Analysis of thermal changes in bone by various insertion torques with different implant designs  

Kim, Min-Ho (Department of Prosthodontics, College of Dentistry, Seoul National University)
Yeo, In-Sung (Department of Prosthodontics, College of Dentistry, Seoul National University)
Kim, Sung-Hun (Department of Prosthodontics, College of Dentistry, Seoul National University)
Han, Jung-Seok (Department of Prosthodontics, College of Dentistry, Seoul National University)
Lee, Jai-Bong (Department of Prosthodontics, College of Dentistry, Seoul National University)
Yang, Jae-Ho (Department of Prosthodontics, College of Dentistry, Seoul National University)
Publication Information
The Journal of Korean Academy of Prosthodontics / v.49, no.2, 2011 , pp. 168-176 More about this Journal
Abstract
Purpose: This study aims at investigating the influence of various insertion torques on thermal changes of bone. A proper insertion torque is derived based on the thermal analysis with two different implant designs. Materials and methods: For implant materials, bovine scapula bone of 15 - 20 mm thickness was cut into 35 mm by 40 - 50 mm pieces. Of these, the pieces having 2 - 3 mm thickness cortical bone were used as samples. Then, the half of the sample was immersed in a bath of $36.5^{\circ}C$ and the other half was exposed to ambient temperature of $25^{\circ}C$, so that the inner and surface temperatures reached $36.5^{\circ}C$ and $28^{\circ}C$, respectively. Two types of implants ($4.5{\times}10\;mm$ Br${\aa}$nemark type, $4.8{\times}10\;mm$ Microthread type) were inserted into bovine scapula bone and the temperature was measured by a thermocouple at 0.2 mm from the measuring point. Finite element method (FEM) was used to analyze the thermal changes at contacting surface assuming that the sample is a cube of $4\;cm{\times}4\;cm{\times}2\;cm$ and a layer up to 2 mm from the top is cortical bone and below is a cancellous bone. Boundary conditions were set on the basis of the shape of cavity after implants. SolidWorks was used as a CAD program with the help of Abaqus 6.9-1. Results: In the in-vitro experiment, the Microhead type implant gives a higher maximum temperature than that of the Br${\aa}$nemark type, which is attributed to high frictional heat that is associated with the implant shape. In both types, an Eriksson threshold was observed at torques of 50 Ncm (Br${\aa}$nemark) and 35 Ncm (Microthread type), respectively. Based on these findings, the Microthread type implant is more affected by insertion torques. Conclusion: This study demonstrate that a proper choice of insertion torque is important when using a specific type of implant. In particular, for the Microthread type implant, possible bone damage may be expected as a result of frictional heat, which compensates for initial high success rate of fixation. Therefore, the insertion torque should be adjusted for each implant design. Furthermore, the operation skills should be carefully chosen for each implant type and insertion torque.
Keywords
Finite element method; Frictional heat; Implant diameter; Insertion torque; Thermal change;
Citations & Related Records
Times Cited By KSCI : 4  (Citation Analysis)
연도 인용수 순위
1 Cochran DL, Buser D, ten Bruggenkate CM, Weingart D, Taylor TM, Bernard JP, Peters F, Simpson JP. The use of reduced healing times on ITI implants with a sandblasted and acid-etched (SLA) surface: early results from clinical trials on ITI SLA implants. Clin Oral Implants Res 2002;13:144-53.   DOI   ScienceOn
2 Sutter F, Weber HP, Sorensen J, Belser U. The new restorative concept of the ITI dental implant system. Design and engineering. Int J Periodont Restorative Dent 1993;13:409-31.
3 Misch CE. Implant design considerations for the posterior regions of the mouth. Implant Dent 1999;8:376-86.   DOI   ScienceOn
4 Strong JT, Misch CE, Bidez MW, Nalluri P. Functional surface area: thread form and diameter optimization for implant body design. Compendium 1998;19:4-9.
5 Adell R, Lekholm U, Rockler B, Branemark PI. A 15-year study of osseointegrated implants in the treatment of the edentulous jaw. Int J Oral Surg 1981;10:387-416.   DOI
6 Adell R, Lekholm U. Surgical procedures. In: Zarb GA. Branemark PI, Albrektsson T, eds. Tissue-integrated prostheses: Osseointegration in clinical dentistry. Chicago: Quintessence Publishing 1985, pp. 1-35.
7 Bass SL, Triplett RG. The effects of preoperative resorption and jaw anatomy on implant success. A report of 303 cases. Clin Oral Implants Res 1991;2:193-8.   DOI   ScienceOn
8 Hutton JE, Heath MR, Chai JY, Harnett J, Jemt T, Johns RB, McKenna S, McNamara DC, van Steenberghe D, Taylor R, et al. Factors related to success and failure rates at 3-year follow-up in a multicenter study of overdentures supported by Branemark implants. Int J Oral Maxillofac Implants 1995;10:33-42.
