Browse > Article
http://dx.doi.org/10.5395/JKACD.2003.28.4.334

ANALYSIS OF TEMPERATURE RISE ON THE SURFACE OF BUCHANAN PLUGGER USING THERMOCOUPLE  

Cho, Jin-Suk (Dept. of Conservative Dentistry, College of Dentistry, DSRI, Chonnam Natl.Uni.)
Hwang, Yun-Chan (Dept. of Conservative Dentistry, College of Dentistry, DSRI, Chonnam Natl.Uni.)
Kim, Sun-Ho (Dept. of Conservative Dentistry, College of Dentistry, DSRI, Chonnam Natl.Uni.)
Hwang, In-Nam (Dept. of Conservative Dentistry, College of Dentistry, DSRI, Chonnam Natl.Uni.)
Choi, Bo-Young (School of Mechanical Systems Engineering, Chonnam Natl. Uni.)
Jeong, Young-Jin (School of Mechanical Systems Engineering, Chonnam Natl. Uni.)
Juhng, Woo-Nam (School of Mechanical Systems Engineering, Chonnam Natl. Uni.)
Oh, Won-Mann (Dept. of Conservative Dentistry, College of Dentistry, DSRI, Chonnam Natl.Uni.)
Publication Information
Restorative Dentistry and Endodontics / v.28, no.4, 2003 , pp. 334-340 More about this Journal
Abstract
This study was performed to evaluate the actual temperature rise on the surface of Buchanan plugger using thermocouple. The heat carrier system 'System B Heatsource'(Model 1005, Analytic Technologies, Redmond, WA, USA) and the Buchanan pluggers of F, FM, M and ML sizes are used for this study. The temperature was set to 200^{\circ}C on digital display and the power level on it was set to 10. Five thermocouples were placed in direct contact with the surface of each size of Buchanan's pluggers at 1 mm increments from the tip to the 4 mm length of shank. The heat control spring was touched for 5 seconds. and the temperature rise on the surface of the pluggers were measured at 1 sec intervals for more than 5 seconds with an accuracy of 0.01 using Data Logger. The data were statistically analyzed by one-way ANOVA. The results were as follows. 1. The position at which the temperature peaked was approximately at 1~2 mm far from the tip of Buchanan plugger (p<0.01). 2. The peak temperature was $215.25{\pm}2.28^{\circ}C$ in F plugger. $185.94{\pm}2.19^{\circ}C$ in FM plugger, $169.51{\pm}9.12^{\circ}C$ in M plugger, and 160.79{\pm}1.27^{\circ}C in ML plugger and the peak temperature was highest in F plugger and followed by. in descending order. FM plugger. M plugger. ML plugger showed the lowest peak temperature (p<0.01). 3. The temperature on the plugger was decreased with the increase of touching time. This results suggest that the actual temperature on the surface of the pluggers does not correlate well with the temperature set on digital display. Heat concentrates around the tip. The larger plugger reveals lower temperature rise relatively.
Keywords
Thermocouple; System B Heatsource; Temperature; Buchanan plugger;
Citations & Related Records
연도 인용수 순위
  • Reference
1 Jurcak JJ, Weller RN, Kulild JC and Donley DL : In vitro intracanal temperatures produced during warm lateral condensation of gutta-percha. J Endod 18:1-3, 1992   DOI   ScienceOn
2 Romero AD, Green DB and Wucherpfennig AL : Heat transfer to the periodontal ligament during root obturation procedures using an in vitro model. J Endod 26: 85-87, 2000   DOI   ScienceOn
3 Eriksson AR and Albrektsson T : Temperature threshold levels for heat-induced bone tissue injury : a vitalmicroscopic study in the rabbit. J Prosthet Dent 50:101-107, 1983   DOI   ScienceOn
4 최성아, 김선호, 황윤찬, 윤창, 오병주, 최보영, 정우남, 정선와, 황인남과 오원만 : 적외선열화상장치를 이용한 Buchanan plugger 표면의 온도상승 분석, 대한치과보존학회지 27:370-381, 2002
5 Buchanan LS : Continuous wave of condensation technique. Dentistry Today 1:60-64, 1996
6 Weller RN, Jurcak JJ, Donley DL and Kulild JC : A new model system for measuring intracanal temperatures. J Endod 17: 491-494, 1991   DOI   ScienceOn
7 Analytic Technology Corp. Instruction guidelines for system B HeatSource model 1005, Glendora, California, USA, 1997
8 Marlin J and Schilder H : Physical properties of guttapercha when subjected to heat and vertical condensation. Oral Surg 36:872-879, 1973   DOI   ScienceOn
9 Schilder H : Filling root canals in three dimensions. Dent Clin North Am 11:723-744, 1967
10 Floren JW, Weller RN, Pashley DH and Kimbrough WF : Changes in root surface temperatures with in vitro use of the System B HeatSource. J Endod 25:593-595, 1999   DOI   ScienceOn
11 Lee FS, Van Cura JE and BeGole E : A comparison of root surface temperatures using different obturation heat source. J Endod 24:617-620, 1998   DOI   ScienceOn
12 Nguyen NT : Obturation of the Root Canal System. Pathways of the Pulp. 6th ed., St. Louis, Mosby Co., 219-271, 1994
13 Hand RE, Hugel EF and Tsaknis PJ : Effect of a warm gutta-percha technique on the lateral periodontium. Oral Surg 42: 395-401, 1976   DOI   ScienceOn
14 Barkhordar RA, Goodis HE, Watanabe L and Koundjian J : Evaluation of temperature rise on the outer surface of teeth during root canal obturation technique. Quint lnt 21:585-588, 1990
15 Eriksson AR, Albrektsson T, Grane B and McQueen D: Thermal injury to bone : a vital-microscopic description of heat effect. lnt J Oral Surg 11:115-121, 1982   DOI
16 Eriksson AR and Albrektsson T : The effect of heat on bone regeneration. Oral Surg 42:705-711, 1984   DOI   ScienceOn