Browse > Article

Effects of drilling process in stability of micro-implants used for the orthodontic anchorage  

Chang, Young-Il (Department of Orthodontics, College of Dentistry, Seoul National University)
Kim, Jong-Wan (Department of Orthodontics, College of Dentistry, Seoul National University)
Publication Information
The korean journal of orthodontics / v.32, no.2, 2002 , pp. 107-115 More about this Journal
Abstract
The aim of this study was to investigate experimentally the mechanical and histological effect of drilling process on the stability of micro-implant used for the orthodontic anchorage. For this purpose, 32 micro-implants(Osas$^{(R)}$, Epoch medical, ${\phi}$1.6 mm) were inserted into maxilla, mandible and palate in two beagle dogs. 16 micro-implants(8 per dog) were inserted after drilling with pilot drilling bur (drill method group). 16 micro-implants(8 per dog) were inserted without drilling (drill-free method group). After 1 week, micro-implants were loaded by means of Ni-Ti coil spring (Ni-Ti springs-extension$^{(R)}$, Ormco) with 200-300 gm force. Following 12 weeks, the micro-implants and the surrounding bone were removed. Before sacrifice, the mobilities were tested with Periotest$^{(R)}$(Siemens). Undecalcified serial sections with the long axis were made and the histologic evaluations were done. The results of this study were as follow ; 1. The osseointegration was found in both of drill-free method group and drill method group 2. Two of drill method group and one of drill-free method group in 32 micro-implants were lost after loading. 3. The mobilities of drill-free method group were less than drill method group 4. The bone contact on surface of micro-implants in drill-free method group was more than drill method group but there was no significant difference between groups. 5. The bone density in threads of micro-implants in drill-free method group was more than drill method group and there was significant difference between groups. These results suggest that drill-free method in insertion of micro-implants is superior to drill method on the stabilities, bone remodeling and osseointegrations under early loading.
Keywords
Orthodontic anchorage; Micro-implant; Drill-free method; Early loading;
Citations & Related Records
연도 인용수 순위
  • Reference
1 Bahr W. Pretapped and self-tapping screws in the human midface : torque measurements and bone screw interface. Int J Oral Maxillofac Surg 1990 : 19 : 51-3   DOI   ScienceOn
2 Bahr W. Pretapped and self-tapping screws in the human midface : torque measurements and bone screw interface. Int J Oral Maxillofac Surg 1990 : 19 : 51-3   DOI   ScienceOn
3 Bahr W. Pretapped and self-tapping screws in the human midface : torque measurements and bone screw interface. Int J Oral Maxillofac Surg 1990 : 19 : 51-3   DOI   ScienceOn
4 Albrektsson T. Direct bone anchorage of dental implants. J Prosthet Dent 1983 : 50 : 255-61   DOI   ScienceOn
5 Maniatopoulos C, Pilliar RM, Smith DC. Threaded versus poroussurfaced design for implant stabilization in bone-endodontic implant model. J Biomed Mater Res 1986 : 20 : 1309-33   DOI   ScienceOn
6 Weiss CM. A comparative analysis of fibre-osteal and osteal integration and other variables that affect long term bone maintenance around dental implants J Oral Irnplantol 1987 : 13 : 467-87
7 Newman MG, Flemrnig TF. Periodontal considerations of implants and implant-associated microbiota. J Dent Educ 1988 : 52 : 737-44
8 Drago CJ, A prospective study to assess osseointegration of dental endosseous implants with the Periotest instrument. Int J Oral Maxillofac Implants 2000 : 15 : 389-95
9 Bahr W. Pretapped and self-tapping screws in the human midface : torque measurements and bone screw interface. Int J Oral Maxillofac Surg 1990 : 19 : 51-3   DOI   ScienceOn
10 Bahr W. Pretapped and self-tapping screws in the human midface : torque measurements and bone screw interface. Int J Oral Maxillofac Surg 1990 : 19 : 51-3   DOI   ScienceOn
11 Bahr W. Pretapped and self-tapping screws in the human midface : torque measurements and bone screw interface. Int J Oral Maxillofac Surg 1990 : 19 : 51-3   DOI   ScienceOn
12 van Steenberghe D, Tricio J, Naert I, Nys M. Damping characteristics of bone-to-implant interfaces. A clinical study with the Periotest device. Clin Oral Implants Res 1995 : 6 : 31-9
13 Lindstrom J, Branemark PI, Albrektsson T. mandibular reconstruction using the preformed autologous bone graft. Scan J Plast Reconstr Surg 1981 : 15 : 29-38   DOI
