Browse > Article
http://dx.doi.org/10.4047/jap.2022.14.5.273

Microbiological cleaning and disinfection efficacy of a three-stage ultrasonic processing protocol for CAD-CAM implant abutments  

Gehrke, Peter (Department of Postgraduate Education, Oral Implantology, Center for Dentistry and Oral Medicine (Carolinum), Johann Wolfgang Goethe University)
Riebe, Oliver (HygCen Germany GmbH)
Fischer, Carsten (Sirius Ceramics Laboratory)
Weinhold, Octavio (Oral Surgery and Implant Dentistry)
Dhom, Gunter (Oral Surgery and Implant Dentistry)
Sader, Robert (Department for Oral, Cranio-Maxillofacial and Facial Plastic Surgery, Medical Center of the Goethe University Frankfurt)
Weigl, Paul (Department of Prosthodontics Department of Postgraduate Education, Oral Implantology, Center for Dentistry and Oral Medicine (Carolinum), Johann Wolfgang Goethe University)
Publication Information
The Journal of Advanced Prosthodontics / v.14, no.5, 2022 , pp. 273-284 More about this Journal
Abstract
PURPOSE. Computer-aided design and manufacturing (CAD-CAM) of implant abutments has been shown to result in surface contamination from site-specific milling and fabrication processes. If not removed, these contaminants can have a potentially adverse effect and may trigger inflammatory responses of the peri-implant tissues. The aim of the present study was to evaluate the bacterial disinfection and cleaning efficacy of ultrasonic reprocessing in approved disinfectants to reduce the microbial load of CAD-CAM abutments. MATERIALS AND METHODS. Four different types of custom implant abutments (total N = 32) with eight specimens in each test group (type I to IV) were CAD-CAM manufactured. In two separate contamination experiments, specimens were contaminated with heparinized sheep blood alone and with heparinized sheep blood and the test bacterium Enterococcus faecium. Abutments in the test group were processed according to a three-stage ultrasonic protocol and assessed qualitatively and quantitatively by determination of residual protein. Ultrasonicated specimens contaminated with sheep blood and E. faecium were additionally eluted and the dilutions were incubated on agar plates for seven days. The determined bacterial counts were expressed as colony-forming units (CFU). RESULTS. Ultrasonic reprocessing resulted in a substantial decrease in residual bacterial protein to less than 80 ㎍ and a reduction in microbiota of more than 7 log levels of CFU for all abutment types, exceeding the effect required for disinfection. CONCLUSION. A three-stage ultrasonic cleaning and disinfection protocol results in effective bacterial decontamination. The procedure is reproducible and complies with the standardized reprocessing and disinfection specifications for one- or two-piece CAD-CAM implant abutments.
Keywords
CAD-CAM abutments; Contamination; Disinfection; Bacterial decontamination; Ultrasonic cleaning;
Citations & Related Records
Times Cited By KSCI : 2  (Citation Analysis)
연도 인용수 순위
1 Yan S, Li M, Komasa S, Agariguchi A, Yang Y, Zeng Y, Takao S, Zhang H, Tashiro Y, Kusumoto T, Kobayashi Y, Chen L, Kashiwagi K, Matsumoto N, Okazaki J, Kawazoe T. Decontamination of titanium surface using different methods: an in vitro study. Materials (Basel) 2020;13:2287.   DOI
2 Jatzwauk L, Schone H, Pietsch H. How to improve instrument disinfection by ultrasound. J Hosp Infect 2001;48:S80-3.   DOI
3 Bentley EM. The value of ultrasonic cleaners in dental practice. Br Dent J 1994;177:53-6.   DOI
4 Guan A, Wang Y, Phillips KS. An extraction free modified o-phthalaldehyde assay for quantifying residual protein and microbial biofilms on surfaces. Biofouling 2018;34:925-34.   DOI
5 Zuhlsdorf B, Martiny H. Intralaboratory reproducibility of the German test method of prEN ISO 15883-1 for determination of the cleaning efficacy of washer-disinfectors for flexible endoscopes. J Hosp Infect 2005;59:286-91.   DOI
6 Kapos T, Evans C. CAD/CAM technology for implant abutments, crowns, and superstructures. Int J Oral Maxillofac Implants 2014;29:117-36.   DOI
7 Zarauz C, Pitta J, Pjetursson B, Zwahlen M, Pradies G, Sailer I. Esthetic outcomes of implant-supported single crowns related to abutment type and material: a systematic review. Int J Prosthodont 2021;34:229-49.   DOI
8 Ivanovski S, Lee R. Comparison of peri-implant and periodontal marginal soft tissues in health and disease. Periodontol 2000 2018;76:116-30.   DOI
9 Gehrke P, Bleuel K, Fischer C, Sader R. Influence of margin location and luting material on the amount of undetected cement excess on CAD/CAM implant abutments and cement-retained zirconia crowns: an in-vitro study. BMC Oral Health 2019;19:111.   DOI
10 Guideline compiled by German Society of Hospital Hygiene (DGKH, DGSV and AK). Validation and routine monitoring of automated cleaning and thermal disinfection processes for medical devices. 5th ed., 2017. p. 1-4.
