Effect of Implant Types and Bone Resorption on the Fatigue Life and Fracture Characteristics of Dental Implants  

Won, Ho-Yeon (Department of Prosthodontics, College of Dentistry, Dankook University)
Choi, Yu-Sung (Department of Prosthodontics, College of Dentistry, Dankook University)
Cho, In-Ho (Department of Prosthodontics, College of Dentistry, Dankook University)
Publication Information
Journal of Dental Rehabilitation and Applied Science / v.26, no.2, 2010 , pp. 121-143 More about this Journal
Abstract
To investigate the effect of implant types and bone resorption on the fracture characteristics. 4 types of Osstem$^{(R)}$Implant were chosen and classified into external parallel, internal parallel, external taper, internal taper groups. Finite elements analysis was conducted with ANSYS Multi Physics software. Fatigue fracture test was performed by connecting the mold to the dynamic load fatigue testing machine with maximum load of 600N and minimum load of 60N. The entire fatigue test was performed with frequency of 14Hz and fractured specimens were observed with Hitachi S-3000 H scanning electron microscope. The results were as follows: 1. In the fatigue test of 2 mm exposed implants group, Tapered type and external connected type had higher fatigue life. 2. In the fatigue test of 4 mm exposed implants group, Parallel type and external connected types had higher fatigue life. 3. The fracture patterns of all 4 mm exposed implant system appeared transversely near the dead space of the fixture. With a exposing level of 2 mm, all internally connected implant systems were fractured transversely at the platform of fixture facing the abutment. but externally connected ones were fractured at the fillet of abutment body and hexa of fixture or near the dead space of the fixture. 4. Many fatigue striations were observed near the crack initiation and propagation sites. The cleavage with facet or dimple fractures appeared at the final fracture sites. 5. Effective stress of buccal site with compressive stress is higher than that of lingual site with tensile stress, and effective stress acting on the fixture is higher than that of the abutment screw. Also, maximum effective stress acting on the parallel type fixtures is higher. It is careful to use the internal type implant system in posterior area.
Keywords
Bone resorption; Fatigue fracture; Implant type;
Citations & Related Records
Times Cited By KSCI : 4  (Citation Analysis)
연도 인용수 순위
1 Mericske-Stern R, Steinlin Schaffner T, Marti P, et al. Peri-implant mucosal aspects of ITI implants supporting overdentures. A five-year longitudinal study. Clin Oral Implants Res 1994;5:9-18.   DOI   ScienceOn
2 Zarb GA, Schmitt A. The longitudinal clinical effectiveness of osseointegrated dental implants: the Toronto study. Part I: Surgical results. J Prosthet Dent 1990;63: 451-7.   DOI   ScienceOn
3 Pfeifer AB, Sendyk CL, Sendyk WR, et al. Fracture of abutment screw supporting a cemented implant retained prosthesis with external hexagon connection: a case report. J Appl Oral Sci 2007;15(2):148-51.   DOI
4 Rangert B, Jemt T. Forces and moments on Branemark implants. Int J Oral Maxillofac Implants 1989;4:241-7.
5 Kim YS, Kim CW, Lim YJ, et al. Application of finite element analysis to evaluate implant fracture. J Korean Acad Prosthodont 2006;44(3):295-313.   과학기술학회마을
6 Bonakdarchiana M, Askaria N, Askarib M. Effect of face form on maximal molar bite force with natural dentition. Archives of Oral Biology 2009;54:201-4.   DOI   ScienceOn
7 Park WH, Lee YS. Three dimensional finite element stress analysis of implant prosthesis according to the different fixture locations and angulations. J Korean Acad Prosthodont 2005;43:61-77.   과학기술학회마을
8 Linkow LI, Donath K, Lemons JE. Retrieval analyses of a blade implant after 231 months of clinical function. Implant Dent 1992;1:37-43.
9 Khraisat A, Stegaroiu R, Nomura S, et al. Fatigue resistance of two implant/abutment joint designs. J Prosthet Dent 2002;88:604-10.   DOI   ScienceOn
10 Kim JH, Lee JB. A comparative study on the correlation between Korean foods and the fractures of PFG and all ceramic crowns for posterior applications. J Korean Acad Prosthodont 2009;47(2): 156-163.   과학기술학회마을   DOI   ScienceOn
11 Kenji W. Higuchi, Tiddo F. Implant survival rates in partially edentulous patients: A 3-year prospective multicenter study. J Oral Maxillofac Surgery 1995;53 (3): 264-8.   DOI   ScienceOn
12 Binon PP, McHugh MJ. The effect of eliminating implant/abutment rotational misfit on screw joint stability. Int J Prosthodont 1996;9:511-9.
13 Carlson B, Carlsson GE. Prosthodontic complications in osseointegrated dental implant treatment. Int J Oral Maxillofac Implants 1994;9:90-4.
14 Naert I, Quirynen M, van Steenberghe D, et al. A study of 589 consecutive implants supporting complete fixed prostheses. Part II: Prosthetic aspects. J Prosthet Dent 1992;68:949-56.   DOI   ScienceOn
15 Sato Y, Shindoi N, Hosokawa R, et al. A biomechanical effect of wide implant placement and offset placement of three implants in the posterior partially edentulous region. J Oral Rehabil 2000;27:15-21.   DOI   ScienceOn
16 Brunski JB, Puleo DA, Nanci A. Biomaterials and biomechanics of oral and maxillofacial implants: current status and future developments. Int J Oral Maxillofac Implants 2000;15:15-46.
