Browse > Article
http://dx.doi.org/10.5572/KOSAE.2011.27.2.133

Selection of Calibration Approaches and Their Impact on the Quantification of Unknown Samples: Case Study on Reduced Sulfur Gases  

Jo, Hyo-Jae (Department of Environment & Energy, Sejong University)
Hong, One-Feel (Department of Environment & Energy, Sejong University)
Kim, Ki-Hyun (Department of Environment & Energy, Sejong University)
Publication Information
Journal of Korean Society for Atmospheric Environment / v.27, no.2, 2011 , pp. 133-141 More about this Journal
Abstract
In this study, different calibration approaches for reduced sulfur compounds (RSCs) were investigated by using thermal desorption coupled with gas chromatography (GC) and pulsed flame photometric detection (PFPD). To evaluate the effects of calibration procedures, gaseous standards of 4 RSCs ($H_2S$, $CH_3SH$, DMS, and DMDS) prepared at 10 ppm level were analyzed at 7 loading injection volumes (40, 60, 80, 100, 160, 240, and 320 ${\mu}L$). The results were then compared with calibration curves made with the Z (zero offset) and N (non-zero offset) method. The concentrations of unknown samples were then quantified by using R (ratio) method in which the slope values are compared between standards and samples. Secondly, in A (average) method, results obtained from a multi-point analysis of unknown samples were also averaged to extract representative values for each sample. Results of both experiments showed that analytical error of low molecular weight components (such as $H_2S$ and $CH_3SH$) was greatly expanded with the Z method. In conclusion, the combined application of N-A method was the more realistic approach to reduce biases in the quantification of RSCs.
Keywords
Reduced sulfur compounds; $H_2S$; Calibration; Standard;
Citations & Related Records
Times Cited By KSCI : 1  (Citation Analysis)
연도 인용수 순위
1 Jeon, S.-J. and G.-S. Heo (1999) Measurement of volatile organic sulfur compoundsin ambient air by canister and tedlar-bag sampling method, J. Korean Soc. Atmos. Environ., 15(4), 417-428. (in Korean with English abstract)
2 Kim, K.-H., S.-I. Oh, Y.-J. Choi, G.-H. Choi, and D.-W. Ju (2003) Calibration methods for the gas chromatographic analysis of ppt-level hydrogen sulfide $(H_2S)$ in air, J. Korean Soc. Atmos. Environ., 19(6), 679-687. (in Korean with English abstract)   과학기술학회마을
3 Kim, K.-H., S.-I. Oh, and Y.-J. Choi (2004) Studies of gas chromatographic analysis of malodorous S compounds in air, Analytical Science & Technology, 17(2), 145-152. (in Korean with English abstract)   과학기술학회마을
4 Kim, K.-H. (2008) A method to test the detectability of GC/PFPD for an extended concentration range with respect to reduced sulfur compounds, J. Separation Sci., 31(10), 1761-1768.   DOI   ScienceOn
5 Nagata, Y. (2003) Odor intensity and door threshold value, J. Japan Air Cleaning Association, 41(2), 17-25.
6 Black, M.S., P.H. Richard, and R.H. Dian (1978) Solid adsorbent preconcentration and gas chromatographic analysis of sulfur gases, Analyt. Chem., 50(7), 848-851.   DOI   ScienceOn