Finite Element Stress Analysis of Implant Prosthesis According to Friction Fit or Slip Fit of Internal Connection System between Implant and Abutment  

Jang, Doo-Ik (Dept of Prosthodontics, College of Dentistry, Chosun University)
Jeong, Seung-Mi (Dept of Dentistry, College of Wonju Medicine, Yonsei University)
Chung, Chae-Heon (Dept of Prosthodontics, College of Dentistry, Chosun University)
Publication Information
Journal of Dental Rehabilitation and Applied Science / v.21, no.2, 2005 , pp. 113-132 More about this Journal
Abstract
The purpose of this study was to assess the stress-induced pattern at the supporting bone, the implant fixture, the abutment and the abutment screw according to a friction-fit joint (Astra; Model 1) or slip- fit joint (Frialit-2; Model 2) in the internal connection system under vertical and inclined loading using finite element analysis. In conclusion, in the internal connection system of the implant and the abutment connection methods, the stress-induced pattern at the supporting bone, the implant fixture, the abutment and the abutment screw according to the abutment connection form had difference among them, and the stress distribution pattern usually had a widely distributed tendency along the inner surface of the implant fixture contacting the abutment post. The magnitude of the stress distributed in the supporting bone, the implant fixture, the abutment and the abutment screw was higher in the friction-fit joint than in the slip-fit joint. But it is considered that the further study is necessary about how this difference in the magnitude of the stress have an effect on the practical clinic.
Keywords
Citations & Related Records
연도 인용수 순위
  • Reference
1 Levine RA, et al. A multicenter retrospective analysis of ITI implant system used for single-tooth replacements: Preliminary results at six or more months of loading. Int J Oral Maxillofacial Implants 1997; 12: 237-242
2 Chung KM, Chung CH, Jeong SM. Finite element analysis of implant prostheis according to platform width of fixture. J Korean Acad Prosthodont 2003; 41:674-688
3 Morimoto K, Kihara A, Takeshita F, Suetsugu T. An experimental study on the tissue compatibility of the titanium blade-vent implant coated with HAP-alumina in the semi- functional state. J Oral Implantol 1987; 13:387-401
4 Rangert B, Jemt T, Jomeus L. Forces and moments on Branemark implants. Int J Oral Maxillofac Implants 1989; 4:241-247
5 Weinberg LA. Force distribution in splinted anterior teeth. Oral Surg Oral Med Oral Pathol. 1957; 10:484-494   DOI   ScienceOn
6 Clelland NL, Lee JK, Bimbenet OC, Gilat A. Use of an axisymmetric finite element method to compare maxillary bone variables for a loaded implant. J Prosthodont 1993; 2:183-189   DOI   ScienceOn
7 Akca K, Iplikcioglu H. Evaluation of the effect of the residual bone angulation on implant-supported fixed prosthesis in mandibular posterior edentulism. Part II : 3-D finite element stress analysis. Implant Dent 2001; 10:239-245
8 Norton MR. Assessment of cold welding properties of internal conical interface two commercially available implant system. J Prosthet Dent 1999; 81:159-166   DOI   PUBMED   ScienceOn
9 Sutter F, Webber HP, Sorensen J, Belser U. The new restorative concept of the ITI dental implant system: design and engineering. Int J Perodont Rest Dent 1993; 13:409-431
10 Patterson EA, Burguete RL, Thoi MH, Johns RB. Distribution of load in an oral orosthesis system: an in vitro study. Int J Oral Maxillofac Implants 1995; 10:552-560
11 Rangert B, Krogh PHJ, Langer B, van Roekel NB. Bending overload and implant fracture: A retrospective clinical analysis. Int J Oral Maxillofac Implants 1995; 10:326-334
12 Holmes DC, Grigsby WR, Goel VK, Keller JC. Comparison of stress transmission in the IMZ implant system with polyoxymethylene or titanium intramobile element: A finite element stress analysis. Int J Oral Maxillofac Implants 1992; 7:450-458
