Browse > Article
http://dx.doi.org/10.4047/jkap.2012.50.3.184

Absorbed and effective dose for periapical radiography using portable and wall type dental X-ray machines  

Han, Won-Jeong (Department of Oral and Maxillofacial Radiology, College of Dentistry, Dankook University)
Publication Information
The Journal of Korean Academy of Prosthodontics / v.50, no.3, 2012 , pp. 184-190 More about this Journal
Abstract
Purpose: The purpose of this study was to measure the absorbed dose and to calculate the effective dose for one periapical radiography using the portable and wall type dental X-ray machines. Materials and methods: Thermoluminescent chips were placed at 25 sites throughout the layers of the head and neck of a tissue-equivalent human skull phantom. The man phantom was exposed with the portable and wall type dental X-ray machines. For one periapical radiography taken by portable dental X-ray machine, the exposure setting was 60 kVp, 2 mA and 0.2 seconds, while for one periapical radiography taken by wall type dental X-ray machine, exposure setting was 70 kVp, 8 mA and 0.074 seconds. Absorbed dose measurements were performed and equivalent doses to individual organs were summed using ICRP 103 to calculate effective dose. Results: In the upper anterior periapical radiography using portable dental X-ray machine and in the lower posterior periapical radiography using both machines, the highest absorbed dose was recorded at the mandible body. The effective dose in upper anterior periapical radiography using portable and wall type dental X-ray machines was $4{\mu}Sv$, $2{\mu}Sv$, respectively. In the lower posterior periapical radiography, the effective dose for each portable and wall type dental X-ray machines was $6{\mu}Sv$, $2{\mu}Sv$. Conclusion: It was recommended that the operator use prudently potable dental X-ray machine because that the effective dose in the periapical radiography using wall type dental X-ray machine was lower than that in the periapical radiography using portable dental X-ray machine.
Keywords
Radiography, Dental, Digital; Radiation Protection; Radiation Dosage; Portable Dental X-ray;
Citations & Related Records
Times Cited By KSCI : 2  (Citation Analysis)
연도 인용수 순위
1 Underhill TE, Chilvarquer I, Kimura K, Langlais RP, McDavid WD, Preece JW, Barnwell G. Radiobiologic risk estimation from dental radiology. Part I. Absorbed doses to critical organs. Oral Surg Oral Med Oral Pathol 1988;66:111-20.   DOI   ScienceOn
2 Han WJ. Absorbed and effective dose from periapical radiography by portable intraoral x-ray machine and panoramic radiography. J Korean Dent Assoc 2012;50:148-58.
3 Coy J, Vandre RH, Davidson WR. Use of the hand-held dental X-ray machine during joint operation, NATO exercise Display Determination-92. Mil Med 1997;162:575-7.
4 Varghese S, Kimmel A, Radmer T, Bradley TG, Bahcall J. In vitro evaluation of the XR-15 portable x-ray unit for forensic odontology. J Forensic Odontostomatol 2004;22:5-8.
5 Hermsen KP, Jaeger SS, Jaeger MA. Radiation safety for the NOMAD portable X-ray system in a temporary morgue setting. J Forensic Sci 2008;53:917-21.   DOI   ScienceOn
6 Charlton DG. Portable dental equipment: dental units and x-ray equipment. Gen Dent 2009;57:336-41.
7 Kim EK. Effect of the amount of battery charge on tube voltage in different hand-held dental x-ray systems. Imaging Sci Dent 2012;42:1-4.   과학기술학회마을   DOI   ScienceOn
8 Lee W. Current status of medical radiation exposure and regulation efforts. J Korean Med Assoc 2011;54:1248-52.   DOI
9 Ludlow JB, Davies-Ludlow LE, Brooks SL, Howerton WB. Dosimetry of 3 CBCT devices for oral and maxillofacial radiology: CB Mercuray, NewTom 3G and i-CAT. Dentomaxillofac Radiol 2006;35:219-26.   DOI   ScienceOn
10 Ludlow JB, Davies-Ludlow LE, Brooks SL. Dosimetry of two extraoral direct digital imaging devices: NewTom cone beam CT and Orthophos Plus DS panoramic unit. Dentomaxillofac Radiol 2003;32:229-34.   DOI   ScienceOn
11 Cho JY, Han WJ, Kim EK. Absorbed and effective dose from periapical radiography by portable intraoral x-ray machine. Korean J Oral Maxillofac Radiol 2007;37:149-56.
12 White SC, Pharoah MJ. Oral radiology: principles and interpretation. 6th ed. Mosby; 2008.
13 Avendanio B, Frederiksen NL, Benson BW, Sokolowski TW. Effective dose and risk assessment from detailed narrow beam radiography. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 1996;82:713-9.   DOI   ScienceOn
14 ICRP. ICRP Publication 103. The 2007 recommendations of the international commission on radiological protection. Elsevier; 2007.
15 Van Dis ML, Miles DA, Parks ET, Razmus TF. Information yield from a hand-held dental x-ray unit. Oral Surg Oral Med Oral Pathol 1993;76:381-5.   DOI   ScienceOn
16 Kim EK. Leakage and scattered radiation from hand-held dental x-ray unit. Korean J Oral Maxillofac Radiol 2007;37:65-8.
17 Choi SC, Choi HM. Absorbed dose in the full-mouth periapical radiography, panoramic radiography and zonography. J Korean Acad Oral Maxillofac Radiol 1999;29:255-60.