Browse > Article

Heat Flow Studies in Low Temperature Detectors  

Kim, Il-Hwan (Korea Research Institute of Standards and Science (KRISS))
Lee, Min-Kyu (Korea Research Institute of Standards and Science (KRISS))
Kim, Yong-Hamb (Korea Research Institute of Standards and Science (KRISS))
Publication Information
Abstract
Low temperature micro-calorimeters have been employed in the field of high resolution alpha spectrometers. These alpha detectors typically consist of a superconducting or metal absorber and a temperature sensor. The temperature sensor can be a transition edge sensor (TES), a metallic magnetic calorimeter (MMC) or other low temperature detectors for an accurate measurement of temperature change due to an alpha particle absorption. We report a recent study of the heat flow between a replaceable absorber and a temperature sensor. A piece of gold foil in $2.4{\times}2.7{\times}0.03\;mm^3$ is used as an absorber. A $40\;{\mu}m$ diameter Au:Er paramagnetic sensor is attached to another small piece of gold foil in $400{\times}200{\times}30\;{\mu}m^3$ to serve as the temperature sensor. This sensor assembly, Au:Er and gold foil, is placed on a miniature SQUID susceptometer in a gradiometric configuration. The thermal connection between the absorber and the sensor was made with three gold bonding wires. The measured thermal conductance shows a linear dependence to the temperature. The values are in a good agreement with Wiedemann-Franz type thermal conductance of the gold wires.
Keywords
Micro-calorimeter; Heat flow; Alpha spectrometry;
Citations & Related Records
연도 인용수 순위
  • Reference
1 I. H. Kim, "Development of cryogenic detectors for high resolution radioactivity measurement", Daejeon University, Master's thesis, DM037063 (2010).
2 F. Pobell, "Matter and Methods at Low Temperatures", Third, Revised and Expanded Edition, Springer, 3.3.5, 72-73 (2006).
3 Basunia M. S. "Nucl. Data Sheets" 107, 3323 (2006), http://www.nndc.bnl.gov/chart/decaysearchdirect.jsp?nuc=241AM&unc=nds.
4 A. Burck et al., "Microstructured Magnetic Calorimeter with Meander-Shaped Pickup Coil", J. Low Temp. Phys. 151, 337-344 (2008).   DOI   ScienceOn
5 Robert D. Horansky et al., "Superconducting calorimetric alpha particle sensors for nuclear nonproliferation applications", Appl. Phys. Lett., 93, 123504 ( 2008).   DOI   ScienceOn
6 Ranitzsch P. C. et al., "Development of cryogenic alpha spectrometers using metallic magnetic calorimeters", submitted in Nucl. Instr. and Mech. A.
7 Sang-Jun Lee et al., "Cryogenic measurement of alpha decay in a $4{\pi}$ absorber", J. Phys. G: Nucl. Part. Phys. 37, 055103 (7pp) (2010).
8 C. Enss(Editors), "Cryogenic Particle Detection", Springer, 151-216.
9 E. Steinbauer et al., "Energy resolution of silicon detectors: approaching the physical limit", Nucl. Instr. and Mech. B, 85, 642-649 (1994).   DOI   ScienceOn