13 Rangert B, Enouard F, Amoux JP, Sarment DP. Load factor control for implants in the posterior partially edentulous segment. Int J Oral Maxillofac Implants 1997; 12:360-370
14 Jemt T, Lekholm U, Grondahl K. A 3-year follow-up study of early single implant restorations ad modem Branemark, Int J Oral Perodont Rest Dent 1990; 10: 340-349
15 Mollersten L, Lockowandt P, Linden LA. Comparison of strength and failure mode of seven implant systems: An in vitro test. J Prosthet Dent 1997;78:582-591   DOI   ScienceOn
16 Merz BR. Hunenbart S, Belser UC. Mechanics of the implant-abutment connection. An 8-degree taper compared to a butt joint connection. Int J Oral Maxillofac Implants 2000; 15: 519-526
17 Krennmair G. Schmidinger S, Waldenberger O. Single-tooth replacement with frialit-2 system: A retrospective clinical analysis of 146 implants. Int J Oral Maxilofac Implanst. 2002; 17:78-85
18 Dan ET, William RL. Tissue-integrated prostheses complication. Int J Oral Maxillofac Implants 1992; 7:477-484
19 Norton MR. In vitro evaluation of the strength of the conical implant-to-abutment joint in two commercially available implant systems.. J Prosthet Dent 2000; 83:567-571   DOI   PUBMED   ScienceOn
20 Weinberg LA. Force distribution in splinted posterior teeth. Oral Surg Oral Med Oral Pathol. 1957; 10:1268-1276   DOI   ScienceOn
21 Matsushita Y, Kithoh M, Mizuta K, Ikeda H, Suetsugu T. Two-dimentional FEM analysis of hydroxyapatite implants: diameter effects on stress distribution. J Oral Implantol 1990; 16:6-11
22 Jemt T, Linden B, Lekholm U. Failures and complications in 127 consecutive placed fixed partial prostheses supported by Branemark implants. Int J Oral Maxillofac Implants 1992; 7:40-44
23 Balfour A, O'Brien GR. Comparative study of antirotational single tooth abutments. J Prosthet Dent 1995; 73:36-43   DOI   ScienceOn
24 Weinberg LA. The biomechanics of force distribution in implant- supported prostheses. Int J Oral Maxillofac Implants 1993; 8:19-31
25 Boggan RS, Strong JT, Misch CE, Bidez MW. Influence of hex geometry and prosthetic table width on static and fatigue strength of dental implants. J Prosthet Dent 1999; 82:436-440   DOI   ScienceOn
26 Kim NH, Chung CH, Son MK, Back DH. A study on the fit of the implant-abutment-screw interface. J Korean Acad Prosthodont 2003; 41:503-518
27 Kirsch A, Neuendorff G, Ackermann KL, Nagel R, Durr W. Camlog connection: Requirements for a reliable implant prosthetic treament concept; tooth for tooth restoration. Quintessenz 1999; 50:1-18
28 Jemt T et al. Osseointegrated implants for single tooth replacement. A l-year report from a multicenter prospective study. Int J Oral Maxillofac Implants 1991; 6:29-36
29 Akca K, Cehreli MC, lplikcioglu H. Evaluation of the mechanical characteristics of the implant abutment complex of a reduced diameter morse taper implant: a nonlinear finite element stress analysis, Clinical Oral Implants Research 2003;14:444-455   DOI   ScienceOn
30 Haraldson T, Zarb GA. A 10-year follow-up study of the masticatory system after treatment with osseointegrated implant bridges. Scand J Dent Rec 1988; 96:243-252
31 Lum LB. A biomechanical rationale for the use of short implants. J. Oral Implantol 1991; 17:126-131
32 Norton MR. An in vitro evaluation of the strength of an internal conical interface compared to a butt joint interface in implant design. Clin Oral Implants Res 1997; 8:290-298   DOI   PUBMED   ScienceOn
33 Sutter F, et al. The new concept of ITI hollow-cylinder and hollow-screw implants : Part 1. Engineering and design. Int J Oral Maxilofac Implant. 1988; 3:161-172
34 Lum LB, Osier J.F. Load transfer from endosteal implants to supporting bone: An analysis using statics. Part one: Horizontal loading. J Oral Implantol 1992; 18:343-348