14 Bahr W. Pretapped and self-tapping screws in the human midface : torque measurements and bone screw interface. Int J Oral Maxillofac Surg 1990 : 19 : 51-3   DOI   ScienceOn
15 Goodman S, Wang JS, Doshi A, Aspenberg P. Difference in bone ingrowth after one versus two daily episodes of micromotion : experiments with titanium chambers in rabbits. J Biomed Mater Res 1993 : 27 : 1419-24   DOI   ScienceOn
16 Bahr W. Pretapped and self-tapping screws in the human midface : torque measurements and bone screw interface. Int J Oral Maxillofac Surg 1990 : 19 : 51-3   DOI   ScienceOn
17 Bahr W. Pretapped and self-tapping screws in the human midface : torque measurements and bone screw interface. Int J Oral Maxillofac Surg 1990 : 19 : 51-3   DOI   ScienceOn
18 Thomsen P, Ericson LE. Light and transmission electron microscopy used to study the tissue morphology close to implants. Biomaterials 1985 : 6 : 421-4   DOI   ScienceOn
19 Kanomi R. Mini-implant for orthodontic anchorage. J Clin Orthod 1997 : 31 : 763-7
20 박효상. micro-implant 를 이용한 교정치료. 나래출판사 2001: 45-6
21 Bahr W. Pretapped and self-tapping screws in the human midface : torque measurements and bone screw interface. Int J Oral Maxillofac Surg 1990 : 19 : 51-3   DOI   ScienceOn
22 Bahr W. Pretapped and self-tapping screws in the human midface : torque measurements and bone screw interface. Int J Oral Maxillofac Surg 1990 : 19 : 51-3   DOI   ScienceOn
23 김종완, 장영일. 교정적 고정원으로 이용되는 drill-free micro-implant에 대한 조직학적 연구. 서울대학교대학원 석사논문. 2002
24 Bahr W. Pretapped and self-tapping screws in the human midface : torque measurements and bone screw interface. Int J Oral Maxillofac Surg 1990 : 19 : 51-3   DOI   ScienceOn
25 Pilliar RM, Cameron HU, Welsh RP, Binnington AG. Radiographic and morphologic studies of load-bearing porous surfaced structural implants. Clin Orthop 1981 : 156 : 249-57
26 Bahr W. Pretapped and self-tapping screws in the human midface : torque measurements and bone screw interface. Int J Oral Maxillofac Surg 1990 : 19 : 51-3   DOI   ScienceOn
27 Costa A, Raffainl M, Melsen B. Miniscrews as orthodontic anchorage: A preliminary report. Int J Adult Orthod Orthognath Surg 1998 : 13 : 201-9
28 Bahr W. Pretapped and self-tapping screws in the human midface : torque measurements and bone screw interface. Int J Oral Maxillofac Surg 1990 : 19 : 51-3   DOI   ScienceOn
29 Bahr W. Pretapped and self-tapping screws in the human midface : torque measurements and bone screw interface. Int J Oral Maxillofac Surg 1990 : 19 : 51-3   DOI   ScienceOn
30 Umemori M, Sugawara J, Mitani H, Nagasaka H, Kawamura H. Skeletal anchorage system for open-bite correction. Am J Orthod Dentofac Orthop 1999 : 115 : 166-74   DOI   ScienceOn
31 Meffert RM., Langer B, Fritz ME. Dental implant: a review. J Periodontol 1992 : 63 : 859-70   DOI
32 Ueda M, Matsuki M, Jacobsson M, Tjellstrorn A. Relationship between insertion torque and removal torque analyzed in fresh temporal bone. Int J Oral Maxillofac Implants 1991 : 6 : 442-7
33 Bahr W. Pretapped and self-tapping screws in the human midface : torque measurements and bone screw interface. Int J Oral Maxillofac Surg 1990 : 19 : 51-3   DOI   ScienceOn
34 Bahr W. Pretapped and self-tapping screws in the human midface : torque measurements and bone screw interface. Int J Oral Maxillofac Surg 1990 : 19 : 51-3   DOI   ScienceOn
35 Babbush CA. Endosseous blade-vent implants: A research review. J Oral Surg 1972 : 30 : 168-75
36 Bahr W. Pretapped and self-tapping screws in the human midface : torque measurements and bone screw interface. Int J Oral Maxillofac Surg 1990 : 19 : 51-3   DOI   ScienceOn
37 Melsen B, Costa A. Immediate loading of implants used for orthodontic anchorage. Clin Orthod Res 2000 : 3 : 23-8   DOI   ScienceOn
38 Albrektsson T, Jacobsson M. Bone-metal interface in osseointegration. J Prosthet Dent 1987 : 57 : 597-607   DOI   ScienceOn
39 Bahr W. Pretapped and self-tapping screws in the human midface : torque measurements and bone screw interface. Int J Oral Maxillofac Surg 1990 : 19 : 51-3   DOI   ScienceOn
40 Bahr W. Pretapped and self-tapping screws in the human midface : torque measurements and bone screw interface. Int J Oral Maxillofac Surg 1990 : 19 : 51-3   DOI   ScienceOn
41 Bahr W. Pretapped and self-tapping screws in the human midface : torque measurements and bone screw interface. Int J Oral Maxillofac Surg 1990 : 19 : 51-3   DOI   ScienceOn