In vivo evaluation of accuracy and consistency of two electronic apex locators  

Pi, Chien-Yun (Department of Conservative Dentistry, Wonju Christian Hospital, Yonsei University)
Kim, Eui-Seong (Department of Conservative Dentistry, Yonsei University)
Jung, Il-Young (Department of Conservative Dentistry, Yonsei University)
Lee, Seung-Jong (Department of Conservative Dentistry, Yonsei University)
Publication Information
Restorative Dentistry and Endodontics / v.35, no.6, 2010 , pp. 453-460 More about this Journal
Abstract
Objectives: To evaluate the accuracy and consistency of two different apex locators at both the Apex and 0.5 marks. Materials and Methods: Twenty-six root canals was scheduled for extraction for periodontal or prosthodontic reasons. Thirteen canals were measured using Root ZX and the rest by i-ROOT. The root canal length was measured both the at 0.5 mark and the Apex mark. The file was then fixed to the toot, and the distance from the file tip to the major foramen of each canal was measured after removing the root dentin under the microscope so that the major foramen and the file tip were seen. Results: 1. When the Apex mark was used, 100% of both the Root ZX and i-ROOT groups were within 0.5 mm of the major foramen. 2. When 0.5 mark was used, 100% of the Root ZX group and 77% of the i-ROOT group were within 0.5 mm of the major foramen. 3. In terms of standard deviation and quartile value, the Apex mark was more consistent than 0.5 mark in the Root ZX group, and 0.5 mark was more consistent in the i-ROOT group, but there was no statistically significant difference when compared with t-test. 4. The root canal length difference between the Apex mark and 0.5 mark was 0.22 mm and 0.46 mm in the Root ZX and i-ROOT groups, respectively. Conclusions: In this study, the Apex mark was the more consistent mark. Therefore, it is recommended to subtract 0.5 mm, which is the average length between the apex and apical constriction, from the root canal length at the Apex mark to obtain the working length clinically.
Keywords
Accuracy; Consistency; Electronic apex locator; Microscope;
Citations & Related Records
Times Cited By KSCI : 1  (Citation Analysis)
연도 인용수 순위
1 Kim E, Marmo M, Lee C, Oh N, Kim I. An in vivo comparison of working length determination by only root-ZX apex locator versus combining root-ZX apex locator with radiographs using a new impression technique. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 2008;105(4):e79-e83.   DOI   ScienceOn
2 Dunlap CA, Remeikis NA, BeGole EA, Rauschenberger CR. An in vivo evaluation of an electronic apex locator that uses the ratio method in vital and necrotic canals. J Endod 1998;24(1):48-50.   DOI   ScienceOn
3 Pagavino G, Pace R, Baccetti T. A SEM study of in vivo accuracy of the Root ZX electronic apex locator. J Endod 1998;24(6):438-441.   DOI   ScienceOn
4 Venturi M, Breschi L. A comparison between two electronic apex locators: an ex vivo investigation. Int Endod J 2007;40(5):362-373.   DOI   ScienceOn
5 Goldberg F, Marroqun BB, Frajlich S, Dreyer C. In vitro evaluation of the ability of three apex locators to determine the working length during retreatment. J Endod 2005;31(9):676-678.   DOI   ScienceOn
6 Kim E, Lee S. Electronic apex locator. Dent Clin North Am 2004;48(1):35-54.   DOI   ScienceOn
7 Yamashita Y. A study of a new electric root canal measuring device using relative values of frequency response: influences of the diameter of apical foramen, the size of electrode, and the concentration of sodium hypochlorite. Jpn J Conserv Dent 1990;33:547-549.
8 Kobayashi C MK, Suda H, Sunada I. New practical model of the division method electronic root canal length measuring device. J Jpn Endodon Assoc 1991;12:143-148.
9 Kobayashi C, Suda H. New electronic canal measuring device based on the ratio method. J Endod 1994;20(3):111-114.   DOI   ScienceOn
10 Hwang GY, Roh BD, Kim ES, Lee SJ. In vitro evaluation of the consistency of two electronic apex locators. J Kor Acad Cons Dent 2008;33(1):20-27.   과학기술학회마을   DOI   ScienceOn
11 Vieyra JP, Acosta J, Mondaca JM. Comparison of working length determination with radiographs and two electronic apex locators. Int Endod J 2010;43(1):16-20.   DOI   ScienceOn
12 Yoon BH, Kim ES, Jung IY, Lee SJ. vitro evaluation of the consistency of two electronic apex locators. Department of Dentistry, The graduate school, Yonsei University. 2009.
13 Fouad AF, Rivera EM, Krell KV. Accuracy of the Endex with variations in canal irrigants and foramen size. J Endod 1993;19(2):63-67.   DOI   ScienceOn
14 Wrbas KT, Ziegler AA, Altenburger MJ, Schirrmeister JF. In vivo comparison of working length determination with two electronic apex locators. Int Endod J 2007;40(2):133-138.   DOI   ScienceOn
15 Seltzer S, Soltanoff W, Sinai I, Goldenberg A, Bender IB. Biologic aspects of endodontics. 3. Periapical tissue reactions to root canal instrumentation. Oral Surg Oral Med Oral Pathol 1968;26(5):694-705.   DOI   ScienceOn
16 Seltzer S, Soltanoff W, Sinai I, Smith J. Biologic aspects of endodontics. IV. Periapical tissue reactions to root-filled teeth whose canals had been instrumented short of their apices. Oral Surg Oral Med Oral Pathol 1969;28(5):724-738.   DOI   ScienceOn
17 Kuttler Y. Microscopic investigation of root apexes. J Am Dent Assoc 1955;50(5):544-552.
18 Cluster L. Exact methods of locating the apical foramen. J Natl Dent Assoc 1918;5:815-819.   DOI
19 Dummer PM, McGinn JH, Rees DG. The position and topography of the apical canal constriction and apical foramen. Int Endod J 1984;17(4):192-198.   DOI
20 Lee SJ, Nam KC, Kim YJ, Kim DW. Clinical accuracy of a new apex locator with an automatic compensation circuit. J Endod 2002;28(10):706-709.   DOI   ScienceOn
21 Suzuki K. Experimental study on iontophoresis. J Jpn Stomatol 1942;16:411-417.
22 Sunada I. New metod for measuring the length of the root canal. J Dent Res 1962;41(2):375-387.   DOI
23 Inoue N. An audiometric method for determining the length of the canals. J Can Dent Assoc 1973;39(9):630-636.