Geometrical design characteristics of orthodontic mini-implants predicting maximum insertion torque  

Katic, Visnja (Department of Pediatric Dentistry and Orthodontics, School of Medicine, University of Rijeka)
Kamenar, Ervin (Department of Mechanical Engineering Design, Faculty of Engineering, University of Rijeka)
Blazevic, David (Department of Mechanical Engineering Design, Faculty of Engineering, University of Rijeka)
Spalj, Stjepan (Department of Pediatric Dentistry and Orthodontics, School of Medicine, University of Rijeka)
Publication Information
The korean journal of orthodontics / v.44, no.4, 2014 , pp. 177-183 More about this Journal
Abstract
Objective: To determine the unique contribution of geometrical design characteristics of orthodontic mini-implants on maximum insertion torque while controlling for the influence of cortical bone thickness. Methods: Total number of 100 cylindrical orthodontic mini-implants was used. Geometrical design characteristics of ten specimens of ten types of cylindrical self-drilling orthodontic mini-implants (Ortho Easy$^{(R)}$, Aarhus, and Dual Top$^{TM}$) with diameters ranging from 1.4 to 2.0 mm and lengths of 6 and 8 mm were measured. Maximum insertion torque was recorded during manual insertion of mini-implants into bone samples. Cortical bone thickness was measured. Retrieved data were analyzed in a multiple regression model. Results: Significant predictors for higher maximum insertion torque included larger outer diameter of implant, higher lead angle of thread, and thicker cortical bone, and their unique contribution to maximum insertion torque was 12.3%, 10.7%, and 24.7%, respectively. Conclusions: The maximum insertion torque values are best controlled by choosing an implant diameter and lead angle according to the assessed thickness of cortical bone.
Keywords
Orthodontic mini-implant; Mini-implant stability; Mini-implant orthodontic anchorage; Implant insertion torque;
Citations & Related Records
Times Cited By KSCI : 2  (Citation Analysis)
연도 인용수 순위
1 Tseng YC, Hsieh CH, Chen CH, Shen YS, Huang IY, Chen CM. The application of mini-implants for orthodontic anchorage. Int J Oral Maxillofac Surg 2006;35:704-7.   DOI   ScienceOn
2 Poggio PM, Incorvati C, Velo S, Carano A. "Safe zones": a guide for miniscrew positioning in the maxillary and mandibular arch. Angle Orthod 2006; 76:191-7.
3 Kanomi R. Mini-implant for orthodontic anchorage. J Clin Orthod 1997;31:763-7.
4 Motoyoshi M. Clinical indices for orthodontic miniimplants. J Oral Sci 2011;53:407-12.   DOI
5 Cheng SJ, Tseng IY, Lee JJ, Kok SH. A prospective study of the risk factors associated with failure of mini-implants used for orthodontic anchorage. Int J Oral Maxillofac Implants 2004;19:100-6.
6 Meredith N. Assessment of implant stability as a prognostic determinant. Int J Prosthodont 1998;11:491-501.
7 Motoyoshi M, Hirabayashi M, Uemura M, Shimizu N. Recommended placement torque when tightening an orthodontic mini-implant. Clin Oral Implants Res 2006;17:109-14.   DOI   ScienceOn
8 Pithon MM, Nojima MG, Nojima LI. In vitro evaluation of insertion and removal torques of orthodontic mini-implants. Int J Oral Maxillofac Surg 2011;40:80-5.   DOI
9 Lim SA, Cha JY, Hwang CJ. Insertion torque of orthodontic miniscrews according to changes in shape, diameter and length. Angle Orthod 2008;78:234-40.   DOI   ScienceOn
10 Xu Z, Zhao L, Wu Y, Wei X, Wang J, Tang N, et al. Histomorphometric and biomechanical analyses of the osseointegration of loaded orthodontic microscrews inserted at different cortical bone thickness sites. Oral Surg Oral Med Oral Pathol Oral Radiol 2012. doi: 10.1016/j.oooo.2012.06.007. [Epub ahead of print]
11 Pithon MM, Nojima MG, Nojima LI. Primary stability of orthodontic mini-implants inserted into maxilla and mandible of swine. Oral Surg Oral Med Oral Pathol Oral Radiol 2012;113:748-54.   DOI
12 Studenmund AH, Cassidy HJ. Instructor's manual to accompany. Using econometrics, a practical guide. Boston: Little Brown; 1987.
13 Miyawaki S, Koyama I, Inoue M, Mishima K, Sugahara T, Takano-Yamamoto T. Factors associated with the stability of titanium screws placed in the posterior region for orthodontic anchorage. Am J Orthod Dentofacial Orthop 2003;124:373-8.   DOI   ScienceOn
14 American Gear Manufacturers Association, American National Standards Institute. Gear nomenclature, definitions of terms with symbols: AGMA standard. Alexandria: American Gear Manufacturers Association;2005.
15 Dalstra M, Cattaneo PM, Melsen B. Load transfer of miniscrews for orthodontic anchorage. Orthodontics 2004;1:53-62.
16 Kondo S, Okada Y, Iseki H, Hori T, Takakura K, Kobayashi A, et al. Thermological study of drilling bone tissue with a high-speed drill. Neurosurgery 2000;46:1162-8.   DOI
17 Brisman DL. The effect of speed, pressure, and time on bone temperature during the drilling of implant sites. Int J Oral Maxillofac Implants 1996;11:35-7.
18 Toews AR, Bailey JV, Townsend HG, Barber SM. Effect of feed rate and drill speed on temperatures in equine cortical bone. Am J Vet Res 1999;60:942-4.
19 Kim JS, Choi SH, Cha SK, Kim JH, Lee HJ, Yeom SS, et al. Comparison of success rates of orthodontic mini-screws by the insertion method. Korean J Orthod 2012;42:242-8.   DOI
20 Cha JY, Yu HS, Hwang CJ. The validation of Periotest values for the evaluation of orthodontic miniimplants' stability. Korean J Orthod 2010;40:167-75.   DOI
21 Aparicio C, Lang NP, Rangert B. Validity and clinical significance of biomechanical testing of implant/ bone interface. Clin Oral Implants Res 2006;17 (Suppl 2):2-7.   DOI   ScienceOn
22 Wilmes B, Panayotidis A, Drescher D. Fracture resistance of orthodontic mini-implants: a biomechanical in vitro study. Eur J Orthod 2011;33:396-401.   DOI
23 Kim SJ, Yoo J, Kim YS, Shin SW. Temperature change in pig rib bone during implant site preparation by low-speed drilling. J Appl Oral Sci 2010; 18:522-7.   DOI
24 Markovic A, Misic T, Milicic B, Calvo-Guirado JL, Aleksic Z, Dinic A. Heat generation during implant placement in low-density bone: effect of surgical technique, insertion torque and implant macro design. Clin Oral Implants Res 2013;24:798-805.   DOI
25 Friberg B, Sennerby L, Roos J, Johansson P, Strid CG, Lekholm U. Evaluation of bone density using cutting resistance measurements and microradiography: an in vitro study in pig ribs. Clin Oral Implants Res 1995;6:164-71.   DOI   ScienceOn
26 Wilmes B, Drescher D. Impact of bone quality, implant type, and implantation site preparation on insertion torques of mini-implants used for orthodontic anchorage. Int J Oral Maxillofac Surg 2011;40:697-703.   DOI