11 Rupp F, Gittens RA, Scheideler L, Marmur A, Boyan BD, Schwartz Z, Geis-Gerstorfer J. A review on the wettability of dental implant surfaces I: theoretical and experimental aspects. Acta Biomater 2014;10:2894-906.   DOI
12 Canullo L, Tallarico M, Radovanovic S, Delibasic B, Covani U, Rakic M. Distinguishing predictive profiles for patient-based risk assessment and diagnostics of plaque induced, surgically and prosthetically triggered peri-implantitis. Clin Oral Implants Res 2016;27:1243-50.   DOI
13 Kampf G, Bloss R, Martiny H. Surface fixation of dried blood by glutaraldehyde and peracetic acid. J Hosp Infect 2004;57:139-43.   DOI
14 Welander M, Abrahamsson I, Berglundh T. The mucosal barrier at implant abutments of different materials. Clin Oral Implants Res 2008;19:635-41.
15 Canullo L, Micarelli C, Lembo-Fazio L, Iannello G, Clementini M. Microscopical and microbiologic characterization of customized titanium abutments after different cleaning procedures. Clin Oral Implants Res 2014;25:328-36.
16 Gehrke P, Tabellion A, Fischer C. Microscopical and chemical surface characterization of CAD/CAM zircona abutments after different cleaning procedures. A qualitative analysis. J Adv Prosthodont 2015;7:151-9.   DOI
17 Happe A, Roling N, Schafer A, Rothamel D. Effects of different polishing protocols on the surface roughness of Y-TZP surfaces used for custom-made implant abutments: a controlled morphologic SEM and profilometric pilot study. J Prosthet Dent 2015;113:440-7.   DOI
18 Piattelli A, Pontes AE, Degidi M, Iezzi G. Histologic studies on osseointegration: soft tissues response to implant surfaces and components. A review. Dent Mater 2011;27:53-60.   DOI
19 Geng D, Mao H, Wang J, Zhu X, Huang C, Chen L, Yang H, Xu Y. Protective effects of COX-2 inhibitor on titanium-particle-induced inflammatory osteolysis via the down-regulation of RANK/RANKL. Acta Biomater 2011;7:3216-21.   DOI
20 Nakajima K, Odatsu T, Shinohara A, Baba K, Shibata Y, Sawase T. Effects of cleaning methods for custom abutment surfaces on gene expression of human gingival fibroblasts. J Oral Sci 2017;59:533-9.   DOI
21 Yang SY, Zhang K, Bai L, Song Z, Yu H, McQueen DA, Wooley PH. Polymethylmethacrylate and titanium alloy particles activate peripheral monocytes during periprosthetic inflammation and osteolysis. J Orthop Res 2011;29:781-6.   DOI
22 Canullo L, Penarrocha D, Micarelli C, Massidda O, Bazzoli M. Hard tissue response to argon plasma cleaning/sterilisation of customised titanium abutments versus 5-second steam cleaning: results of a 2-year post-loading follow-up from an explanatory randomised controlled trial in periodontally healthy patients. Eur J Oral Implantol 2013;6:251-60.
23 Kim S, Choi C, Cha Y, Chang JS. The efficacy of convenient cleaning methods applicable for customized abutments: an in vitro study. BMC Oral Health 2021;21:78.   DOI
24 Rupp F, Liang L, Geis-Gerstorfer J, Scheideler L, Huttig F. Surface characteristics of dental implants: A review. Dent Mater 2018;34:40-57.   DOI
25 Mehl C, Kern M, Zimmermann A, Harder S, Huth S, Selhuber-Unkel C. Impact of cleaning procedures on adhesion of living cells to three abutment materials. Int J Oral Maxillofac Implants 2017;32:976-84.   DOI
26 Ozcan M, Bernasconi M. Adhesion to zirconia used for dental restorations: a systematic review and meta-analysis. J Adhes Dent 2015;17:7-26.
27 Inokoshi M, De Munck J, Minakuchi S, Van Meerbeek B. Meta-analysis of bonding effectiveness to zirconia ceramics. J Dent Res 2014;93:329-34.   DOI
28 Bergamo E, Zahoui A, Luri Amorin Ikejiri L, Marun M, Peixoto da Silva K, G Coelho P, Soares S, A Bonfante E. Retention of zirconia crowns to Ti-base abutments: effect of luting protocol, abutment treatment and autoclave sterilization. J Prosthodont Res 2021;65:171-5.   DOI
29 DIN EN 14885:2018. Chemical disinfectants and antiseptics - Application of European Standards for chemical disinfectants and antiseptics. 2018.
