Browse > Article
http://dx.doi.org/10.14368/jdras.2013.29.2.111

Effects of Coronal Thread Pitch in Scalloped Implant with 2 Different Connections on Loading Stress using 3 Dimensional Finite Element Analysis  

Choi, Kyung-Soo (Department of Prosthodontics, School of Dentistry, Pusan National University)
Park, Seong-Hun (School of Mechanical Engineering, Pusan National University)
Lee, Jae-Hoon (School of Mechanical Engineering, Pusan National University)
Huh, Jung-Bo (Department of Prosthodontics, School of Dentistry, Pusan National University)
Yun, Mi-Jung (Department of Prosthodontics, School of Dentistry, Pusan National University)
Jeon, Young-Chan (Department of Prosthodontics, School of Dentistry, Pusan National University)
Jeong, Chang-Mo (Department of Prosthodontics, School of Dentistry, Pusan National University)
Publication Information
Journal of Dental Rehabilitation and Applied Science / v.29, no.2, 2013 , pp. 111-118 More about this Journal
Abstract
Purpose of present study is to investigate the effects of thread pitch in coronal portion in scalloped implant with 2 different connections on loading stress using 3 dimensional finite element analysis. Scalloped implant with 4 different thread pitches (0.4mm, 0.5mm, 0.6, and 0.7mm) in the coronal part was modeled with 2 different implant-abutment connections. Platform matching connection had the same implant and abutment diameter so that they were in flush contact at the periphery while platform mismatching connection had smaller abutment diameter than implant so that their connection was made away from periphery of implant-bone interface. Occlusal loading of 100N force was applied vertically and 30 degree obliquely to all 8 models and the maximum von Mises bone stress was identified. Loading stress as highly concentrated in cortical bone. Platform mismatching scalloped implant with small thread pitch (0.4mm) model had consistently lowest maximum von Mises bone stress in vertical and oblique loads. Platform matching model had lowest maximum von Mises bone stress with 0.6mm thread pitch in vertical load and with 0.4mm thread pitch in oblique load. Platform mismatching connection had important roles in reducing maximum von Mises bone stress. Scalloped implant with smaller coronal thread pitch showed trend of reducing maximum von Mises bone stress under load.
Keywords
Coronal thread pitch; Connection; Scalloped implant; 3D finite element analysis;
Citations & Related Records
연도 인용수 순위
  • Reference
1 Lindquist LW, Carlsson GE, Jemt T. A prospective 15-year follow-up study of mandibular fixed prosthesis supported by osseointegrated implants. Clinical results and marginal bone loss. Clin Oral Implants Res 1996;7:329336.   DOI   ScienceOn
2 Jemt T, Lekholm U,Adell R. Osseointegration in the treatment of partially edentulous patients: A preliminary study of 876 consecutively placed fixtures. Int J Oral Maxillofac Implants 1989;4:211217.
3 Adell R, Lekholm U, Rockler B, Brnemark P-I. A 15-year study of osseointegrated implants in the treatment of the edentulous jaw. Int J Oral Surg 1981;10:387416.   DOI
4 Brunski JB. Biomaterials and biomechanics in dental implant design. Int J Oral Maxillofac Implants 1988 Summer;3(2):85-97.
5 Steigenga JT, al-Shammari KF, Nociti FH, Misch CE, Wang HL. Dental implant design and its relationship to long-term implant success. Implant Dent 2003;12 (4):306-17.   DOI   ScienceOn
6 Albrektsson T, Sennerby L. State of the art in oral implants. J Clin Periodontol 1991 Jul;18(6):474-81.   DOI
7 Whrle, PS. Nobel perfect esthetic scalloped implant: Rationale for a new design. Clin Implant Dent Relat Res.2003 5: 64-73.   DOI   ScienceOn
8 Becker WC, Ochsenbein C, Tibbetts L, Becker BE. Alveolar bone anatomic profiles as measured from dry skulls. Clinical ramifications. J Clin Periodontol 1997 Oct;24(10): 727-31.   DOI   ScienceOn
9 Nowzari H, Chee W, Yi K, Pak M, Chung WH, Rich S. Scalloped dental implants: a retrospective analysis of radiographic and clinical outcomes of 17 Nobel- Perfect implants in 6 patients. Clin Implant Dent Relat Res 2006;8:1-10.   DOI   ScienceOn
10 Kan J, Rungcharassaeng K, Lieddelow G, Henry P, Goodacre C. Peri-implant tissue response following immediate provisional restoration of scalloped implants in the esthetic zone: A one-year pilot prospective multi-center study. J Prosthet Dent 2007;97:109-118.   DOI   ScienceOn
11 Tymstra N, Raghoebar GM, Vissink A, Den Hartog L, Stellingsma K, Meijer HJ. Treatment outcome of two adjacent implant crowns with different implant platform designs in the aesthetic zone: a 1-year randomized clinical trial. J Clin Periodontol 2010;38: 74-85.
12 Abrahamsson I, Berglundh T. Tissue characteristics at microthreaded implants: An experimental study in dogs. Clin Implant Dent Relat Res 2006; 8: 107-113.   DOI   ScienceOn
13 Lee D, Choi Y, Park K, Kim C, Moon I. Effect of microthread on the maintenance of marginal bone level: a 3-year prospective study. Clin Oral Implants Res 2007;18:465-470.   DOI   ScienceOn
14 Shin Y, Han C, Heo S, Kim S, Chun H. Radiographic evaluation of marginal bone level around implants with different neck designs after 1 year. The Int J Oral Maxillofac Implants 2006;21:789-794.
15 Choi KS, Lozada JL, Kan JY, Lee SH, Kim CS, Kwon TG. Study of an experimental microthreaded scalloped implant design: proximal bone healing at different interimplant distances in a canine model. Int J Oral Maxillofac Implants 2010;25:195-202.
16 Katranji A, Misch K, Wang HL. Cortical bone thickness in dentate and edentulous human cadavers. J Periodontol 2007;78:974-8.   DOI   ScienceOn
17 Chun H, Shin H, Han C, Lee S. Influence of implant abutment type on stress distribution in bone under various loading conditions using finite element analysis. Int J Oral Maxillofac Implants 2006;21:195-
18 Merz BR, Hunenbart S, Belser UC, Mechanics of the implant-abutment connection: an 8-degree taper compared to a butt joint connection. The Int J Oral Maxillofac Implants 2000;15:789-794.
19 Abkowitz S, Burke JJ, Hiltz RH. Titanium in Industry. New York: Van Nostrand Co Inc, 1955.
20 Misch CE, Suzuki JB, Misch-Dietsh FM, Bidez MW. A positive correlation between occlusal trauma and peri-implant bone loss: literature support. Implant Dent. 2005 Jun;14(2):108-16.   DOI   ScienceOn
21 Oh TJ, Yoon J, Misch C E, Wang HL. The causes of early implant bone loss: myth or science? J Periodontol 2002;73:322-333.   DOI   ScienceOn
22 Choi KS, Park SH, Lee JH, Jeon YC, Yun MJ, Jeong CM. Stress distribution of scalloped implant with different microthread and connection configurations using 3 dimensional finite element analysis. Int J Oral Maxillofac Implants (manuscript accepted).