9 Friberg B, Jemt T, Lekholm U. Early failures in 4,641 consecutively placed Branemark dental implants: a study from stage 1 surgery to the connection of completed prostheses. Int J Oral Maxillofac Implants 1991;6:142-6.
10 Branemark PI, Hansson BO, Adell R, Breine U, Lindstrom J, Hallen O, Ohman A. Osseointegrated implants in the treatment of the edentulous jaw. Experience from a 10-year period. Scand J Plast Reconstr Surg Suppl 1977;16:1-132.
11 O'Sullivan D, Sennerby L, Meredith N. Measurements comparing the initial stability of five designs of dental implants: a human cadaver study. Clin Implant Dent Relat Res 2000;2:85-92.   DOI   ScienceOn
12 Lundskog J. Heat and bone tissue. An experimental investigation of the thermal properties of bone and threshold levels for thermal injury. Scand J Plast Reconstr Surg 1972;9:1-80.
13 Eriksson AR, Albrektsson T. Temperature threshold levels for heat-induced bone tissue injury: a vital-microscopic study in the rabbit. J Prosthet Dent 1983;50:101-7.   DOI   ScienceOn
14 Sevitt S. Local vascular changes in burned skin. Proc R Soc Med 1954;47:225-8.
15 Park JH, Lim YJ, Kim MJ, Kwon HB. The effect of various thread designs on the initial stability of taper implants. J Adv Prosthodont 2009;1:19-25.   DOI
16 Norton MR. An in vitro evaluation of the strength of an internal conical interface compared to a butt joint interface in implant design. Clin Oral Implants Res 1997;8:290-8.   DOI   ScienceOn
17 Huh JB, Ko SM. The comparative study of thermal inductive effect between internal connection and external connection implant in abutment preparation. J Korean Acad Prosthodont 2007;45:60-70.
18 Cho JM, Cho U, Yun MJ, Jeong CM, Jeon YC. Influence of implant diameter and length changes on initial stability J Korean Acad Prosthodont 2009;47:335-41.   DOI
19 Kim SH, Kim SJ, Lee KW, Han DH. The effects of local factors on the survival of dental implants: A 19 year retrospective study. J Korean Acad Prosthodont 2010;48:28-40.   DOI
20 Norton MR. Assessment of cold welding properties of the internal conical interface of two commercially available implant systems. J Prosthet Dent 1999;81:159-66.   DOI   ScienceOn
21 Norton MR. In vitro evaluation of the strength of the conical implant-to-abutment joint in two commercially available implant systems. J Prosthet Dent 2000;83:567-71.   DOI   ScienceOn
22 Fish J, Belytschko T. A first course in finite elements. John Willey & Sons, 2007.
23 Chen HL, Gundjian AA. Specific heat of bone. Med Biol Eng Comput 1976;14:548-50.   DOI   ScienceOn
24 Davim JP, Marques N. Dynamical experimental study of friction and wear behaviour of bovine cancellous bone sliding against a metallic counterface in a water lubricated environment. J Mater Proc Tech 2004;152:389-94.   DOI   ScienceOn
25 Esposito M, Thomsen P, Ericson LE, Sennerby L, Lekholm U. Histopathologic observations on late oral implant failures. Clin Implant Dent Relat Res 2000;2:18-32.   DOI   ScienceOn
26 Sean RH Davidson, David F. James. Measurement of thermal conductivity of bovine cortical bone. Med Eng Phy 2000;22:741-7.   DOI   ScienceOn
27 Hosokawa A, Otani T. Ultrasonic wave propagation in bovine cancellous bone. J Acoust Soc Am 1997;101:558-62.   DOI   ScienceOn
28 Esposito M, Hirsch J, Lekholm U, Thomsen P. Differential diagnosis and treatment strategies for biologic complications and failing oral implants: a review of the literature. Int J Oral Maxillofac Implants 1999;14:473-90.
29 Truhlar RS, Morris HF, Ochi S, Winkler S. Second-stage failures related to bone quality in patients receiving endosseous dental implants: DICRG Interim Report No. 7. Dental Implant Clinical Research Group. Implant Dent 1994;3:252-5.   DOI   ScienceOn
30 Jaffin RA, Berman CL. The excessive loss of Branemark fixtures in type IV bone: a 5-year analysis. J Periodontol 1991;62:2-4.   DOI
31 Levine RA, Clem DS 3rd, Wilson TG Jr, Higginbottom F, Saunders SL. A multicenter retrospective analysis of the ITI implant system used for single-tooth replacements: preliminary results at 6 or more months of loading. Int J Oral Maxillofac Implants 1997;12:237-42.
32 Goodacre CJ, Bernal G, Rungcharassaeng K, Kan JY. Clinical complications with implants and implant prostheses. J Prosthet Dent 2003;90:121-32.   DOI   ScienceOn
33 Eriksson AR, Albrektsson T. Temperature threshold levels for heat-induced bone tissue injury: a vital-microscopic study in the rabbit. J Prosthet Dent 1983;50:101-7.   DOI   ScienceOn