Evaluation of canal preparation with Ni-Ti rotary files by micro computed tomography  

Lee, Jeong-Ho (Deportment of Conservative Dentistry, College of Dentistry, Seoul Notional University)
Kim, Mi-Ja (Deportment of Conservative Dentistry, College of Dentistry, Seoul Notional University)
Seok, Chang-In (Deportment of Conservative Dentistry, College of Dentistry, Seoul Notional University)
Lee, Woo-Cheol (Deportment of Conservative Dentistry, College of Dentistry, Seoul Notional University)
Baek, Seung-Ho (Deportment of Conservative Dentistry, College of Dentistry, Seoul Notional University)
Publication Information
Restorative Dentistry and Endodontics / v.29, no.4, 2004 , pp. 378-385 More about this Journal
Abstract
The purpose of this study was to compare the effects of preparation with GT files and profiles .04 in shaping of root canals and reconstruct the three-dimensional root canal system using micro computed tomography 40 canals of the extracted human mandibular molars were used, and randomly distributed into two experimental groups. In group 1. canals were prepared by GT files. In group 2, Profiles .04. were used. Apical preparation size was #30. For each tooth pre and post operative cross-sectional images were obtained by the micro CT at 50 micron intervals. Pre and post operative cross-sectional images of 1, 2, 3, 5, and 8mm from the apex were compared. For each section. canal area and centering ratio were determined. For each tooth pre- and post-operative root canal volume from the furcation to the apex of the roots was calculated by three-dimensional image software. Following results were obtained: 1. At 8mm from the apex, area of dentin removed by GT rotary file was significantly larger than that by Profile .04. And at the other levels there was not a significant difference. 2. There was a trend for GT rotary file to remain more centered in the canals than Profile .04 at all levels. But at 3mm level. there was a statistically significant difference. 3. In root canal volume increments after instrumentation, there was no significant difference between two groups.
Keywords
Micro CT; Canal preparation; Ni-Ti rotary file; Three dimensional reconstruction;
Citations & Related Records
연도 인용수 순위
  • Reference
1 Bergmans L, Van Cleynenbreugel J, Wevers M, Lambrechts P. A methodology for quantitative evaluation of root canal instrumentation using microcomputed tomography. Int Endod J 34(5):390-398, 2001   DOI   ScienceOn
2 Peters OA, Laib A, Ruegsegger P, Barbakow F. Three-dimensional analysis of root canal geometry by high resolution computed tomography. J Dent Res 79(6):1405-1409, 2000   DOI   ScienceOn
3 Gambill JM, Alder M, Del Rio CE. Comparison of nickel titanium and stainless steel hand-file instrumentation using computed tomography. J Endod 22(7):369-375, 1996   DOI   ScienceOn
4 Garip Y, Gunday M. The use of computed tomography when comparing nickel-titanium and stainless steel files during preparation of simulated curved canals. Int Endod J 34(6):452-457, 2001   DOI   ScienceOn
5 Schneider SW. A comparison of the canal preparations in straight and curved root canals. Oral surg 32:271-275, 1971   DOI   ScienceOn
6 Peters OA, Laib A, Gohring TN, Barbakow F. Changes in root canal geometry after preparation assessed by high-resolution computed tomography. J Endod 27(1):1-6, 2001   DOI   ScienceOn
7 Blaskovic-Subat V, Smojver I, Maricic B, Sutalo J. A computerized method for the evaluation of root canal morphology. Int Endod J 28:290-296, 1995   DOI
8 Bjorndal L, Carlsen O, Thuesen G, Darvann T, Kreiborg S. External and internal macromorphology in 3D-reconstructed maxillary molars using computerized X-ray microtomography. Int Endod J 32(1):3-9, 1999   DOI   ScienceOn
9 Tachibana H, Matsumoto K. Applicability of X-ray computerized tomography in endodontics. Endodont Dent Traumatol 6:16-20, 1990   DOI
10 Kum KY, Spangberg L, Cha BY, Jung IY , Lee SJ, Lee CY. Shaping ability of three Profile rotary instrumentation techniques in simulated resin root canals. J Endod 26(12):719-723, 2000   DOI   ScienceOn
11 Kim YT. A comparison of three stainless steel instruments in the preparation of curved root canals in vitro J Kor Acad Cons Dent 26(1):9-15, 2001
12 Gluskin AH, Brown DC, Buchanan LS. A reconstructed computerized tomographic comparison of Ni-Ti rotary GT files versus traditional instruments in canals shaped by novice operators. Int Endod J 34(6):476-484, 2001   DOI   ScienceOn
13 Calhoun G, Montgomery S. The effects of four instrumentation technique on root canal shape. J Endod 14:273-277, 1988   DOI   ScienceOn
14 Bramante CM, Berbert A, Borges RP. A methology for evauation of root canal instrumentation. J Endod 13:243-245, 1987   DOI   ScienceOn
15 Rhodes JS, Ford TR, Lynch JA, Liepins PJ, Curtis RV. Micro-computed tomography: a new tool for experimental endodontology. Int Endod J 32(3):165-170, 1999   DOI   ScienceOn
16 Peters OA. Schonenberger K, Laib A. Effects of four Ni-Ti preparation techniques on root canal geometry assessed by micro computed tomography. Int Endod J 34(3):221-230, 2001   DOI   ScienceOn
17 Berutti E. Computerized analysis of the instrumentation of the root canal system. J Endod 19(5):236-238, 1993   DOI   ScienceOn
18 Dowker SE, Davis GR, Elliott JC. X-ray microtomogra phy. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 83(4):510-516, 1997   DOI   ScienceOn
19 Nielsen RB, Alyassin AM, Peters DD, Carnes DL, Lancaster J. Microcomputed tomography: an advanced system for detailed endodontic research. J Endod 21(11):561-568, 1995   DOI   ScienceOn
20 Lyroudia K, Mikrogeorgis G, Bakaloudi P, Kechagias E, Nikolaidis N, Pitas L. Virtual endodontics: three-dimensional tooth volume representations and their pulp cavity access. J Endod 28(8):599-602, 2002   DOI   ScienceOn
21 Schilder H. Cleaning and shaping the root canal. DCNA 18:269-296, 1967
22 Hulsmann M, Gambal A, Bahr R. An improved technique for the evaluation of root canal preparation. J Endod 25(9):599-602, 1999   DOI   ScienceOn
23 Kosa DA, Marshall G, Baumgartner JC. An analysis of canal centering using mechanical instrumentation techniques. J Endod 25(6) :441-445, 1999   DOI   ScienceOn
24 SkyScan 1072 Instruction manual $\copyright$ SkyScan 1998-2001, ver3 for 1072-scanners 4NT and 5NT
25 Kuttler S, Garala M, Perez R, Dorn SO. The endodontic cube: a system designed for evaluation of root canal anatomy and canal preparation. J Endod 27(8):533-536, 2001   DOI   ScienceOn