30 Canullo L, Tallarico M, Chu S, Penarrocha D, Ozcan M, Pesce P. Cleaning, disinfection, and sterilization protocols employed for customized implant abutments: an international survey of 100 universities worldwide. Int J Oral Maxillofac Implants 2017;32:774-8.   DOI
31 Bollen CM, Papaioanno W, Van Eldere J, Schepers E, Quirynen M, van Steenberghe D. The influence of abutment surface roughness on plaque accumulation and peri-implant mucositis. Clin Oral Implants Res 1996;7:201-11.   DOI
32 Gehrke P, Abazari C, Schlichter K, Fischer C, Duddeck D, Romanos GE, Weigl P. Qualitative and semi-quantitative assessment of processing-related surface contamination of one- and two-piece CAD/CAM abutments before and after ultrasonic cleaning. Materials (Basel) 2020;13:3225.   DOI
33 Infection control recommendations for the dental office and the dental laboratory. ADA Council on Scientific Affairs and ADA Council on Dental Practice. J Am Dent Assoc 1996;127:672-80.   DOI
34 Canullo L, Micarelli C, Iannello G. Microscopical and chemical surface characterization of the gingival portion and connection of an internal hexagon abutment before and after different technical stages of preparation. Clin Oral Implants Res 2013;24:606-11.   DOI
35 Canullo L, Penarrocha-Oltra D, Marchionni S, Bagan L, Penarrocha-Diago MA, Micarelli C. Soft tissue cell adhesion to titanium abutments after different cleaning procedures: Preliminary results of a randomized clinical trial. Med Oral Patol Oral Cir Bucal 2014;19:e177-83.
36 Mishra PK, Wu W, Rozo C, Hallab NJ, Benevenia J, Gause WC. Micrometer-sized titanium particles can induce potent Th2-type responses through TLR4-independent pathways. J Immunol 2011;187:6491-8.   DOI
37 ISO 17664-1. Processing of health care products - Information to be provided by the medical device manufacturer for the processing of medical devices - Part 1: Critical and semi-critical medical devices. International Standard Organization (ISO); Geneva; Switzerland, 2021.
38 Park JH, Olivares-Navarrete R, Baier RE, Meyer AE, Tannenbaum R, Boyan BD, Schwartz Z. Effect of cleaning and sterilization on titanium implant surface properties and cellular response. Acta Biomater 2012;8:1966-75.   DOI
39 Canullo L, Tallarico M, Penarrocha-Oltra D, Monje A, Wang HL, Penarrocha-Diago M. Implant abutment cleaning by plasma of argon: 5-year follow-up of a randomized controlled trial. J Periodontol 2016;87:434-42.   DOI
40 Kern M. On the scientific evidence that the sterilization of customized implant abutments is required. Eur J Oral Implantol 2015;8:219.
41 Technical information report AMI TIR30:2011 (R2016). A compendium of processes, materials, test methods, and acceptance criteria for cleaning reusable medical devices. Textbook, Association for the Advancement of Medical Instrumentation, 2011, p. 9-17.
42 German guideline for the validation of manual cleaning and manual chemical disinfection of medical devices compiled by DGKH-German Society for Hospital Hygiene. DGSV-German Society for Sterile Supply, AKI-working group instrument reprocessing in cooperation with the VAH-association for applied hygiene. 5th ed., 2017. Supplement; 5-14.
43 Gehrke P, Smeets R, Gosau M, Friedrich RE, Madani E, Duddeck D, Fischer C, Tebbel F, Sader R, Hartjen P. The influence of an ultrasonic cleaning protocol for cad/cam abutment surfaces on cell viability and inflammatory response in vitro. In Vivo 2019;33:689-98.   DOI
44 DIN EN ISO 15883-5. Washer-disinfectors - Part 5: Performance requirements and test method criteria for demonstrating cleaning efficacy. International Standard Organization (ISO); Geneva; Switzerland, 2021.
45 Lang R, Hiller KA, Kienbock L, Friedl K, Friedl KH. Influence of autoclave sterilization on bond strength between zirconia frameworks and Ti-base abutments using different resin cements. J Prosthet Dent 2022;127:617.e1-617.e6.
46 Gehrke P, Sing T, Fischer C, Spintzyk S, Geis-Gerstorfer J. Marginal and internal adaptation of hybrid abutment assemblies after central and local manufacturing, respectively. Int J Oral Maxillofac Implants 2018;33:808-14.   DOI
47 Pils D, Baeppler RJ, Junker R, Kielbassa AM, Nothdurft FP. Application of a standard autoclaving protocol does not harm structural integrity of two-piece zirconia abutments under detachment force testing. Clin Oral Investig 2019;23:3133-7.   DOI
48 Wiedenmann F, Liebermann A, Spintzyk S, Eichberger M, Stawarczyk B. Influence of different cleaning procedures on tensile bond strength between zirconia abutment and titanium base. Int J Oral Maxillofac Implants 2019;34:1318-27.   DOI