17 Taylor TD, Agar JR, Vogiatzi T. Implant prosthodontics: current perspective and future directions. Int J Oral Maxillofac Implants 2000;15:66-75.
18 Patterson EA, Johns RB. Theoretical analysis of the fatigue life of fixture screws in osseointegrated dental implants. Int J Oral Maxillofac Implants 1992;7: 26-33.
19 Piattelli A, Piattelli M, Scarano A, et al. Light and scanning electron microscopic report of four fractured implants. Int J Oral Maxillofac Implants 1998;13: 561-4.
20 Tosun T, Karabuda C, Cuhadaroglu C. Evaluation of sleep bruxism by polysomnographic analysis in patients with dental implants. Int J Oral Maxillofac Implants 2003;18:286-92.
21 Morgan MJ, James DF, Pilliar RM. Fractures of the fixture component of an osseointegrated implant. Int J Oral Maxillofac Implants 1993;8:409-14.
22 Heather JC, John KS, Mark CV. Fractures related to occlusal overload with single posterior implants. A clinical report. J Prosthet Dent 2008;99:251-6.   DOI   ScienceOn
23 Balshi TJ, Hernandez RE, Pryszlak MC, et al. A comparative study of one implant versus two replacing a single molar. Int J Oral Maxillofac Implants 1996;11:372-8.
24 Cranin AN, Dibling JB, Simons A, et al. Report of the incidence of implant insert fracture and repair of Core-Vent dental implants. Int J Oral Maxillofac Implants 1990;16:184-8.
25 Lanyon LE, Goodship AE, Pye CJ, et al. Mechanically adaptive bone remodelling. J Biomech 1982;15:141-54.   DOI   ScienceOn
26 Quirynen M, Naert I, van Steenberghe D. Fixture design and overload influence marginal bone loss and fixture success in the Branemark system. Clin Oral Implants Res 1992;3:104-11.   DOI   ScienceOn
27 Machtei EE, Horwitz J, Peled M, et al. Fracture of dental implants: literature review and report of a case. Implant Dent 2002;11:137-43.   DOI   ScienceOn
28 Rangert B, Krogh PH, Langer B, et al. Bending overload and implant fracture: a retrospective clinical analysis. Int J Oral Maxillofac Implants 1995;10: 326-34.
29 Ragnar A, Eriksson B, Lekholm U. A longtermfollow up study of osseointegrated implants in the treatment of totally edentulous jaws. Int J Oral Maxillofac Implants 1990;5:347-59.
30 Takeshita F, Sutsugu T, Higuchi Y, et al. Histologic study of failed hollow implants. Int J Oral Maxillofac Implants 1996;11:245-50.
31 Levine RA, Clem DS, Wilson TG Jr, et al. Multicenter retrospective analysis of the ITI implant system used for single-tooth replacements: results of loading for 2 or more years. Int J Oral Maxillofac Implants 1999;14:516-20.
32 Eckert SE, Meraw SJ, Cal E, et al. Analysis of incidence and associated factors with fractured implants: a retrospective study. Int J Oral Maxillofac Implants 2000;15:662-7.
33 Tolman DE, Laney WR. Tissue-integrated prosthesis complications. Int J Oral Maxillofac Implants 1992;7: 477-84.
34 Jemt T, Lekholm U. Oral implant treatment in posterior partially edentulous jaws: A 5-year follow up report. Int J Oral Maxillofac Implants 1993;8: 635-40.
35 Choe HC, Lee JK, Chung CH. Analyses of fractured implant fixture after prolonged implantation. Metals and Materials International 2004;10(4):327-40.   DOI   ScienceOn
36 Albrektsson T, Dahl E, Enbom L, et al. Osseointegrated oral implants. A Swedish multicenter study of 8139 consecutively inserted Nobelpharma implants. J Periodontol 1988;59:287-96.   DOI
37 Albrektsson T. A multicenter report on osseointegrated oral implants. J Prosthet Dent 60:75-84.
38 Schwarz MS. Mechanical complications of dental implants. Clin Oral Implants Res 2000;1:156-8.
39 Adell R, Lekholm U, Rockler B, et al. A 15-year study of osseointegrated implants in the treatment of the edentulous jaw. Int J Oral Surg 1981;10:387-416.   DOI
40 Balshi TJ, Hernandez RE, Pryszlak MC, et al. An analysis and management of fractured implants. a clinical report. Int J Oral Maxillofac Implants 1996;11(5):660-5.
41 Adell R, Eriksson B, Lekholm U, et al. Long-term follow-up study of osseointegrated implants in the treatment of totally edentulous jaws. Int J Oral Maxillofac Implants 1990;5:347-59.
42 Albrektsson T, Zarb G, Worthington P, et al. The long-term efficacy of currently used dental implants: a review and proposed criteria of success. Int J Oral Maxillofac Implants 1986;1:11-25.
43 Albrektsson T, Blomberg S, Branemark A, et al. Edentulousness an oral handicap patient reactions to treatment with jawbone anchored prostheses. J Oral Rehabil 1987;14:503-11.   DOI