Browse > Article

A semispherical SQUID magnetometer system using high sensitivity double relaxation oscillation SQUIDs for magnetoencephalographic measurements  

Lee, Yong-Ho (Korea Research Institute of Standards and Science)
Hyukchan Kwon (Korea Research Institute of Standards and Science)
Kim, Jin-Mok (Korea Research Institute of Standards and Science)
Kim, Kwoong (Korea Research Institute of Standards and Science)
Park, Yong-Ki (Korea Research Institute of Standards and Science)
Publication Information
Progress in Superconductivity and Cryogenics / v.5, no.1, 2003 , pp. 21-26 More about this Journal
Abstract
We designed and constructed a multichannel superconducting quantum interference device (SQUID) magnetometer system to measure magnetic fields from the human brain. We used a new type of SQUID, the double relaxation oscillation SQUID (DROS). With high flux-to-voltage transfers of the DROS, about 10 times larger than the dc SQUIDs, simple flux-locked loop circuits could be used for SQUID operation. Also the large modulation voltage of the DROS, typically being 100 $mutextrm{V}$, enabled stable flux-locked loop operation against the thermal offset voltage drift of the preamplifier. The magnetometers were fabricated using the Nb/AlOx/Nb junction technology. The SQUID system consists of 37 signal magnetometers, distributed on a semispherical surface, and 11 reference channels were installed to pickup background noises. External feedback was used to eliminate the magnetic coupling with the adjacent channels. The liquid helium dewar has a capacity of 29 L and boil-off rate of about 4 L/d with the total 48 channel insert. The magnetometer system has an average noise level of 3 fT/√Hz at 100 Hz, inside a shielded loon, and was applied to measure auditory-evoked fields.
Keywords
Citations & Related Records
연도 인용수 순위
  • Reference
1 (Double) relaxation oscillation SQUIDs with high flux-to-voltage transfer:Simulations and experiments /
[ D.J.Adelerhof;H.Nijstad;F.Flokstra;H.Rogalla ] / J. Appl. Phys.   DOI   ScienceOn
2 Advanced SQUID readout electronics. /
[ D.Drung;H. Weinstock, Dordecht(ed.) ] / SQUID Sensors: Fundamentals, Fabrication and Application.
3 SQUID magnetometers for biomagnetism and nondestructive testing: Important questions and initial answers /
[ J.P.Wikswo, Jr. ] / IEEE Trans. Supercond.   DOI   ScienceOn
4 Correction in the principal component elimenation method for excluding correlated noises in neuromagnetic evoked field measurements. /
[ K.Kim;Y.H.Lee;H.Kwon;J.M.Kim;C.S.Kang;Y.K.Park;I.S.Kim ] / IEEE Trans. Biomed. Eng.
5 Double relaxation oscillation SQUID with reference junction for biomagnetic multichannel applications. /
[ Y.H.Lee;H.C.Kwon;J.M.Kim;Y.K.Park;J.C.Park ] / Appl. Supercond.   DOI   ScienceOn
6 High sensitivity double relaxation oscillation superconducting quantum interference devices with large transfer from flux to voltage. /
[ D.J.Adelerhof;J.Kawai;G.Uehara;H.Kado ] / Rev. Sci. Instrum   DOI   ScienceOn
7 Magnetoencephalography-theory, instrumentation,and applications to noninvasive studies of the working human brain /
[ M.Hamalainen;R.Hari;R.J.Ilmoniemi;J.Knuutila;O.V.Lounasmaa ] / Rev. Mod. Phys.   DOI   ScienceOn
8 Modular middle-scale SQUID magnetometer systen for neuromagnetic research. /
[ Y.Hirata;S.Kuriki ] / IEEE Trans. Electron.
9 DC SQUIDs: Design,optimization and practical applications /
[ R.Cantor;H. Weinstock, Dordrecht(ed.) ] / SQUID Sensors: Fundamentals,Fabrication and Application.