Browse > Article

Standardization of Critical Temperature Measurement based on IEC International Standard  

Lee, Kyu-Won (Korea Research Institute of Standards and Science)
Kim, Kyu-Tae (Korea Research Institute of Standards and Science)
Publication Information
Progress in Superconductivity and Cryogenics / v.5, no.1, 2003 , pp. 123-127 More about this Journal
Abstract
For disseminating a ney IEC international standard of critical trmperature of NbTi,Nb$_3$Sn and Bi-2223 Composite Suterconductors, we develpted a measuring system and studied standardization of test method. The measuring system consisted of cryogenic reservoir, base plate, thermometer, voltmeter and current source. Various specimens of the Nbti, Nb$_3$Sn and Bi-2223 composite superconductors were tested using this system for measuring their critical temperatures. After measuring the resistance-temperature relation, the data were compensated with thermoelectric voltages for NbTi Nb$_3$Sn specimens. NbTi specimens showed 9.2 K ~ 9.5 K of transition temperature and Nb$_3$Sn specimen showed about 18 K. Bi-2223 specimens showed 104 K ~ 107 K of transition temperature.
Keywords
critical temperature; composite superconductor; resistance measurement; standardization;
Citations & Related Records
Times Cited By KSCI : 1  (Citation Analysis)
연도 인용수 순위
1 Fabrication of RRR measuring system for disseminating IEC international standard /
[ Kyu Won Lee;Kyu Tae Kim ] / Progress in Superconductivity   과학기술학회마을
2 /
[] / IEC 61788-10. Superconductivity-Part10: Critical Temperature Measuremet-Critical temperature of NbTi. Nb3Sn.and Bi-system oxide composite Superconductors by a esistance method