Browse > Article
http://dx.doi.org/10.14347/kadt.2013.35.3.201

A Study on the Screw Loosening Torque According to the Type of Tightening the Implant Fixture and Abutment  

Park, Soo-Chul (Department of Dental Technology, Gimcheon University)
Kim, Hong-Sik (Department of Dental Technology, Gimcheon University)
Ham, Sung-Won (Department of Chemical Engineering, Kyungil University)
Publication Information
Journal of Technologic Dentistry / v.35, no.3, 2013 , pp. 201-207 More about this Journal
Abstract
Purpose: In this study, the loosening torque test was conducted with three implant products that are produced, approved and sold in Korea, which are manufactured in different fixture and abutment tightening methods (internal submerged type, internal morse taper type, and external type) to examine the loosening torque of the screw according to the method of tightening the implant fixture and abutment. Methods: In the loosening torque test, the three types of fixtures and abutments with different tightening methods were tightened by rotating them clockwise with a $30N{\cdot}cm$ force using a driver equipped with an electric torque meter. The results of the test are as follows. Results: The loosening torque values of the internal submerged type, internal morse taper type and external type implants were $24.10{\pm}0.742N{\cdot}cm$, $29.10{\pm}1.710N{\cdot}cm$, and $26.60{\times}1.636N{\cdot}cm$, respectively. Conclusion: The screw loosening torque values of the three fixture and abutment tightening methods were analyzed via Kruskal Wallis test layout, and they were significantly different (p<0.05).
Keywords
screw loosening torque; fixture; abutment;
Citations & Related Records
Times Cited By KSCI : 6  (Citation Analysis)
연도 인용수 순위
1 Koak JY, Heo SJ, Chang IT, Yim SH, Lee JY, Lee KR. The study on the removal torque of the diamond like carbon coated titanium abutment screws. J Korean Acad Prosthodont, 41(2), 128-135, 2003.   과학기술학회마을
2 Jamiyandorj O, Kim JH, Kim MS, Park YB, Shim JS. Comparison of removal torque between prefabricated and customized abutment screw. J Korean Acad Prosthodont, 50(4), 243-248, 2012.   과학기술학회마을   DOI   ScienceOn
3 Sakong J, Park SC. Concentration of elemental ions released from non-precious dental casting alloys. J Korean Acad dental Technology, 35 (1), 1-18, 2013.   DOI   ScienceOn
4 Carlson B, Carlsson GE. Prosthodontic complications in osseointegrated dental implant treatment. J Implant Dentistry, 3(4), 264, 1994.
5 Ekfeldt A. Carlsson GE, Borjesson G. Clinical evaluation of single-tooth restorations supported by osseointegrated implants: a retrospective study. Int J Oral Maxillofac Implants, 9(2), 179-183, 1994.
6 Eom TG, Suh SW, Jeon GR, Shin JW, Jeong CM. Effect of Tightening Torque on Abutment- Fixture Joint Stability using 3-Dimensional Finite Element Analysis. J Korean Acad Prosthodont, 47(2), 125-135, 2009.   과학기술학회마을   DOI   ScienceOn
7 Hemmings KW, Schmitt A, Zarb GA. Complications and maintenance requirements for fixed prostheses and overdentures in the edentulous mandible: a 5-year report. Int J Oral Maxillofac Implants, 9(2), 191-196, 1994.
8 Holmes DC, William RG, Vijay KG, John CK. Comparison of stress transmission in the IMZ implant system with polyoxymethylene or titanium intramobile element: a finite element stress analysis. Int J Oral Maxillofac Implants, 7(4), 450-458, 1992.
9 Jorn'eus L, Jemt T, Carlsson L. Loads and designs of screw joints for single crowns supported by osseointegrated implants. Int J Oral Maxillofac Implants, 7(3), 353-359, 1992.
10 Jemt T, Pattersson P. A 3-year follow-up study on single implant treatment. J Dent, 21(4), 203-208, 1993.   DOI   ScienceOn
11 Kallus T, Bressing C, Odont D. Loose gold screws frequently occur in full-arch fixed prostheses supported by osseointegrated implants after 5 years. Int J Oral Maxillofac Implants, 9(2), 169-178, 1994.
12 Lee CJ, Yang SE, Kim SG. Evaluation of reverse torque value of abutment screws on CAD/CAM custom-made implant abutments. J Korean Acad Prosthodont, 50(2), 128-134, 2012.   과학기술학회마을   DOI   ScienceOn
13 Park JK, Jeong CM, Jeon YC, Yoon JH. Influence of tungsten carbide/carbon coating of implantabutment screw on screw loosening. J Korean Acad Prosthodont, 46(2), 137-147, 2008.
14 Park SC, Choi SM, Kang JH. Corrosion Resistance Evaluation in the Co-Cr Alloys for the Full and Removable Partial Denture Metal Frameworks and the Porcelain-fused-tometal Crown. J Korean Acad dental Technology, 34(3), 237-245, 2012.   DOI
15 The performance and stability of the dental implant assessment tests guidelines. Korea Food & Drug Administration, 2006.