Comparison of Accuracy of Implant Torque Controllers  

Kim, Dae-Gon (Department of Prosthodontics, College of Dentistry, Kangnung National University)
Cho, Lee-Ra (Department of Prosthodontics, College of Dentistry, Kangnung National University)
Park, Chan-Jin (Department of Prosthodontics, College of Dentistry, Kangnung National University)
Publication Information
Journal of Dental Rehabilitation and Applied Science / v.24, no.2, 2008 , pp. 157-168 More about this Journal
Abstract
Tightening of the screws in implant restorations should be accurate and precise. If applied torque is too low, screw loosening would be occurred. With too high torque, the screw fracture might take place. Various torque generating devices are developed and employed to apply a proper torque. The purpose of this investigation was to determine and compare the accuracy of the torque controllers. In this study, 4 types of torque controllers were used; electronic torque controller, torque limiting device, torque indicating device and contra angle torque driver. Digital torque gauge was employed to measure the de-torque value. Thirty cycles of tightening and loosening were done with each torque controller. All implant torque controllers have shown slight errors and deviations. The torque liming device exhibited the most accurate data. No significant difference was found among the mean de-torque values of the electronic torque controller, torque indicating device and contra angle torque driver. In the limitation of this study, it would be recommended that the implant torque controllers should be checked whether uniformed and precise torque can be generated and a measuring error should be corrected.
Keywords
de-torque; torque; torque controllers;
Citations & Related Records
연도 인용수 순위
  • Reference
1 Misch CE. Contemporary Implant Dentistry
2 Jemt T. Failures and complications in 391 consecutively inserted fixed prostheses supported by Brånemark implants in edentulous jaws: a study of treatment from the time of prosthesis placement to the first annual checkup. Int J Oral Maxillofac Implants 1991;6(3):270-6
3 Jung RE, Pjetursson BE, Glauser R, Zembic A, Zwahlen M, Lang NP. A systematic review of the 5-year survival and complication rates of implant-supported single crowns. Clin Oral Implants Res 2008;19(2):119-30   DOI   ScienceOn
4 Jaarda MJ, Razzoog ME, Gratton DG. Comparison of "look-alike" implant prosthetic retaining screws. J Prosthodont 1995;4(1):23-7   DOI   ScienceOn
5 Weinberg LA, Kruger B. A comparison of implant/prosthesis loading with four clinical variables. Int J Prosthodont 1995;8(5):421-33
6 Rangert B, Jemt T, Jörneus L. Forces and moments on Branemark implants. Int J Oral Maxillofac Implants 1989;4(3):241-7
7 Bickford JH. An introduction to the design and behavior of bolted joints. Marcel Dekker, New York 1995;515-64
8 Jaarda MJ, Razzoog ME, Gratton DG. Effect of preload torque on the ultimate tensile strength of implant prosthetic retaining screws. Implant Dent 1994 ;3(1):17-21   DOI   ScienceOn
9 Goheen KL, Vermilyea SG, Vossoughi J, Agar JR. Torque generated by handheld screwdrivers and mechanical torquing devices for osseointegrated implants. Int J Oral Maxillofac Implants. 1994;9(2): 149-55
10 McGlumphy EA, Mendel DA, Holloway JA. Implant screw mechanics. Dent Clin North Am 1998;42(1): 71-89
11 Priest G. Single-tooth implants and their role in preserving remaining teeth: a 10-year survival study. Int J Oral Maxillofac Implants 1999;14(2):181-8
12 Binon PP. The external hexagonal interface and screw-joint stability: A primer on threaded fasteners in implant dentistry. Quint Dent Tech 2000;23: 91-104
13 Cehreli MC, Akça K, Tönük E. Accuracy of a manual torque application device for morse-taper implants: a technical note. Int J Oral Maxillofac Implants. 2004 ;19(5):743-8
14 Dellinges M, Curtis D. Effects of infection control procedures on the accuracy of a new mechanical torque wrench system for implant restorations. J Prosthet Dent 1996;75(1):93-8   DOI   ScienceOn
15 Khraisat A, Hashimoto A, Nomura S, Miyakawa O. Effect of lateral cyclic loading on abutment screw loosening of an external hexagon implant system. J Prosthet Dent 2004;91(4):326-34   DOI   ScienceOn
16 Patterson EA, Johns RB. Theoretical analysis of the fatigue life of fixture screws in osseointegrated dental implants. Int J Oral Maxillofac Implants 1992 ;7(1):26-33
17 Gutierrez J, Nicholls JI, Libman WJ, Butson TJ. Accuracy of the implant torque wrench following time in clinical service. Int J Prosthodont 1997;10(6):562-7
18 Burguete RL, Johns RB, King T, Patterson EA. Tightening characteristics for screwed joints in osseointegrated dental implants. J Prosthet Dent. 1994 ;71(6):592-9   DOI   ScienceOn
19 Jorneus L, Jemt T, Carlsson L. Loads and designs of screw joints for single crowns supported by osseointegrated implants. Int J Oral Maxillofac Implants 1992;7(3):353-9
20 Standlee JP, Caputo AA. Accuracy of an electric torque-limiting device for implants. Int J Oral Maxillofac Implants 1999;14(2):278-81
21 Lang LA, May KB, Wang RF. The effect of the use of a counter-torque device on the abutment-implant complex. J Prosthet Dent 1999;81(4):411-7   DOI   ScienceOn
22 Haack JE, Sakaguchi RL, Sun T, Coffey JP. Elongation and preload stress in dental implant abutment screws. Int J Oral Maxillofac Implants 1995;10(5):529-36
23 Standlee JP, Caputo AA, Chwu MY, Sun TT. Accuracy of mechanical torque - limiting devices for implants. Int J Oral Maxillofac Implants 2002;17(2): 220-4
24 Binon PP. Evaluation of the effectiveness of a technique to prevent screw loosening. J Prosthet Dent 1998;79(4):430-2   DOI   ScienceOn
25 Martin WC, Woody RD, Miller BH, Miller AW. Implant abutment screw rotations and preloads for four different screw materials and surfaces. J Prosthet Dent 2001;86(1):24-32   DOI   ScienceOn
26 Winkler S, Ring K, Ring JD, Boberick KG. Implant screw mechanics and the settling effect: overview. J Oral Implantol 2003;29(5):242-5   DOI   ScienceOn
27 Mitrani R, Nicholls JI, Phillips KM, Ma T. Accuracy of electronic implant torque controllers following time in clinical service. Int J Oral Maxillofac Implants 2001 ;16(3):394-9
28 Breeding LC, Dixon DL, Nelson EW, Tietge JD. Torque required to loosen single-tooth implant abutment screws before and after simulated function. Int J Prosthodont 1993;6(5):435-9
29 Dellinges MA, Tebrock OC. A measurement of torque values obtained with hand-held drivers in a simulated clinical setting. J Prosthodont 1993;2(4): 212-4   DOI   ScienceOn
30 Jaarda MJ, Razzoog ME, Gratton DG. Effect of preload torque on the ultimate tensile strength of implant prosthetic retaining screws. Implant Dent 1994 ;3(1):17-21   DOI   ScienceOn