as an Instrument for Very Precise Weighing, Meas. Sci. Technol. 3, pp.141-159   DOI   ScienceOn
10 Balachandran, W., Halimic, M., Hodzic, M., Tariq, M., Enab, Y. and Cecelja, F., 1995, Optimal Digital Control and Filtering for Dynamic Weighing System, IEEE Instrumentation/ Measurement Technology Conf. - IMTC/95, Waltham, Boston, MA, USA, April 24- 26, pp. 293-298   DOI