Evaluating Accuracy according to the Evaluator and Equipment Using Electronic Apex Locators  

Yu, Beom-Young (Department of Dental Science, Graduate School, Kyungpook National University)
Son, Keunbada (Department of Dental Science, Graduate School, Kyungpook National University)
Lee, Kyu-Bok (Advanced Dental Device Development Institute (A3DI), Kyungpook National University)
Publication Information
Journal of Korean Dental Science / v.13, no.2, 2020 , pp. 52-58 More about this Journal
Abstract
Purpose: Using two types of electronic apex locators, this study aimed to investigate the differences in accuracy according to the evaluator and equipment. Materials and Methods: Artificial teeth of the lower first premolars and two mandibular acrylic models (A and B) were used in this study. In the artificial teeth, the pulp chamber was opened and the access cavity was prepared. Using calibrated digital Vernier calipers, the distance from the top of the cavity and the root apex was measured to assess the actual distance between two artificial teeth. The evaluation was conducted by 20 dentists, and each evaluator repeated measurements for each electronic apex locator five times. The difference between the actual distance from the top of the cavity to the root apex and the distance measured using electronic measuring equipment was compared. For statistical analysis, the Friedman test the Mann-Whitney U-test were conducted and the differences between groups were analyzed (α=0.05). Result: As for the accuracy of measurement according to the two types of electronic apex locators, the value of the measurement error was 0.4753 mm in Dentaport ZX and 0.3321 mm in E-Cube Plus. Moreover, electronic apex locators Dentaport ZX and E-Cube Plus showed statistically significant differences (P<0.05). As for the difference in the accuracy of the two types of electronic apex locators according to the evaluator, the resulting values differed depending on the evaluator and showed a statistically significant difference (P<0.001). Conclusion: Electronic apex locator E-Cube Plus showed higher accuracy than did Dentaport ZX. Nevertheless, both types of electronic apex locators showed 100% accuracy in finding the region within root apex ±0.5 mm zone. Furthermore, according to the evaluator, the two electronic apex locators showed different resulting values.
Keywords
Apical foramen; Apicoectomy; Dimensional measurement accuracy; Tooth apex; Tooth root;
Citations & Related Records
연도 인용수 순위
  • Reference
1 Shabahang S, Goon WW, Gluskin AH. An in vivo evaluation of Root ZX electronic apex locator. J Endod. 1996; 22: 616-8.   DOI
2 Connert T, Judenhofer MS, Hulber-J M, Schell S, Mannheim JG, Pichler BJ, Lost C, ElAyouti A. Evaluation of the accuracy of nine electronic apex locators by using micro-CT. Int Endod J. 2018; 51: 223-32.   DOI
3 Kim E, Lee SJ. Electronic apex locator. Dent Clin North Am. 2004; 48: 35-54.   DOI
4 Kobayashi C. Electronic canal length measurement. Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 1995; 79: 226-31.   DOI
5 Chang HS, Park SH, Cho KM, Kim JW. Accuracy of electronic apex locator in using different Root canal irrigant. J Dent Rehabil Appl Sci. 2012; 28: 139-45.
6 Seltzer S, Soltanoff W, Sinai I, Smith J. Biologic aspects of endodontics. IV. Periapical tissue reactions to root-filled teeth whose canals had been instrumented short of their apices. Oral Surg Oral Med Oral Pathol. 1969; 28: 724-38.   DOI
7 Suzuki K. Experimental study on iontophoresis. Jpn J Stomatol. 1942; 16: 411-29.   DOI
8 Tosun G, Erdemir A, Eldeniz AU, Sermet U, Sener Y. Accuracy of two electronic apex locators in primary teeth with and without apical resorption: a laboratory study. Int Endod J. 2008; 41: 436-41.   DOI
9 Pagavino G, Pace R, Baccetti T. A SEM study of in vivo accuracy of the Root ZX electronic apex locator. J Endod. 1998; 24: 438-41.   DOI
10 Cho YL, Son WH, Hwang HK. An accuracy of the several electronic apex locators on the mesial root canal of the mandibular molar. J Korean Acad Conserv Dent. 2005; 30: 477-85.   DOI
11 Fouad AF, Krell KV, McKendry DJ, Koorbusch GF, Olson RA. Clinical evaluation of five electronic root canal length measuring instruments. J Endod. 1990; 16: 446-9.   DOI
12 Ricard O, Roux D, Bourdeau L, Woda A. Clinical evaluation of the accuracy of the Evident RCM Mark II apex locator. J Endod. 1991; 17: 567-9.   DOI
13 Ounsi HF, Naaman A. In vitro evaluation of the reliability of the Root ZX electronic apex locator. Int Endod J. 1999; 32: 120-3.   DOI
14 Frank AL, Torabinejad M. An in vivo evaluation of Endex electronic apex locator. J Endod. 1993; 19: 177-9.   DOI
15 Pi CY, Kim ES, Jung IY, Lee SJ. In vivo evaluation of accuracy and consistency of two electronic apex locators. J Korean Acad Conserv Dent. 2010; 35: 453-60.   DOI
16 Mayeda DL, Simon JH, Aimar DF, Finley K. In vivo measurement accuracy in vital and necrotic canals with the Endex apex locator. J Endod. 1993; 19: 545-8.   DOI
17 Fouad AF, Rivera EM, Krell KV. Accuracy of the Endex with variations in canal irrigants and foramen size. J Endod. 1993; 19: 63-7.   DOI
18 Gutmann JL, Leonard JE. Problem solving in endodontic working-length determination. Compend Contin Educ Dent. 1995; 16: 288, 290, 293-4 passim; quiz 304.
19 Seltzer S, Soltanoff W, Sinai I, Goldenberg A, Bender IB. Biologic aspects of endodontics. 3. Periapical tissue reactions to root canal instrumentation. Oral Surg Oral Med Oral Pathol. 1968; 26: 694-705.   DOI
20 Park SY, Lee DK, Hwang HK. An evaluation of the accuracy of Root ZX according to the conditions of major apical foramen. Restor Dent Endod. 2012; 37: 68-73.   DOI
21 Custer LE. Exact methods of locating the apical foramen. J Natl Dent Assoc. 1918; 5: 815-9.   DOI
22 Sunada I. New method for measuring the length of the root canal. J Dent Res. 1962; 41: 375-87.   DOI
23 Inoue N. An audiometric method for determining the length of root canals. J Can Dent Assoc (Tor). 1973; 39: 630-6.
24 Yamashita Y. A study of a new electronic root canal measuring device using relative values of frequency response: influences of the diameter of apical foramen, the size of electrode, and the concentration of sodium hypochlorite. Jpn J Conserv Dent. 1990; 33: 547-59.
25 Kobayashi C, Suda H. New electronic canal measuring device based on the ratio method. J Endod. 1994; 20: 111-4.   DOI