Abstract
The objective of the present study was to perform a fracture analysis on fractured implant fixture after use in vivo and make clear the cause & mechanisms of failure. In case of fatigue fracture, the fractured surface represents fatigue striation. Fatigue striation indicate the progression of the crack front under cyclic loading, are characteristic of stage 2 crack growth. The site of crack initiation and stage 1 crack growth were not easily identified in any of the failure, presumably because of the complex microstructural features of the polycrystalline sample. In case of fractured by overload, dimpled or cleavage surface were observed. Using the interpretation of characteristic markings(ratchet mark, fatigue striation, dimple, cleavage et al) in fracture surfaces, failure events containing the crack origin, crack propagation, material deficiency could be understand.
Keywords
Fracture analysis; implant fixture; scanning electron microscope;
Citations & Related Records
연도 인용수 순위
  • Reference
1 Taskonak B, Yan J, Mecholsky JJ. Fractographic analyses of zirconia-based fixed partial dentures. Dent Mater 2008;24:1077-1082.   DOI   ScienceOn
2 Green NT, Machtei EE, Horwitz J, Peled M. Fracture of dental implants: Literature review and report of a case. Implant Dent 2002;11:137-143.   DOI   ScienceOn
3 Kelly JR, Campbell SD, Bowen HK. Fracture-surface analysis of dental ceramics. J Prosthet Dent 1989;62:536-541.   DOI   ScienceOn
4 Scherrer SS, Quinn JB, Quinn GD, Kelly JR. Failure analysis of ceramic clinical cases using qualitive fractography. Int J Prosthodont 2006;19:185-192.
5 Mills K, Davis JR, Destefani JD, Dieterich D. ASM Handbook, volume 12, Fractography. ASM international; 1992.
6 Rangert B, Krogh PH, Langer B, Roekel NV. Bending overload and implant failure: a retrospective clinical analysis. Int J Oral Maxillofac Impants 1995;10:326-334.
7 Powell GW, Mahmond SE, Mills K. Failure analysis and prevention. ASM Handbook, volume 11. ASM international; 1992.
8 Jones DD, Schroeder JR, Greene KA, Benzing B. Fractography of modern engineering materials: Composites and metals. 1985.
9 최병학. 금속 손상 진단. 서울: 진영사; 1997. pp1-33.
10 Zavanelli RA, Pessanha Henriques GE, Ferreira I. Corrosion-fatigue life of commercially pure titanium and Ti-6Al-4V alloys in different storage enviroments. J Prosthet Dent 2000;84:274-279.   DOI   ScienceOn
11 Bryant SR, Jankowski DM, Kim KS. Does the type of implant prosthesis affect outcomes for the completely edentulous arch. Int J Oral Maxillofac Implants 2007;22:117-139.
12 Hoshaw SJ, Brunski JB, Cochran GV. Mechanical loading of Branemark implants affects interfacial bone modeling and remodeling. Int Oral Maxillofac Implants 1994;9:345-360.
13 Jabbari YS, Fournelle R, Ziebert G, Toth J, Lacopino M. Mechanical behavior and failure analysis of prosthetic retaining screws after long-term use in vivo. part4: failure analysis of 10 fractured retaining screws retrieved from three patients. J Prosthodont 2008;17:201-210.   DOI   ScienceOn
14 Tonetti MS, Schmitt A. Pathogenesis of implant failures. Periodontology 2000 1994;4:127-138.   DOI   ScienceOn
15 Eckert SE, Woollan PC. Retrospective review of 1170 endosseous implants placed in partially edentulous jaws. J Prosthet Dent 1998;79:415-421.   DOI   ScienceOn
16 Morgan MJ, James DF, Pillar RM. Fractures of the fixture component of an osseointegrated implant. Int J Oral Maxillofac Implants 1993;8:409-414.
17 Capodiferro S, Favia G, Frenza GD, Grassi R. Clinical management and microscopic characterization of fatigue-induced failure of a dental implant. Case report. Head&Face Medicine 2006;2:18-23.   DOI   ScienceOn
18 Piattelli A, Piattelli M. Scarano A, Montesani L. Light and scanning electron microscopic report of four fractured implant. Int J Oral Maxillofac Implants 1998;13:561-564.
19 Balshi TJ. An analysis and management of fractured implants: a clinical report. Int J Oral Maxillofac Implants 1996;11:660-666.
20 Eckert SE, Meraw SJ, Cal E, Ow R. Analysis of incidence and associated factors with fractured implants: A retrospective study. Int J Oral Maxillofac Implants 2000;15:662-667.
21 Zarb GA, Schmitt A. The longitudinal clinical effectiveness of osseointegrated dental implants: The toronto study. Part III: Problems and complications encountered. J Prosthet Dent 1990;64:185-194.   DOI   ScienceOn
22 Quirynen M, Assche NV, Botticelli D, Berglundh T. How does the timing of implant placement to extraction affect outcome? Int J Oral Maxillofac Implant 2007;22:203-223.
23 Branemark PI, Adell R, Lekholm U, Rockler B. A 15-year study of osseointegrated implants in the treatment the edentulous jaw. Int J Oral Surg 1981; 10:387-416.   DOI
24 Jokstad A, Carr AB. What is the effect on outcomes of time to loading of a fixed or removable prosthesis placed on implant? Int J Oral Maxillofac implants 2007;22:19-48.
25 Weber HP, Sukotjo C. Does the type of implant prosthesis affect outcomes in the partially edentulous patient? Int J Oral Maxillofac Implants 2007;22: 140-172.
26 Davis DM, Waters NE. An investigation into the fracture behavior of a particulate-filled bis-GMA resin. J Dent Res 1987;66:1128-1133.   DOI   ScienceOn
27 Yamini S, Young R. The mechanical properties of epoxy resins. J Mater Sci Mater Med 1980;15:1823- 1831.
28 David DM, Waters NE. Fractography of a bis-GMA resins. J Dent Res 1989;68:1194-1198.   DOI   ScienceOn
29 Gapido CG, Kobayashi H, Miyakawa O, Kohno S. Fatigue resistance of cast occlusal rests using Co-Cr and Ag-Pd-Cu-Au alloys. J Prosthet Dent 2003;90: 261-269.   DOI   ScienceOn
30 Bates JF, Scott J. Studies related to the fracture of partial dentures-Fractography of Cobalt-chromium alloys. J Biomed Mater Res 1973;7:419-429.
31 Scherrer SS, Quinn GD, Quinn JB. Fractographic failure analysis of $Procera^{{\circledR}}$ Allceram crown using stereo and scanning electron microscopy. Dent Mater 2008;24:1107-1113.   DOI   ScienceOn
32 Iwama CY, Preston JD. Cobalt-Chromium-Titanium alloy for removable partial dentures. Int J Prosthodont 1997;10:309-317.